Published on in Vol 15, No 8 (2013): August

Standardized, App-Based Disinfection of iPads in a Clinical and Nonclinical Setting: Comparative Analysis

Standardized, App-Based Disinfection of iPads in a Clinical and Nonclinical Setting: Comparative Analysis

Standardized, App-Based Disinfection of iPads in a Clinical and Nonclinical Setting: Comparative Analysis

Original Paper

1PL Reichertz Institute for Medical Informatics, Hannover Medical School, Hannover, Germany

2Institute for Medical Microbiology and Hospital Epidemiology, Hannover Medical School, Hannover, Germany

3Institute of Cell Biology in the Center of Anatomy, Hannover Medical School, Hannover, Germany

*all authors contributed equally

Corresponding Author:

Urs-Vito Albrecht, Dr med, MPH

PL Reichertz Institute for Medical Informatics

Hannover Medical School

Carl-Neuberg-Str 1

Hannover, 30625

Germany

Phone: 49 511 532 ext 3508

Fax:49 511 532 2517

Email: albrecht.urs-vito@mh-hannover.de


Background: With the use of highly mobile tools like tablet PCs in clinical settings, an effective disinfection method is a necessity. Since manufacturers do not allow cleaning methods that make use of anything but a dry fleece, other approaches have to be established to ensure patient safety and to minimize risks posed by microbiological contamination.

Objective: The ability of isopropanol wipes to decontaminate iPads was evaluated prospectively in a observer blinded, comparative analysis of devices used in a clinical and a nonclinical setting.

Methods: 10 new iPads were randomly deployed to members of the nursing staff of 10 clinical wards, to be used in a clinical setting over a period of 4 weeks. A pre-installed interactive disinfection application (deBac-app, PLRI MedAppLab, Germany) was used on a daily basis. Thereafter, the number and species of remaining microorganisms on the surface of the devices (13 locations; front and back) was evaluated using contact agar plates. Following this, the 10 iPads were disinfected and randomly deployed to medical informatics professionals who also used the devices for 4 weeks but were forbidden to use disinfecting agents. The quality of a single, standardized disinfection process was then determined by a final surface disinfection process of all devices in the infection control laboratory. No personal data were logged with the devices. The evaluation was performed observer blinded with respect to the clinical setting they were deployed in and personnel that used the devices.

Results: We discovered a 2.7-fold (Mann-Whitney U test, z=-3.402, P=.000670) lower bacterial load on the devices used in the clinical environment that underwent a standardized daily disinfection routine with isopropanol wipes following the instructions provided by “deBac-app”. Under controlled conditions, an average reduction of the mainly Gram-positive normal skin microbiological load of 99.4% (Mann-Whitney U test, z=-3.1798, P=.001474) for the nonclinical group and 98.1% (Mann-Whitney U test, z=3.1808, P=.001469) for the clinical group was achieved using one complete disinfecting cycle.

Conclusions: Normal use of tablet PCs leads to a remarkable amount of microbial surface contamination. Standardized surface disinfection with isopropanol wipes as guided by the application significantly reduces this microbial load. When performed regularly, the disinfection process helps with maintaining a low germ count during use. This should reduce the risk of subsequent nosocomial pathogen transmission. Unfortunately, applying a disinfection procedure such as the one we propose may lead to losing the manufacturer’s warranty for the devices; this remains an unsolved issue.

J Med Internet Res 2013;15(8):e176

doi:10.2196/jmir.2643

Keywords



Infections are called nosocomially acquired if they occur during a hospital stay. They have an enormous clinical and economical impact for health care systems [Rosenthal VD, Bijie H, Maki DG, Mehta Y, Apisarnthanarak A, Medeiros EA, INICC members. International Nosocomial Infection Control Consortium (INICC) report, data summary of 36 countries, for 2004-2009. Am J Infect Control 2012 Jun;40(5):396-407. [CrossRef] [Medline]1]. In addition, multidrug resistant pathogens represent an increasing problem in hospitals these days [Gagliotti C, Balode A, Baquero F, Degener J, Grundmann H, Gür D, EARS-Net Participants (Disease Specific Contact Points for AMR). Escherichia coli and Staphylococcus aureus: bad news and good news from the European Antimicrobial Resistance Surveillance Network (EARS-Net, formerly EARSS), 2002 to 2009. Euro Surveill 2011;16(11):pii/19834 [FREE Full text] [Medline]2]. Besides the hands of health care workers (HCW), contaminated medical devices and surfaces play an important role in the transmission of bacterial pathogens. This necessitates considerable effort for environmental infection control in order to prevent the spread of all kinds of microorganisms between patients [Sehulster L, Chinn RY, CDC, HICPAC. Guidelines for environmental infection control in health-care facilities. Recommendations of CDC and the Healthcare Infection Control Practices Advisory Committee (HICPAC). MMWR Recomm Rep 2003 Jun 6;52(RR-10):1-42 [FREE Full text] [Medline]3].

Mobile devices such as mobile phones or personal digital assistants (PDAs) represent a rather novel “surface” in the hospital setting that may also play an important role in the transmission of nosocomial pathogens. Nowadays, such devices are frequently used by physicians and other medical staff for both clinical practice and educational purposes [Dasari KB, White SM, Pateman J. Survey of iPhone usage among anaesthetists in England. Anaesthesia 2011 Jul;66(7):630-631. [CrossRef] [Medline]4,Payne KB, Wharrad H, Watts K. Smartphone and medical related App use among medical students and junior doctors in the United Kingdom (UK): a regional survey. BMC Med Inform Decis Mak 2012;12:121 [FREE Full text] [CrossRef] [Medline]5]. The number and availability of medical applications (apps) on smartphones is constantly rising and includes drug guides, medical calculators, coding and billing apps, textbooks and other reference materials, classification and treatment algorithms, as well as information regarding general medical knowledge [Franko OI, Tirrell TF. Smartphone app use among medical providers in ACGME training programs. J Med Syst 2012 Oct;36(5):3135-3139. [CrossRef] [Medline]6]. However, contamination of a device’s surface occurs every time it is being touched by a user [Gould D. Commentary: Ulger F et al. (2009). Are we aware how contaminated our mobile phones with nosocomial pathogens? Nurs Crit Care 2009;14(4):213-214. [CrossRef] [Medline]7], and there are several reports showing that these devices may then serve as vectors for transmission of pathogens to patients [Borer A, Gilad J, Smolyakov R, Eskira S, Peled N, Porat N, et al. Cell phones and Acinetobacter transmission. Emerg Infect Dis 2005 Jul;11(7):1160-1161 [FREE Full text] [CrossRef] [Medline]8,Goldblatt JG, Krief I, Klonsky T, Haller D, Milloul V, Sixsmith DM, et al. Use of cellular telephones and transmission of pathogens by medical staff in New York and Israel. Infect Control Hosp Epidemiol 2007 Apr;28(4):500-503. [CrossRef] [Medline]9]. In a review of data from studies published between 2002 and 2008, Brady et al [Brady RR, Verran J, Damani NN, Gibb AP. Review of mobile communication devices as potential reservoirs of nosocomial pathogens. J Hosp Infect 2009 Apr;71(4):295-300. [CrossRef] [Medline]10] showed that 9-25% of mobile communication devices were contaminated with pathogenic bacteria. More recent prevalence studies report contamination rates as high as 44-95% [Gould D. Commentary: Ulger F et al. (2009). Are we aware how contaminated our mobile phones with nosocomial pathogens? Nurs Crit Care 2009;14(4):213-214. [CrossRef] [Medline]7,Ramesh J, Carter AO, Campbell MH, Gibbons N, Powlett C, Moseley H, et al. Use of mobile phones by medical staff at Queen Elizabeth Hospital, Barbados: evidence for both benefit and harm. J Hosp Infect 2008 Oct;70(2):160-165. [CrossRef] [Medline]11-Sadat-Ali M, Al-Omran AK, Azam Q, Bukari H, Al-Zahrani AJ, Al-Turki RA, et al. Bacterial flora on cell phones of health care providers in a teaching institution. Am J Infect Control 2010 Jun;38(5):404-405. [CrossRef] [Medline]15]. Like mobile phones, tablet PCs (for example the iPad) are also frequently touched during patient care and bacteria may reside on their surfaces (Figure 1).

Before our study began, tests conducted with microbiological swabs showed that brand new iPad devices are not significantly contaminated with bacteria or fungi. However, the extent of contamination of tablets that have already been used remains yet unknown. Considering that bacteria may survive for days and weeks on inanimate surfaces [Kramer A, Schwebke I, Kampf G. How long do nosocomial pathogens persist on inanimate surfaces? A systematic review. BMC Infect Dis 2006;6:130 [FREE Full text] [CrossRef] [Medline]16], there is a need to determine the extent of contamination and to implement proper routine decontamination measures.

The present study was set up to determine (1) the microbiological flora (qualitative and quantitative) on tablet PCs as a result of use under the usual conditions that can be found in clinical as well as in nonclinical settings and (2) the quality of a standardized disinfection process as guided by an app specifically programmed for this purpose.

Figure 1. Aluminium backside of a tablet PC with fingerprints and other residue visible under fluorescent light and corresponding scanning electron microscopy pictures of cocci on the device in 2 magnifications (Bars: B1=5µm and B2=10µm).
View this figure

Clinical Setting

A set of 10 Apple iPads was randomly distributed to be used by nursing staff after obtaining informed consent. Various wards of Hannover Medical School, a tertiary care German university hospital, were included. They covered nonsurgical as well as surgical specialties. For the clinical setting, the disinfection study was an add-on to a larger trial dealing with various aspects [Albrecht U, Behrends M, Schmeer R, Matthies HK, von Jan U. Usage of Multilingual Mobile Translation Applications in Clinical Settings. JMIR Mhealth and Uhealth 2013 Apr 23;1(1):e4. [CrossRef]17] of using iPads on the wards of the Hannover Medical School. Altogether, approximately 160 staff members on the wards had the opportunity to use the devices. It was not possible to determine how many different individuals had used the iPads. As mentioned in [Albrecht U, Behrends M, Schmeer R, Matthies HK, von Jan U. Usage of Multilingual Mobile Translation Applications in Clinical Settings. JMIR Mhealth and Uhealth 2013 Apr 23;1(1):e4. [CrossRef]17], the return rate for questionnaires dealing with the overall project was approximately 26% (42/160); this can be assumed as the minimum of actual individual users. Regarding age and gender, the demographics of the participants who had returned the questionnaire paralleled the values for the nursing staff at the Hannover Medical School, where, at the end of 2012, 83% out of the total 2596 employees of the nursing staff were female (data obtained from the human resources department). For our study, 85% (36/42) were female and the age distribution of 69% (29/42) for those below 45 years of age and 31% (13/42) for those 45 years of age or older was also comparable. It was not possible to determine how they had used the devices and whether all or only some of them had disinfected the devices aided by deBac-app.

There were no additional accessories such as protective cases, polyurethane foils, rubber or silicone covers, since these may add additional, hard-to-disinfect niches for contaminating pathogens. Instead, we recommended disinfection of the plain surface of every device on a daily basis: once at the beginning of the working shift as well as anytime when an obvious contamination had occurred. Standardization of the disinfection process was achieved by using the “deBac-app” (from the MedAppLab, Hannover, Germany), which was preinstalled on all iPads. This app was designed by our research group and is an interactive cleaning and disinfection guide that is available free from Apple’s App Store [PLRI MedAppLab. deBac-app. 2013.   URL: https://itunes.apple.com/us/app/debac-app/id453976138?mt=8&ign-mpt=uo%3D2 [accessed 2013-02-08] [WebCite Cache]18]. It provides users with simple-to-follow instructions on how to properly disinfect the entire device (

Multimedia Appendix 1

Standardized disinfection process of an iPad as guided by the corresponding application (“deBac-app”), the app documents the cleaning attempts of the device’s frame, front and back side.

PDF File (Adobe PDF File), 132KBMultimedia Appendix 1). Every disinfection process was logged locally on the respective device. No personal data were stored or transmitted to the observers. The study time period of this clinical setting study arm was set to 4 weeks between August 28 and September 19, 2011, in order to achieve a steady state in terms of usage and reprocessing. A flowchart of the study’s timeline is shown in Figure 2.

Figure 2. Flowchart on the timeline of the 2 settings of iPad usage (clinical and nonclinical) and contact points used for microbiological sampling of an iPad (surface material was glass on the front side [points 1-6], aluminium [points 7-12], and plastic [point 13] on the backside).
View this figure

Nonclinical Setting

10 iPads were also used for a 4-week time period, between September 23 and October 21, 2011, by 10 members of the medical information technology staff (30% or 3/10 female, aged 23-63 years, mean 41.7, SD 10.6) in the nonclinical study arm. The devices were randomly distributed to the staff after obtaining informed consent for participation in the study. As these staff members generally do not have contact to patients, no regular disinfection of the tablet PCs was performed (Figure 2). All 10 participants of the nonclinical part of the study belonged to the P.L. Reichertz Institute for Medical Informatics at the Hannover Medical School and had volunteered to participate. For the nonclinical setting, there were no dropouts during the course of the study and all participants stated that they had frequently used their devices during the 4-week period.

Microbiological Testing

After 4 weeks of usage, all devices were examined for growth of microorganisms on their surfaces as soon as they arrived at the microbiological laboratory. Culture media with a contact area of 25 cm2 (CASO contact agar plates, Heipha diagnostica Dr. Müller GmbH) were used. These culture media support growth of most aerobic bacteria, molds, and yeasts. Since it is known that the adherence and survival of microorganisms may vary depending upon the type of surfaces material [Sadat-Ali M, Al-Omran AK, Azam Q, Bukari H, Al-Zahrani AJ, Al-Turki RA, et al. Bacterial flora on cell phones of health care providers in a teaching institution. Am J Infect Control 2010 Jun;38(5):404-405. [CrossRef] [Medline]15], we decided to perform the microbiological sampling for 13 different contact points of each iPad (Figure 2, right), including locations on the front and the back side of the device. These contact points covered all types of material to be found on the surface of the devices (glass, plastic, and aluminium). The contact plates were incubated at 37°C for 18 hours under aerobic conditions. Colony forming units (CFU) were then counted, and species differentiation was carried out in the microbiology laboratory of our facility according to the national guidelines of the German Institute for Standardization DIN EN ISO 15189 as certified by the German Accreditation Council (DAR). The evaluation was conducted in such a way that the laboratory was unaware of the setting to which the iPads had been deployed.

Electron Microscopy

For photo documentation of bacterial contamination on the tablet PC’s surfaces, scanning electron microscopy (SEM) was applied (Figure 1). Specimens were fixed in 0.1M Na-Cacodylate-HCl buffer (pH 7.3) containing 3% glutaraldehyde for at least 4 hours at 4°C. After washing in the buffer of the fixative, the cells were postfixed in 2% OsO4 buffered in Na-Cacodylate 0.1 M for 90 minutes at room temperature, dehydrated in ascending concentrations of acetone, and subsequently dried in a Balzers CPD 030 critical point dryer (Bal-Tec-AG). After mounting on aluminium stubs with conductive carbon cement (Plano) and sputter coating with gold in a Polaron E 5400 sputter coater, the samples were investigated in a Philips SEM 505 scanning electron microscope at an acceleration voltage of 10kV. Images were recorded using the SEM software version 2.0 [Gebert A, Preiss G. A simple method for the acquisition of high-quality digital images from analog scanning electron microscopes. J Microsc 1998 Sep;191(3):297-302. [Medline]19].

Final Reprocessing

After primary microbiological testing, the devices underwent final reprocessing performed by laboratory staff (Figure 2). For the clinical setting, with one exception, all devices were disinfected using isopropanol wipes (mikrozid-AF, Schülke & Mayr GmbH) using the 6-step disinfection process guided by deBac-app as it was described above (

Multimedia Appendix 1

Standardized disinfection process of an iPad as guided by the corresponding application (“deBac-app”), the app documents the cleaning attempts of the device’s frame, front and back side.

PDF File (Adobe PDF File), 132KBMultimedia Appendix 1). The remaining 10th iPad did not get reprocessed and, thus, served as a negative control for the disinfection process. For the 10 iPads that had been used in the nonclinical setting, we chose a slightly different approach: 8 of them were disinfected as described, while 1 remained without treatment and 1 was simply cleaned (but not disinfected) by using a new “soft, lint-free cloth”, without any liquid cleaning agents, as recommended in the instructions of the manufacturer of the iPad [Apple Inc. How to clean Apple products. 2013.   URL: http://support.apple.com/kb/ht3226 [accessed 2013-02-08] [WebCite Cache]20]. A final microbiological testing as described above was performed following the different, aforementioned types of reprocessing performed in the laboratory (Figure 2).

Nasal Swabs

People may be physiologically colonized by Staphylococcus aureus in the anterior nose, and some of these strains even show multidrug resistance, so-called methicillin-resistant Staphylococcus aureus (MRSA) [Stenehjem E, Rimland D. MRSA nasal colonization burden and risk of MRSA infection. Am J Infect Control 2013 May;41(5):405-410. [CrossRef] [Medline]21]. For a comparison of the Staphylococcus aureus colonization status of the 10 medical informatics professionals and the surface of their devices, nasal swabs (Transystem, Lot 9275, Hain Lifescience) were taken from the users after informed consent was obtained. Swabs were cultured on Columbia 5% sheep blood agar (Becton, Dickinson) overnight at 37°C. Species identification and susceptibility testing were then performed according to laboratory standard operation protocols.

Statistical Analysis

We expected a very strong effect of the applied method on reduction of the CFU according to the literature [Panousi MN, Williams GJ, Girdlestone S, Hiom SJ, Maillard JY. Evaluation of alcohol wipes used during aseptic manufacturing. Lett Appl Microbiol 2009 May;48(5):648-651. [CrossRef] [Medline]22]. Therefore, a smaller sample size was expected to be sufficient to demonstrate the efficacy of the disinfection. Calculated for a paired nonparametric test [Noether GE. Sample Size Determination for Some Common Nonparametric Tests. Journal of the American Statistical Association 1987 Jun;82(398):645 [FREE Full text]23], a sample size of 6 iPads per group was considered sufficient to show a significant effect of a reduction of 98% with beta=.20. The sample size calculation was performed with nQuery Advisor V.7, Statistical Solutions. Since a normal distribution of bacteria on the devices and sampled locations could not be confirmed by descriptive statistics, the Wilcoxon signed-rank test and the Mann-Whitney U test were applied (IBM SPSS Statistics Version 20). All tests were performed two-sided, with alpha=.05. Microsoft Excel 2007 was used for qualitative descriptive and quantitative data analysis. Intraclass correlation could not be confirmed following Shrout and Fleiss’ two-way random single measures (consistency) approach computed with SPSS [Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull 1979 Mar;86(2):420-428. [CrossRef] [Medline]24].


Qualitative and Quantitative Analysis of Microbiological Flora

A total of 6811 colonies representing microbial growth were detected during the initial testing of the iPads after use: 1842 CFU on tablet PCs from the hospital wards where the devices had been disinfected regularly, compared to 4,969 CFU recovered from tablet PCs from the nonclinical setting where daily disinfections had not been carried out (Mann-Whitney U test, z=-3.402, P=.000670). The distribution of pathogens on the various parts of the tablet PCs is shown in Table 1. Note that most pathogens found in both study arms (clinical and nonclinical) were gram-positive bacteria. A more detailed analysis of the species is shown in Table 2. The majority of microorganisms were members of the physiological microbiota of the human skin. The distribution did not differ significantly between both study arms. The main bacterial genera grown from iPads from the clinical setting were staphylococci (n=1104; 59.9%) and micrococci (n=469; 25.7%). The same types of bacteria were mainly found on iPads from the nonclinical setting (staphylococci: n=3678; 74.2% and micrococci: n=1051; 21.2%). However, the cultured microorganisms also included several pathogens. The most frequently identified pathogen was Staphylococcus aureus (non-MRSA only), which was found in nasal swabs from 2 medical informatics professionals as well as on their tablet PCs, but this species was also detected on tablet PCs from 2 other staff members who were not colonized themselves (Wilcoxon signed-rank test, z=-1.414, P=.157).

Determination of the Quality of the Guided Standardized Disinfection Process

The percentage of reduction of pathogens on iPads that underwent the standardized disinfection protocol is shown in Table 3 and Figures 3 and Dasari KB, White SM, Pateman J. Survey of iPhone usage among anaesthetists in England. Anaesthesia 2011 Jul;66(7):630-631. [CrossRef] [Medline]4. A significant overall reduction on microbes was achieved for both the clinical setting (98.1%; Wilcoxon signed-rank test, z=-3.1808, P=.001469) and the nonclinical setting (99.4%; Mann-Whitney U test, z=-3.1798; P=.001474). Note that bacilli are capable of forming spores. If doing so, those spores then show an extraordinarily high resistance towards disinfection processes (as they cannot at all be inactivated by alcohol-based disinfection) and other extreme environmental effects [Leggett MJ, McDonnell G, Denyer SP, Setlow P, Maillard JY. Bacterial spore structures and their protective role in biocide resistance. J Appl Microbiol 2012 Sep;113(3):485-498. [CrossRef] [Medline]25]. Still, a reduction of Bacillus spp. of 88% was achieved in our study in both settings. The reduction rates of all other bacterial and fungal species were as high as 99%.

Re-Sampling of the Two Untreated (Control) iPads

As mentioned above and shown in Figure 2, two tablet PCs were sampled once again without any additional reprocessing step in order to check for a potential germ-reducing effect due to the first sampling process itself. Re-sampling revealed reduction rates of 11.4% (228 of 246 CFU) on the nondisinfected iPad from the clinical setting and 22.4% (595 of 767 CFU) on the nondisinfected iPad from nonclinical setting only.

Determination of the Quality of Cleaning With a Soft, Lint-Free Cloth Without Liquid Cleaning Agents

As already noted (see Figure 2), one tablet PC from the nonclinical study was cleaned only with a brand-new fleece according to the instructions of the manufacturer. The initial CFU count of this device at arrival in the laboratory was 891 CFU; 427 CFU remained after cleaning with a fleece (reduction rate: 51.1%). Removal of bacteria was rather higher for the glass surface of the front (231 of 234 CFU; 98.7% reduction), but almost no reduction (5 of 77 CFU; reduction rate: 6.5%) was achieved on the plastic part of the device. Cleaning the aluminium resulted in a CFU reduction from 580 to 352 corresponding to 38.3%.

Table 1. Recovery of pathogens found on the devices’ surfaces on initial arrival at the laboratory (shown as cumulative number of CFU from 10 tablet PCs each). Comparison of the total number of microorganisms: Mann-Whitney U test, z=-3.402; P=.000670.


Clinical settingNonclinical setting


Total CFUMedian CFUIQRTotal CFUMedian CFUIQR
Total
1842162125.754969440273.75
Gram positive bacteria1825160.5122.754916437.5283

Front (glass)77258.562.251,672167104.25

Back (plastic)21422.5274814635.5

Back (aluminium)83963682763300.5183.25
Gram negative bacteria910.755224.5

Front (glass)611170.53.75

Back (plastic)000501

Back (aluminium)300.753001
Other
801.5100

Front (glass)601.5100

Back (plastic)100000

Back (aluminium)100000
Table 2. Distribution of species of microorganisms from the surface of the iPads on initial arrival of the devices at the laboratory (shown as cumulative number of colony forming units from 20 tablet PCs; n=6811).


CFU%Gram stain
Physiological human skin flora




Staphylococcus epidermidis178326.2positive

Micrococcus luteus150922.2positive

Staphylococcus hominis125618.4positive

Staphylococcus capitis97714.3positive

Staphylococcus warneri1942.9positive

Other coagulase-negative staphylococci3635.3positive

Bacillus spp.3094.5positive

Corynebacterium spp.1171.7positive

Other species200.3positive
Pathogenic microorganisms




Staphylococcus aureus (non-MRSAa)2183.2positive

Pseudomonas spp.360.5negative

Aspergillus spp. / molds90.1N/A

Acinetobacter spp.80.1negative

Other species120.2negative

aMRSA: methicillin resistant Staphylococcus aureus.

Figure 3. Histogram of CFU-count per localization samples taken from 6 corresponding devices in a clinical and nonclinical setting, stratified for position number, side, and material.
View this figure
Figure 4. Reduction of CFU in percent per position, side, and material after disinfection.
View this figure
Table 3. Reduction of bacteria on the surface of 6 iPads after standard disinfection procedure stratified by the type of previous usage (clinical vs nonclinical), the sample location (front vs back), the type of material (glass vs aluminium vs plastic), and type of Gram stain (positive vs negative).

On laboratory arrivalAfter standardized disinfection

Total CFUMedian CFUIQRTotal CFUMedian CFUIQRCFU reduction, %
Total (clinical setting)75312165.751421.598.1
Gram positive bacteria749121651421.597.2

Front (glass)28039.542.75510.7596.9

Back (plastic)1311832.7510098.5

Back (aluminium)3385137.2580.51.7596.5
Gram negative bacteria40.51000100.0

Front (glass)200.75000100.0

Back (plastic)000000100.0

Back (aluminium)200.75000100.0
Total (nonclinical setting)2751440811514.2599.4
Gram positive bacteria2739437.578.251515.2599.5

Front (glass)816148.5107.51012.7597.9

Back (plastic)31556.532000100.0

Back (aluminium)1608300.59050099.7
Gram negative bacteria121.53.25000100.0

Front (glass)80.52.5000100.0

Back (plastic)100000100.0

Back (aluminium)300.75000100.0

Principal Findings

Without any doubt, mobile devices provide numerous advantages in a hospital setting, but despite these benefits, the potential risk of pathogen transmission must be taken into account [Brady RR, Verran J, Damani NN, Gibb AP. Review of mobile communication devices as potential reservoirs of nosocomial pathogens. J Hosp Infect 2009 Apr;71(4):295-300. [CrossRef] [Medline]10]. There are several conclusions that can be drawn from the set-up and from the findings of our study.

As has been shown for other mobile devices [Borer A, Gilad J, Smolyakov R, Eskira S, Peled N, Porat N, et al. Cell phones and Acinetobacter transmission. Emerg Infect Dis 2005 Jul;11(7):1160-1161 [FREE Full text] [CrossRef] [Medline]8,Goldblatt JG, Krief I, Klonsky T, Haller D, Milloul V, Sixsmith DM, et al. Use of cellular telephones and transmission of pathogens by medical staff in New York and Israel. Infect Control Hosp Epidemiol 2007 Apr;28(4):500-503. [CrossRef] [Medline]9], an extensive surface contamination also takes place when iPads are being used. Every fingerprint on the surface (Figure 1) will leave residue on the glass, aluminum, and plastic parts of the device (Figure 1) and may contain a large number of bacteria. An increased awareness of this fact is required when those devices are used during patient care. Brady et al [Brady RR, Fraser SF, Dunlop MG, Paterson-Brown S, Gibb AP. Bacterial contamination of mobile communication devices in the operative environment. J Hosp Infect 2007 Aug;66(4):397-398. [CrossRef] [Medline]26] questioned 90 HCW (surgeons, anesthesiologists, and medical students) regarding this issue. At least 53% of them carried one mobile device (16% carried even more than one) including PDAs, mobile phones, and pagers. When asked about their cleaning habits, the HCW admitted that 80% of the PDAs, 85% of the mobile phones, and 96% of the pagers had never been cleaned by the owner.

Most of the pathogens are members of the resident or transient flora of its user (skin and/or anterior nose). Whatever microorganisms are present on the hands will be found on the mobile phone [Brady RR, Wasson A, Stirling I, McAllister C, Damani NN. Is your phone bugged? The incidence of bacteria known to cause nosocomial infection on healthcare workers' mobile phones. J Hosp Infect 2006 Jan;62(1):123-125. [CrossRef] [Medline]27,Jeske HC, Tiefenthaler W, Hohlrieder M, Hinterberger G, Benzer A. Bacterial contamination of anaesthetists' hands by personal mobile phone and fixed phone use in the operating theatre. Anaesthesia 2007 Sep;62(9):904-906. [CrossRef] [Medline]28] or the tablet PC later on. This stresses the need for proper hand hygiene of HCW as it has been addressed by the World Health Organization in the international “clean hands campaigns” recently [Kilpatrick C, Pittet D. WHO SAVE LIVES: Clean Your Hands global annual campaign. A call for action: 5 May 2011. Infection 2011 Apr;39(2):93-95. [CrossRef] [Medline]29]. Patients, too, should be educated about the role of their own mobile devices brought to the hospital because these devices will also become contaminated [Brady RR, Hunt AC, Visvanathan A, Rodrigues MA, Graham C, Rae C, et al. Mobile phone technology and hospitalized patients: a cross-sectional surveillance study of bacterial colonization, and patient opinions and behaviours. Clin Microbiol Infect 2011 Jun;17(6):830-835. [CrossRef] [Medline]30]. Especially patients who harbor multidrug resistant bacteria should be discouraged to share their mobile phone with others.

As shown by the repeated sampling by contact plates described above, microorganisms may easily spread from the surface of the tablet PC when touched again. A transmission of pathogens that have caused nosocomial outbreaks has been shown for mobile phones [Borer A, Gilad J, Smolyakov R, Eskira S, Peled N, Porat N, et al. Cell phones and Acinetobacter transmission. Emerg Infect Dis 2005 Jul;11(7):1160-1161 [FREE Full text] [CrossRef] [Medline]8,Akinyemi KO, Atapu AD, Adetona OO, Coker AO. The potential role of mobile phones in the spread of bacterial infections. J Infect Dev Ctries 2009;3(8):628-632 [FREE Full text] [Medline]12]. One would assume that a much larger device such as a tablet PC is even more likely to serve as a vehicle of infectious agents. HCW should therefore be encouraged to perform alcohol-based hand rubs after using their mobile device [Morioka I, Tabuchi Y, Takahashi Y, Oda Y, Nakai M, Yanase A, et al. [Bacterial contamination of mobile phones shared in hospital wards and the consciousness and behavior of nurses about biological cleanliness]. Nihon Eiseigaku Zasshi 2011 Jan;66(1):115-121 [FREE Full text] [Medline]31].

Cleaning with a fleece as recommended by the manufacturer of the tablet PC showed a reduction of about 50% of microorganisms. However, a sufficient reduction of the microbiological load will be achieved only when proper disinfection is performed. A cleaning phase for visible contamination and a disinfection phase as a final decontamination step are considered most effective according to infection control as recommended in the guideline for environmental cleaning in health care facilities from the Centers for Disease Control and Prevention and the Healthcare Infection Control Practices Advisory Committee [Sehulster L, Chinn RY, CDC, HICPAC. Guidelines for environmental infection control in health-care facilities. Recommendations of CDC and the Healthcare Infection Control Practices Advisory Committee (HICPAC). MMWR Recomm Rep 2003 Jun 6;52(RR-10):1-42 [FREE Full text] [Medline]3].

As has been shown for the disinfection of mobile phones [Sumritivanicha A, Chintanavilas K, Apisarnthanarak A. Prevalence and type of microorganisms isolated from house staff's mobile phones before and after alcohol cleaning. Infect Control Hosp Epidemiol 2011 Jun;32(6):633-634. [CrossRef] [Medline]32], a disinfection procedure for iPads that makes use of isopropanol wipes is very effective in reducing and inactivating residual bacteria. However, one has to keep in mind that this procedure may cause a loss of the warranty for this product. It is noteworthy that Apple’s recommendations for the cleaning process of the iPad, available on the company’s website, have significantly changed in the past. The version from December 15, 2010, stated that “it is also safe to use isopropyl alcohol 70% or a similar product” for this purpose. However, in the meantime, this statement has been withdrawn. Instead, it is now specified that “liquid damage is not covered under the Apple product warranty or AppleCare Protection Plans” and specifically to “Avoid getting moisture in openings. Don’t use window cleaners, household cleaners, aerosol sprays, solvents, alcohol, ammonia, or abrasives to clean the iPad” [Apple Inc. How to clean Apple products. 2013.   URL: http://support.apple.com/kb/ht3226 [accessed 2013-02-08] [WebCite Cache]20].

Regular disinfection serves to maintain a significantly lower load of pathogens. Our study results imply that disinfection followed by the deBac-app has the ability to reduce microbiological flora in a quality manner. We would like to recommend using a standardized scheme for the disinfection process as described and controlled by the deBac application. Reprocessing of tablet PCs should be performed at least once a day, preferably at the beginning of the working shift. Additional courses of disinfection should also be carried out any time that visible contamination has occurred. Furthermore, we recommend disinfecting the device after using it in a patient room under isolation precautions (eg, if the patient harbors some type of multidrug resistant organism). The guided disinfection procedure ensures that all parts of the surface get thoroughly treated. Furthermore, all steps of reprocessing are documented and may be filed in infection control records.

Limitations

Regarding the study design, we were willing to accept the following limitations: in accordance with our in-house regulations, machinery used in a clinical environment has to be disinfected. Therefore, it was impossible to learn about the baseline colonization by installing a control group on the wards and allowing this control group to use the devices without any disinfection. A randomized controlled trial or a controlled design with carefully matched comparison groups using standard practices as compared to the deBac condition would allow verification of the assumption that, when using mobile devices such as iPads in a clinical environment, performing an app-based disinfection process is more effective in reducing microbiological flora than simply using regular hand hygiene. This will have to be addressed in forthcoming studies.

The nursing staff was not provided with additional, paper-bound cleaning instructions since we wanted users to refer to the information available on the devices. We also refrained from collecting any personal data from the devices since our in-house data protection policy had to be followed. Only the information available from the anonymous cleaning protocols acquired from within the app was used for the evaluation and process documentation. The entries found in the log files demonstrated daily usage, but it is unknown to the observers how often and in which way the machines were used when the deBac-app was not running. This, of course, may bias the amount of CFU that were found on the surfaces.

As for the sensitivity and specificity of the various microbiological tests conducted, all laboratory methods that were used during the course of the study, including all microbiological tests for identification of pathogens, have been certified according to national guidelines. However, the sensitivity and specificity of taking the samples remains unclear as there is no so-called “gold standard” to compare with. It is known from optimized protocols that recovery may come up with a sensitivity of 98% and a specificity for a particular pathogen of interest of 95% [Hodges LR, Rose LJ, O'Connell H, Arduino MJ. National validation study of a swab protocol for the recovery of Bacillus anthracis spores from surfaces. J Microbiol Methods 2010 May;81(2):141-146. [CrossRef] [Medline]33]. However, standard environmental sampling still remains an unsolved problem these days [Moore G, Griffith C. Problems associated with traditional hygiene swabbing: the need for in-house standardization. J Appl Microbiol 2007 Oct;103(4):1090-1103. [CrossRef] [Medline]34,Galvin S, Dolan A, Cahill O, Daniels S, Humphreys H. Microbial monitoring of the hospital environment: why and how? J Hosp Infect 2012 Nov;82(3):143-151. [CrossRef] [Medline]35]. Unfortunately, it is impossible compare the Staphylococcus aureus strains cultured from the nasal swabs of the users with those found on the devices themselves since the isolates had been discarded in the meantime. However, to us, it seems highly probable that we found corresponding strains here, as Staphylococcus aureus does not represent a typical “environmental” bacterium. Furthermore, it is well known that people tend to frequently touch their noses.

Conclusions

Cleaning the devices with disinfecting wipes can be considered efficient and effective. Nevertheless, one must be aware of the potential danger of damaging the devices: there will definitely be a breach in warranty if liquid seeps into the device in any way and causes damage. On the other hand, although tablet PCs were originally developed for the consumer market, once they are used in the medical field, standardized methods for their disinfection must be implemented and closely followed. Also, manufacturers should become aware of the needs of the medical community regarding such devices. Thus, they might avoid building devices that—while being alluring for the medical sector—do not respect the demands for hygiene required for medicinal products. However, the most efficient personal action one can take to avoid transmission of bacteria, viruses, and other pathogens remains the proper disinfection of the hands before and after every patient interaction—this is a fact independent of the kind of device or any operating system or stated purpose.

Future studies should also take the specific profession of the staff as well as their level in the hierarchy into account. Their attitude towards using the deBac-app–based procedure compared to regular hand hygiene using alcohol-based disinfection solutions should also be evaluated. Also, the expenditure of time for implementing the procedure and other cost implicating variables will need to be addressed as the gathered results would be important factors for decision makers.

Acknowledgments

The authors would like to thank Gisela Rademacher and the technical staff of the Institute for Medical Microbiology and Hospital Epidemiology at Hannover Medical School as well as Prof Dr Herbert Matthies (PLRI) for their effort, expertise, and contribution while conducting this study. We would also like to thank Dr. Marianne Behrends (PLRI) as well as Regina Schmeer, MSc (Nursing Department) for giving us the chance to evaluate the iPads that were deployed as mentioned in [Albrecht U, Behrends M, Schmeer R, Matthies HK, von Jan U. Usage of Multilingual Mobile Translation Applications in Clinical Settings. JMIR Mhealth and Uhealth 2013 Apr 23;1(1):e4. [CrossRef]17] with respect to hygiene. The study was financed solely based on institutional funds of the Hannover Medical School. We acknowledge support by Deutsche Forschungsgemeinschaft for the publication costs. The paper was screened against the Transparent Reporting of Evaluations with Nonrandomized Designs (TREND) checklist [Des Jarlais DC, Lyles C, Crepaz N, TREND Group. Improving the reporting quality of nonrandomized evaluations of behavioral and public health interventions: the TREND statement. Am J Public Health 2004 Mar;94(3):361-366. [Medline]36] to assess methodological weaknesses.

Conflicts of Interest

None declared.

Multimedia Appendix 1

Standardized disinfection process of an iPad as guided by the corresponding application (“deBac-app”), the app documents the cleaning attempts of the device’s frame, front and back side.

PDF File (Adobe PDF File), 132KB

  1. Rosenthal VD, Bijie H, Maki DG, Mehta Y, Apisarnthanarak A, Medeiros EA, INICC members. International Nosocomial Infection Control Consortium (INICC) report, data summary of 36 countries, for 2004-2009. Am J Infect Control 2012 Jun;40(5):396-407. [CrossRef] [Medline]
  2. Gagliotti C, Balode A, Baquero F, Degener J, Grundmann H, Gür D, EARS-Net Participants (Disease Specific Contact Points for AMR). Escherichia coli and Staphylococcus aureus: bad news and good news from the European Antimicrobial Resistance Surveillance Network (EARS-Net, formerly EARSS), 2002 to 2009. Euro Surveill 2011;16(11):pii/19834 [FREE Full text] [Medline]
  3. Sehulster L, Chinn RY, CDC, HICPAC. Guidelines for environmental infection control in health-care facilities. Recommendations of CDC and the Healthcare Infection Control Practices Advisory Committee (HICPAC). MMWR Recomm Rep 2003 Jun 6;52(RR-10):1-42 [FREE Full text] [Medline]
  4. Dasari KB, White SM, Pateman J. Survey of iPhone usage among anaesthetists in England. Anaesthesia 2011 Jul;66(7):630-631. [CrossRef] [Medline]
  5. Payne KB, Wharrad H, Watts K. Smartphone and medical related App use among medical students and junior doctors in the United Kingdom (UK): a regional survey. BMC Med Inform Decis Mak 2012;12:121 [FREE Full text] [CrossRef] [Medline]
  6. Franko OI, Tirrell TF. Smartphone app use among medical providers in ACGME training programs. J Med Syst 2012 Oct;36(5):3135-3139. [CrossRef] [Medline]
  7. Gould D. Commentary: Ulger F et al. (2009). Are we aware how contaminated our mobile phones with nosocomial pathogens? Nurs Crit Care 2009;14(4):213-214. [CrossRef] [Medline]
  8. Borer A, Gilad J, Smolyakov R, Eskira S, Peled N, Porat N, et al. Cell phones and Acinetobacter transmission. Emerg Infect Dis 2005 Jul;11(7):1160-1161 [FREE Full text] [CrossRef] [Medline]
  9. Goldblatt JG, Krief I, Klonsky T, Haller D, Milloul V, Sixsmith DM, et al. Use of cellular telephones and transmission of pathogens by medical staff in New York and Israel. Infect Control Hosp Epidemiol 2007 Apr;28(4):500-503. [CrossRef] [Medline]
  10. Brady RR, Verran J, Damani NN, Gibb AP. Review of mobile communication devices as potential reservoirs of nosocomial pathogens. J Hosp Infect 2009 Apr;71(4):295-300. [CrossRef] [Medline]
  11. Ramesh J, Carter AO, Campbell MH, Gibbons N, Powlett C, Moseley H, et al. Use of mobile phones by medical staff at Queen Elizabeth Hospital, Barbados: evidence for both benefit and harm. J Hosp Infect 2008 Oct;70(2):160-165. [CrossRef] [Medline]
  12. Akinyemi KO, Atapu AD, Adetona OO, Coker AO. The potential role of mobile phones in the spread of bacterial infections. J Infect Dev Ctries 2009;3(8):628-632 [FREE Full text] [Medline]
  13. Datta P, Rani H, Chander J, Gupta V. Bacterial contamination of mobile phones of health care workers. Indian J Med Microbiol 2009;27(3):279-281 [FREE Full text] [CrossRef] [Medline]
  14. Ulger F, Esen S, Dilek A, Yanik K, Gunaydin M, Leblebicioglu H. Are we aware how contaminated our mobile phones with nosocomial pathogens? Ann Clin Microbiol Antimicrob 2009;8:7 [FREE Full text] [CrossRef] [Medline]
  15. Sadat-Ali M, Al-Omran AK, Azam Q, Bukari H, Al-Zahrani AJ, Al-Turki RA, et al. Bacterial flora on cell phones of health care providers in a teaching institution. Am J Infect Control 2010 Jun;38(5):404-405. [CrossRef] [Medline]
  16. Kramer A, Schwebke I, Kampf G. How long do nosocomial pathogens persist on inanimate surfaces? A systematic review. BMC Infect Dis 2006;6:130 [FREE Full text] [CrossRef] [Medline]
  17. Albrecht U, Behrends M, Schmeer R, Matthies HK, von Jan U. Usage of Multilingual Mobile Translation Applications in Clinical Settings. JMIR Mhealth and Uhealth 2013 Apr 23;1(1):e4. [CrossRef]
  18. PLRI MedAppLab. deBac-app. 2013.   URL: https://itunes.apple.com/us/app/debac-app/id453976138?mt=8&ign-mpt=uo%3D2 [accessed 2013-02-08] [WebCite Cache]
  19. Gebert A, Preiss G. A simple method for the acquisition of high-quality digital images from analog scanning electron microscopes. J Microsc 1998 Sep;191(3):297-302. [Medline]
  20. Apple Inc. How to clean Apple products. 2013.   URL: http://support.apple.com/kb/ht3226 [accessed 2013-02-08] [WebCite Cache]
  21. Stenehjem E, Rimland D. MRSA nasal colonization burden and risk of MRSA infection. Am J Infect Control 2013 May;41(5):405-410. [CrossRef] [Medline]
  22. Panousi MN, Williams GJ, Girdlestone S, Hiom SJ, Maillard JY. Evaluation of alcohol wipes used during aseptic manufacturing. Lett Appl Microbiol 2009 May;48(5):648-651. [CrossRef] [Medline]
  23. Noether GE. Sample Size Determination for Some Common Nonparametric Tests. Journal of the American Statistical Association 1987 Jun;82(398):645 [FREE Full text]
  24. Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull 1979 Mar;86(2):420-428. [CrossRef] [Medline]
  25. Leggett MJ, McDonnell G, Denyer SP, Setlow P, Maillard JY. Bacterial spore structures and their protective role in biocide resistance. J Appl Microbiol 2012 Sep;113(3):485-498. [CrossRef] [Medline]
  26. Brady RR, Fraser SF, Dunlop MG, Paterson-Brown S, Gibb AP. Bacterial contamination of mobile communication devices in the operative environment. J Hosp Infect 2007 Aug;66(4):397-398. [CrossRef] [Medline]
  27. Brady RR, Wasson A, Stirling I, McAllister C, Damani NN. Is your phone bugged? The incidence of bacteria known to cause nosocomial infection on healthcare workers' mobile phones. J Hosp Infect 2006 Jan;62(1):123-125. [CrossRef] [Medline]
  28. Jeske HC, Tiefenthaler W, Hohlrieder M, Hinterberger G, Benzer A. Bacterial contamination of anaesthetists' hands by personal mobile phone and fixed phone use in the operating theatre. Anaesthesia 2007 Sep;62(9):904-906. [CrossRef] [Medline]
  29. Kilpatrick C, Pittet D. WHO SAVE LIVES: Clean Your Hands global annual campaign. A call for action: 5 May 2011. Infection 2011 Apr;39(2):93-95. [CrossRef] [Medline]
  30. Brady RR, Hunt AC, Visvanathan A, Rodrigues MA, Graham C, Rae C, et al. Mobile phone technology and hospitalized patients: a cross-sectional surveillance study of bacterial colonization, and patient opinions and behaviours. Clin Microbiol Infect 2011 Jun;17(6):830-835. [CrossRef] [Medline]
  31. Morioka I, Tabuchi Y, Takahashi Y, Oda Y, Nakai M, Yanase A, et al. [Bacterial contamination of mobile phones shared in hospital wards and the consciousness and behavior of nurses about biological cleanliness]. Nihon Eiseigaku Zasshi 2011 Jan;66(1):115-121 [FREE Full text] [Medline]
  32. Sumritivanicha A, Chintanavilas K, Apisarnthanarak A. Prevalence and type of microorganisms isolated from house staff's mobile phones before and after alcohol cleaning. Infect Control Hosp Epidemiol 2011 Jun;32(6):633-634. [CrossRef] [Medline]
  33. Hodges LR, Rose LJ, O'Connell H, Arduino MJ. National validation study of a swab protocol for the recovery of Bacillus anthracis spores from surfaces. J Microbiol Methods 2010 May;81(2):141-146. [CrossRef] [Medline]
  34. Moore G, Griffith C. Problems associated with traditional hygiene swabbing: the need for in-house standardization. J Appl Microbiol 2007 Oct;103(4):1090-1103. [CrossRef] [Medline]
  35. Galvin S, Dolan A, Cahill O, Daniels S, Humphreys H. Microbial monitoring of the hospital environment: why and how? J Hosp Infect 2012 Nov;82(3):143-151. [CrossRef] [Medline]
  36. Des Jarlais DC, Lyles C, Crepaz N, TREND Group. Improving the reporting quality of nonrandomized evaluations of behavioral and public health interventions: the TREND statement. Am J Public Health 2004 Mar;94(3):361-366. [Medline]


CFU: colony forming units
HCW: health care workers
IQR: interquartile range
MRSA: methicillin resistant Staphylococcus aureus
PDA: personal digital assistant
PLRI: Peter L Reichertz Institute for Medical Informatics


Edited by G Eysenbach; submitted 31.03.13; peer-reviewed by A Friedrich, B Baskerville; comments to author 25.04.13; revised version received 26.05.13; accepted 13.06.13; published 14.08.13

Copyright

©Urs-Vito Albrecht, Ute von Jan, Ludwig Sedlacek, Stephanie Groos, Sebastian Suerbaum, Ralf-Peter Vonberg. Originally published in the Journal of Medical Internet Research (http://www.jmir.org), 14.08.2013.

This is an open-access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in the Journal of Medical Internet Research, is properly cited. The complete bibliographic information, a link to the original publication on http://www.jmir.org/, as well as this copyright and license information must be included.


Cookie Consent

We use our own cookies and third-party cookies so that we can show you this website and better understand how you use it, with a view to improving the services we offer. If you continue browsing, we consider that you have accepted the cookies.