Wearable Sensors in Gait Assessment for Parkinson Disease and Stroke in Real-World Environments: Systematic Review and Meta-Analysis

Wearable Sensors in Gait Assessment for Parkinson Disease and Stroke in Real-World Environments: Systematic Review and Meta-Analysis

1Data Analytics & Rehabilitation Technology (DART) at Lake Lucerne Institute, Rubistrasse 9, Vitznau, Switzerland

2Faculty of Medicine, University of Zurich, Zurich, Switzerland

3Department of Electrical, Electronic and Information Engineering "Guglielmo Marconi", University of Bologna, Bologna, Italy

4Rehabilitation Engineering Laboratory, ETH Zurich, Zurich, Switzerland

5Department of Neuology, University of Zurich and University Hospital Zurich, Zurich, Switzerland

6cereneo Center for Neurology and Rehabilitation, Vitznau, Switzerland

7Lake Lucerne Institute, Vitznau, Switzerland

8Health Sciences and Technologies, Interdepartmental Center for Industrial Research (CIRI-SDV), University of Bologna, Bologna, Italy

*these authors contributed equally

Corresponding Author:

Chris Easthope Awai, PhD


Background: Real-world gait assessment has gained momentum in populations with walking impairments, offering insights beyond standardized tests and supporting the integration of remote monitoring into clinical care. However, its potential remains limited by the lack of validated population- and context-specific digital biomarkers.

Objective: The primary objective was to summarize and critically evaluate the current state of real-world gait assessment using wearable sensors in Parkinson disease (PD) and stroke, addressing methodological approaches, sensor configurations, validation strategies, and outcome reporting. The secondary objective was to quantify methodological heterogeneity by reporting pooled means with 95% confidence intervals (CIs) and prediction intervals (PIs) of commonly reported gait parameters.

Methods: PubMed, Scopus, Web of Science, and Embase were searched for English-language studies published up to April 20, 2026. Eligible studies used wearable sensors to assess gait quality parameters in real-world settings in at least 5 individuals with PD or poststroke. Laboratory- or rehabilitation-only studies, non–peer-reviewed articles, abstracts, and studies before 2014 were excluded. Methodological quality was assessed with a checklist adapted from Hubble et al. Gait features available in at least 5 reports were included in a meta-analysis using a random-effects model with Hartung-Knapp-Sidik-Jonkman adjustment, deriving 95% CIs and PIs to characterize the pooled estimate and its dispersion.

Results: Of 2489 records, 43 reports were included: 37 on PD (2774 participants; mean age 67.89, SD 8.68 years) and 6 on stroke (217 participants; mean age 63.83, SD 11.84 years). Methodological quality was high, but external validity was consistently the weakest domain in both populations. Sensor configurations most commonly consisted of a lower back accelerometer. In total, 13 distinct walking bout definitions were identified, whereas 17 reports provided none, indicating substantial terminological heterogeneity. In PD, 4 gait parameters met the meta-analysis threshold, each showing a PI far wider than its CI, precluding generalization to future settings. For example, gait speed pooled at 0.84 m/s (95% CI 0.77‐0.91; 95% PI 0.56‐1.12), and number of steps per day at 7202 (95% CI 3975‐10,429), the latter with a PI extending below zero, a mathematically impossible range reflecting extreme dispersion. No stroke parameter was reported consistently enough to permit meta-analysis, underscoring a critically underdeveloped evidence base.

Conclusions: Unlike previous reviews focusing on single diseases or specific hardware, this review contrasts PD and stroke to highlight shared methodological challenges. Real-world gait assessment in PD is advancing but fragmented, whereas the stroke evidence base remains sparse. The observed variance is compounded by procedural and algorithmic inconsistencies that cannot currently be separated from genuine clinical differences. The methodological groundwork established in PD offers a foundation for early adoption of standardized terminology, validation procedures, and core outcomes, essential for transitioning wearable sensors from exploratory tools to reliable clinical and remote-monitoring instruments.

Trial Registration: PROSPERO CRD42024531665; https://www.crd.york.ac.uk/PROSPERO/view/CRD42024531665

J Med Internet Res 2026;28:e93413

doi:10.2196/93413

Keywords



Neurological disorders are the leading cause of disability-adjusted life-years and a major contributor to global disability, requiring long-term access to care and treatment, monitoring, and rehabilitation to improve patient outcomes and quality of life [1,2]. Among these conditions, Parkinson disease (PD) is the fastest-growing neurodegenerative disorder, while stroke remains the leading cause of long-term disability worldwide [3]. Despite differing pathophysiology, both conditions share a clinically important consequence, namely, characteristic gait disturbances that substantially impair mobility in daily life and rank among the most prevalent contributors to reduced quality of life and functional independence [4-6]. In PD, gait is typically marked by freezing episodes, reduced gait speed, shorter step length, prolonged double-support time, and increased stride-to-stride variability [7,8]. Poststroke gait, in contrast, is characterized by persistent asymmetry, reduced gait speed, and altered step patterns that may limit recovery and community reintegration [9]. Beyond these phenotypic differences, both populations share a fundamental clinical challenge: the reliable and meaningful assessment of gait in everyday life. In PD and stroke, gait impairments are chronic, multifactorial, and subject to temporal fluctuations, shaped by a complex interplay of individual, task-related, and environmental factors that isolated point-in-time clinical assessments cannot fully capture [10-13].

This challenge is formally recognized in the World Health Organization’s International Classification of Functioning, Disability and Health, which differentiates between 2 constructs central to understanding functional ability [14]. Capacity refers to an individual’s maximal functioning in a standardized or controlled setting, whereas performance reflects how that individual actually moves within their natural, everyday environment [15]. Assessing both constructs provides valuable and complementary clinical insight; yet, current practice focuses predominantly on capacity, leaving performance systematically underassessed [10]. Wearable sensor technologies address this gap directly by enabling continuous, unobtrusive monitoring under real-world conditions, shifting the focus from what patients can do in the clinic to what they actually do in daily life, thereby enhancing the ecological validity of gait assessment and supporting clinical decision-making [16-18].

Early wearable-based approaches relied primarily on pedometers and activity trackers, capturing step counts, walking distance, and activity bouts [19]. While informative about overall physical activity levels, these measures cannot characterize gait quality. Features such as spatiotemporal variability, asymmetry, coordination, and disease-specific motor symptoms may be more sensitive to motor dysfunction than traditional clinical rating scales; yet, they remain beyond the reach of simple activity monitors [20,21]. This is reflected in a systematic review of 30 studies on walking performance after stroke, in which the most commonly reported outcomes were step counts, step timing, walking distance, and walking bouts [22]. Inertial measurement units (IMUs) represent a meaningful advance in this regard, enabling the extraction of richer, more granular gait outcomes under real-world conditions. Studies have shown that IMU-derived digital mobility outcomes can distinguish people with PD from healthy controls and differentiate real-world from supervised walking, underscoring their clinical potential [23]. Recent advances include algorithms tailored to patient-specific characteristics and validation approaches that extend beyond simple laboratory tasks [24,25]. However, important challenges remain. Accurately assessing gait in severely impaired individuals is particularly difficult in the presence of slow gait speed, assistive device use, frequent breaks, and short walking bouts [24,26-28]. Broader methodological barriers, including inconsistent data collection protocols, lack of standardization, limited algorithm validation, and unclear clinical relevance of many sensor-derived outcomes, further limit progress in the field.

Existing reviews have only partially addressed these challenges. In PD, reviews have largely focused on wearable technologies for motor symptom detection and management, the identification of digital biomarkers for remote monitoring, or specific sensor configurations and data-processing approaches [23,29-33]. In stroke, recent reviews have centered predominantly on physical activity monitoring or rehabilitation-oriented assessment frameworks [34-37]. To our knowledge, no review has yet systematically examined wearable-based real-world gait assessment across both populations, despite their shared assessment challenges and the growing clinical need for validated, ecologically meaningful outcomes. As the field develops rapidly, a timely synthesis is needed to clarify the current state of the evidence and highlight where methodological and terminological heterogeneity remain pronounced. This systematic review and meta-analysis therefore evaluates the current state of wearable sensor-based real-world gait assessment in PD and stroke, examining sensor setups, IMU-derived outcomes, validation strategies, and methodological approaches, while identifying the key barriers and areas where greater consensus will be essential for future progress. Where gait features were available in at least 5 reports, a meta-analysis was conducted to quantitatively synthesize the available evidence and characterize the range of values reported across the literature.


Protocol and Registration

This systematic review and meta-analysis was registered with PROSPERO (International Prospective Register of Systematic Reviews: CRD42024531665). It was conducted in accordance with the PRISMA (Preferred Reporting Items for Systematic Reviews and Meta-Analyses) 2020 guidelines, including the PRISMA 2020 for Abstracts checklist [38], the PRISMA-S (Preferred Reporting Items for Systematic Reviews and Meta-Analyses Literature Search Extension) 2021 guidelines [39], and the framework outlined by Khan et al [40]. All checklists are provided in Checklists 1-3.

Search Strategy and Selection Criteria

A systematic search in PubMed, Scopus, Web of Science, and Embase was performed for English-language studies published up to April 20, 2026. The reference lists of included studies and previous systematic reviews were screened for additional potentially eligible studies. No supplementary search methods, such as gray literature sources or expert contacts, were used. The following terms were searched as keywords: “Parkinson’s disease,” “stroke,” “gait,” “accelerometry,” “wearable sensor,” and “real-world.” The comprehensive search string for each database is provided in Multimedia Appendix 1. Two independent reviewers (SN and JA-S) screened all records at each stage of the selection process, including title, abstract, and full-text assessment. Disagreements were resolved by consensus or, when needed, by a third reviewer (ACN). Authors were contacted for clarification of study details when necessary.

Eligibility criteria were based on the population, intervention, comparison, outcomes, and study design framework [41]. Eligible studies enrolled at least 5 individuals with PD or stroke and used body-worn sensors to assess and report gait quality parameters in real-world settings. Studies were excluded if gait assessments were conducted solely in laboratory or rehabilitation settings, or if they were systematic reviews, non–peer-reviewed articles, or abstracts. The search was restricted to studies published from 2014 onward, as advances in sensor miniaturization, battery life, and data transmission around this period substantially increased the feasibility of prolonged gait monitoring in daily-life settings. Articles were also considered ineligible if they did not report extractable real-world gait outcomes of interest. Studies presenting only classification results, prediction scores, or model performance metrics without corresponding gait parameters were therefore excluded.

Data Extraction and Analysis

Screening and data extraction were performed using Covidence (Veritas Health Innovation) templates to ensure consistency. One reviewer (SN) extracted the following information: publication details (title, year, first author, and country), study characteristics (sample size and clinical concept of interest), participant characteristics (age, sex, and diagnosis), data and algorithm origin and accessibility (study location and availability of data and algorithms), study setting (real-world or combined laboratory and real-world), methods (monitoring duration, sensor type and placement, and algorithm validation), gait-related outcome parameters (macro and micro), and methodological details (study aim and walking bout definition). All extracted data were checked by a second reviewer (JA-S), and discrepancies were resolved by consensus or, when necessary, by a third reviewer (ACN). Missing or unclear information was marked as “not available” if it was not directly stated in the report or could not be clearly traced or verified from the cited references.

Gait parameters were categorized into macro and micro characteristics [42]. Macro gait characteristics describe higher-level daily ambulatory patterns, including volume and variability measures such as total walking time per day, number and duration of walking bouts, and total step count, as well as advanced nonlinear metrics such as the power-law distribution (α) and within-bout variability (S2) [25,43,44]. Micro gait characteristics focus on spatiotemporal parameters, including their variability and asymmetry, within individual walking bouts. Studies meeting inclusion criteria were synthesized narratively, and a meta-analysis of data collected in real-world settings was performed when at least 5 eligible reports provided comparable gait parameters from distinct populations to quantitatively characterize the range of values reported across the literature.

Methodological Quality

Methodological quality was assessed independently by 2 reviewers (SN and JA-S) using a checklist adapted from Hubble et al [45], which evaluates studies across 4 domains: reporting, external validity, internal validity, and confounding and selection bias. As the original checklist was designed for interventional and cohort study designs, it was adapted to fit the observational nature of the reports included in this review. This resulted in a modified 18-item version in which 8 items were reworded or added to better align the tool with this review’s focus on real-world monitoring and algorithm reporting (Multimedia Appendix 2). Specifically, in the reporting domain, 5 of the 10 original items were retained, and 5 were reworded. In the external validity domain, 2 of the 3 original items were retained, and 2 were added. In the internal validity domain, 3 of the 7 original items were retained, and 1 was added. The original 6-item confounding and selection bias domain was excluded entirely, as its content was specific to cohort study methodology and was not relevant to the included reports. The adapted checklist was piloted prior to full application. Disagreements in quality scoring were resolved by consensus or, when necessary, by a third reviewer (ACN). Quality scores were expressed as percentages and used to descriptively characterize the methodological quality of the included reports. Given the heterogeneity of study designs and the exploratory nature of the review, quality scores were not used to exclude reports or to weight the meta-analysis.

Meta-Analysis Methods

When reports provided data from both laboratory and real-world settings, only real-world data were extracted and included in the quantitative synthesis. To avoid double counting in the meta-analysis, each dataset was included only once per outcome. When multiple reports were based on the same dataset, the report with the largest number of participants was selected. When outcomes were reported separately for distinct subgroups, such as fallers and nonfallers, subgroup data were first combined and treated as a single study sample. When outcomes were reported at multiple time points, the time point with the largest sample size was used. Prior to quantitative pooling, strict data extraction rules were applied to address clinical and methodological heterogeneity to prevent statistical distortion.

Reports were excluded from the meta-analysis and reserved for qualitative synthesis only if they met either of the following criteria: (1) fragmented bout definitions, whereby parameters were presented only in disjointed subcategories without providing an overall aggregate mean for steady-state walking (eg, isolating 10‐ to 60-second bouts from >60-second bouts); and (2) incompatible temporal units, whereby macro gait parameters were reported in scaled time frames such as steps per hour, as extrapolating these to daily totals introduces artificial variance, given the nonuniform nature of daily ambulation. Prior to pooling, all extracted parameters were harmonized to standard units (eg, meters per second for gait speed, steps per minute for cadence, meters for spatial parameters, and seconds for temporal parameters). For each included report, the mean, SD, and variance of gait parameters were extracted or computed. When reports provided medians and IQRs, the SD was estimated as SD=(p75–p25)/1.349. This conversion was required in 3 reports for gait speed and 2 reports for cadence. While this formula assumes approximate normality, it allowed these datasets to be included in the quantitative synthesis. A random-effects model was selected a priori, as the true effect size was expected to vary across studies due to differences in sensor configurations, gait detection algorithms, and population characteristics. To provide a more conservative estimation of the uncertainty around the pooled averages, the Hartung-Knapp-Sidik-Jonkman adjustment was applied. Between-study variance (τ²) was estimated using the restricted maximum likelihood method. In addition to 95% confidence intervals (CIs), 95% prediction intervals (PIs) were calculated to quantify the expected range of true effects in future real-world assessments. Formal assessments of small-study effects (eg, funnel plot asymmetry) were not performed because of the small number of reports available per parameter. In addition, according to Cochrane guidelines, such tests are not recommended in the presence of extreme between-study heterogeneity, reflected here in PIs far wider than the corresponding CIs, as any observed asymmetry likely reflects clinical and methodological variation rather than true publication bias [46]. All analyses were conducted in R (version 4.5; R Foundation for Statistical Computing) using RStudio and the meta package (version 8.3).

Certainty of Evidence

The certainty of evidence was assessed for each meta-analyzed outcome using the GRADE (Grading of Recommendations Assessment, Development, and Evaluation) approach. As all evidence was observational, certainty started at low and was evaluated across 5 domains. Risk of bias was judged using a modified quality score (mean of the reporting and internal-validity domain scores from Multimedia Appendix 2); external validity was excluded to avoid double-counting with the indirectness domain. Reports scoring below 75% on this modified score were classified as having relevant concerns, and risk of bias was rated serious if such reports contributed at least half of the pooled weight for an outcome. Inconsistency was based on PI width relative to the CI [47]. Indirectness was based on participant representativeness and algorithm validation setting. Imprecision was based on the relative width of the 95% CI around the pooled mean, with a threshold of 30% for a serious rating. Funnel-plot asymmetry testing was not performed, as k<10 for 2 outcomes, and very wide PIs render asymmetry uninterpretable for the remaining 2. Assessments were performed independently by 2 reviewers (SN and JA-S), with disagreements resolved by consensus.


Study Selection and Characteristics

The search identified 2486 records, with 3 additional records identified through manual reference checks. After duplicate removal and title and abstract screening, 71 reports were selected for full-text review. Of these, 43 reports met the inclusion criteria, comprising 37 reports on PD (2774 cumulative participants; mean age 67.89, SD 8.68 years) [24,26-28,43,44,48-78] and 6 reports on stroke (217 cumulative participants; mean age 63.83, SD 11.84 years) [25,79-83]. Because several reports draw on shared parent cohorts (eg, ICICLE-GAIT [Incidence of Cognitive Impairment in Cohorts with Longitudinal Evaluation-GAIT], V-TIME [Virtual Reality-Treadmill Combined Intervention for Enhancing Mobility and Reducing Falls in the Elderly], Mobilise-D, and FallRiskPD), these totals represent the sum of report-level sample sizes and include some individuals more than once, rather than counts of unique participants. The PRISMA flow diagram is shown in Figure 1. The remaining 28 reports were excluded because of a lack of gait quality parameter or real-world data or because they were non–peer-reviewed articles. Not all included reports contributed data to the meta-analysis, as reported gait parameters varied widely. After excluding duplicate datasets, 21 unique datasets were available for the meta-analysis in PD. The evidence base differed substantially between the 2 populations, with 37 PD reports compared with 6 stroke reports, several of which relied on overlapping datasets. All stroke-related findings should therefore be interpreted with caution and considered preliminary throughout this review. Study characteristics are summarized in Table 1.

‎
Figure 1. The PRISMA (Preferred Reporting Items for Systematic Reviews and Meta-Analyses) flow diagram of the study selection process.
Table 1. Characteristics of the included reportsa.
ReportsSample, nClinical concept of interestGait
assessment
Monitoring
period
Digital outcomeAlgorithm validationSensors, n
Parkinson
Weiss et al (2014)b [70]107Fall riskLabc+RWd3 daysGait macro and microNot available1
Weiss et al (2015)b [71]107Cognitive functionLab+RW3 daysGait macro and microNot available1
Weiss et al (2015)b [73]72GaitLab+RW3 daysGait macro and microNot available1
Toosizadeh et al (2015) [72]15GaitLab+RW1 dayGait speedIndoor in prior study [84]1
Del Din et al (2016)e [67]47GaitLab+RW7 daysGait macro and microIndoor in prior study [85]1
Mancini et al (2018) [69]24Freezing of gaitRW7 daysGait macro and microNot available3
Del Din et al (2019)e,f [44]170Fall riskLab+RW7 daysGait macro and microIndoor in prior study [85]1
Galperin et al (2019)f [68]125Gait, physical activityLab+RW7 daysGait macro and microNot available1
Del Din et al (2020)f [43]128Fall riskLab+RW7 daysGait macroIndoor in prior study [85]1
Terashi et al (2020) [54]106Cognitive function, motor abnormalitiesRW1 dayGait microNot available1
Coates et al (2020)e [53]5GaitRW7 daysSample entropy of stride timeIndoor in prior study [85]1
Shah et al (2020) [55]16GaitLab+RW7 daysGait macro and microIndoor in prior study [86]3
Shah et al (2020)g [61]29GaitRW7 daysGait macro and microIndoor in prior study [86]3
Shah et al (2020)g [62]29GaitRW7 daysGait macro and microIndoor in prior study [86]3
Bouça-Machado et al (2021) [56]24GaitLab+RW3 daysGait macro and microIndoor in prior study [85]1
Ullrich et al (2021)h [57]12Gait testsRW14 daysGait microIndoor in prior study [87]2
Adams et al (2021) [49]17Gait, physical activity, tremorLab+RW2 daysGait macro and microNot available5
Atrsaei et al (2021)i [48]27GaitLab+RW1 dayGait speedIndoor in prior study [88]1
Atrsaei et al (2021) [74]26Fear of fallingLab+RW14 daysGait macro and gait speedIndoor in prior study [88]1
Corrà et al (2021)i [63]27Gait speedLab+RW1 dayGait speedIndoor in prior study [88]2
Rehman et al (2022)e [50]47Parkinson disease classificationLab+RW7 daysGait macro and microIndoor in prior study [85]1
Roth et al (2022)h [51]38Stairs, fall riskRW14 daysGait micro and stairsOutdoor in prior study [89]2
Ullrich et al (2023)h [58]35Fall riskRW14 daysGait microIndoor in prior study [87]2
Caballol et al (2023) [59]39Gait, freezing of gait, dyskinesiaRW7 daysGait macro and microNot available1
Cohen et al (2023) [60]46GaitLab+RW7 daysGait macro and microNot available1
Salis et al (2023)j [24]20GaitLab+RW2.5 hoursGait macro and microIndoor within study7
Micó-Amigo et al (2023)j [52]20GaitRW2.5 hoursGait macro and micro, gait eventsOutdoor within study1
Kirk et al (2023)e [28]62GaitRW7 daysGait speedIndoor in prior study [85]1
Romijnders et al (2023)j [26]18GaitRW2.5 hoursGait micro, gait eventsOutdoor within study2
Shah et al (2023) [75]34Fall riskLab+RW7 daysGait microIndoor in prior study [86]3
Mirelman et al (2024)f [64]587GaitRW7 daysGait macro and microIndoor in prior study [85]1
Kirk et al (2024)j [27]15Gait speedLab+RW2.5 hoursGait speedIndoor and outdoor within study1
Zampogna et al (2024) [65]48Freezing of gait, dyskinesiaRW6‐8 daysGait micro, freezing, dyskinesiaNot available1
Nishi et al (2024) [66]41Gait speedLab+RW7 daysGait microIndoor in prior study [85]1
Kirk et al (2026) [76]39ContextualizationLab+RW7 daysGait macro and microIndoor and outdoor in prior study [27,52]1
Sidoroff et al (2026) [77]41GaitLab+RW7 daysGait macro and microOutdoor in prior study [90]2
Yarnall et al (2026) [78]531GaitLab+RW7 daysGait macro and microIndoor and outdoor in prior study [27,52]1
Stroke
Sanchez et al (2015) [79]23Gait recoveryRW3 ×1 daysGait macro and microNot available3
Punt et al (2016)k [80]56Fall riskRW7 daysGait macro and microIndoor in prior study [91]1
Moore et al (2017) [25]25GaitLab+RW2×7 daysGait macro and microIndoor in prior study [85]1
Punt et al (2017)k [81]40Fall riskLab+RW7 daysGait macro and microIndoor within study1
van de Port et al (2020)k [82]38GaitLab+RW7 daysGait macro and microIndoor in prior study [91]1
Felius et al (2025) [83]35GaitLab+RW2 daysGait speed, number of stridesNot available1

aFor the meta-analysis, only data from the real-world setting were used.

bData originates from the same study population [92].

cLab: gait was assessed in a laboratory or clinical setting.

dRW: gait was assessed in a real-world setting.

eData originates from the ICICLE-GAIT (Incidence of Cognitive Impairment in Cohorts with Longitudinal Evaluation-GAIT) study (June 2009-December 2011).

fData originates from the V-TIME (Virtual Reality-Treadmill Combined Intervention for Enhancing Mobility and Reducing Falls in the Elderly) study (November 2012-April 2015).

gData originates from the same study population.

hData originates from the FallRiskPD study (March 2019-June 2021).

iData originates from the same study population.

jData originates from the Mobilise-D study (July 2020-September 2021).

kData originates from the same study population.

Research activity in this field has increased markedly, with more than 70% of included reports published in 2020 or later [24,26-28,43,48-66,74-78,82,83]. The United Kingdom contributed the largest number of reports (n=15), including 9 from multicenter collaborations [24,26,27,43,44,52,64,68,78]. Among the included reports, 18 PD reports [24,26-28,43,44,49,50,52,53,55,61,62,64,67,72,76,78] and 2 stroke reports [79,80] also included a comparison group of older adults. In total, 9 reports provided open-access datasets [24,26,27,49,51,52,57,58,78], 17 reported data availability upon request [25,28,48,50,54-56,59,60,62-65,74,76,77,83], and the remaining reports did not address data availability [43,44,53,61,66-73,75,79-82]. Algorithm transparency was limited overall. A total of 5 reports made their algorithms publicly available [26,52,64,76,78], and 1 offered access upon request [48].

Methodological Quality

Methodological quality was evaluated across the 3 retained domains: reporting, external validity, and internal validity. Ratings for individual reports are provided in Multimedia Appendix 2. Overall quality scores in PD ranged from 69.4% to 94.4%, with a mean of 83.5% (SD 5.8%). Overall quality scores in stroke ranged from 72.2% to 91.7%, with a mean of 81.9% (SD 7.2%). Of the 37 PD reports, 32 were rated as high quality (scores above 75%), and 5 as moderate (scores between 50.1% and 75%). Of the 6 stroke reports, 4 were rated as high quality, and 2 as moderate. Reporting quality was high in both PD (mean 94.1%, SD 5.9%) and stroke (mean 93.3%, SD 6.8%). The main reporting deficiencies across both populations concerned protocol descriptions, walking bout and step detection definitions, and the characterization of participants. In stroke, an additional reporting gap was identified in the presentation of statistical results, with several reports presenting significance thresholds rather than exact P values for their main outcomes. External validity scored lowest across both populations, with mean scores of 49.3% (SD 15.9%) in PD and 45.8% (SD 18.8%) in stroke. The primary issues were failure to confirm participant representativeness, absence of participation bias assessment, and lack of algorithm validation within the target population. Internal validity scores were substantially higher in both PD (mean 91.2%, SD 9.3%) and stroke (mean 89.6%, SD 5.1%), with the main concern being that some reports did not fully address the appropriateness of statistical methods. Across both populations, external validity represented the most consistently weak domain.

Methodological Protocols

The included reports addressed a broad range of research objectives, reflecting the heterogeneity of the field. In PD, study designs ranged from technical algorithm validation to clinical characterization and intervention evaluation, while stroke reports were more limited in number and scope. Of the 37 PD reports, 23 assessed gait in both laboratory and real-world settings [24,27,43,44,48-50,55,56,60,63,66-68,70-78], while 14 reports focused exclusively on real-world [26,28,51-54,57-59,61,62,64,65,69]. Disease severity was characterized using the Unified Parkinson’s Disease Rating Scale across all PD reports [24,26-28,43,44,48-78], providing a consistent clinical reference point. Four reports focused on technical validation, examining the accuracy of gait detection algorithms in laboratory or real-world contexts [24,26,27,52]. Construct validity was assessed in 20 reports through known-groups, convergent, and discriminant approaches. These included comparisons between people with PD and healthy controls, fallers and nonfallers, people with and without freezing of gait, and disease severity groups, as well as associations between home-based gait parameters, in-clinic assessments, clinical scales, and patient-reported outcomes [28,44,48,49,51,53-55,61-65,68-73,78].

Four reports examined methodological questions relevant to real-world gait assessment, including how environment, walking bout length, and data aggregation strategies influence gait parameters and disease classification or fall risk [50,58,66,67]. Additional methodological contributions included GPS-derived indoor-outdoor contextualization and the detection of unsupervised gait tests within continuous sensor data [57,76]. Beyond validation, wearable sensors were used to evaluate intervention effects and clinical determinants of mobility. Three reports assessed ambulatory outcomes in the context of rehabilitation, examining both efficacy and effectiveness [56,60] or effectiveness alone [43]. Two reports investigated treatment-related effects, including changes following treatment adjustments [59] and differences in gait and turning markers between off- and on-levodopa states across clinical and daily-life settings [75]. In addition, one report investigated the influence of fear of falling on mobility [74], and another compared mobility profiles between PD and atypical parkinsonian syndromes [77].

The 6 stroke reports were more limited in methodological scope. A total of 4 assessed gait in both laboratory and real-world settings [25,81-83], while 2 focused exclusively on real-world settings [79,80]. One report evaluated the feasibility, validity, and test-retest reliability of a single lower back accelerometer for measuring gait after stroke in both settings [25]. Construct validity was supported through comparisons between fall-prone and nonfall-prone individuals with stroke and associations between real-world gait parameters and clinical measures including balance, gait speed, and fear of falling [79,80,82]. Predictive validity was examined in 2 reports investigating whether laboratory-based and daily-life gait characteristics could predict prospective falls or community walking outcomes [81,83]. Notably, none of the stroke reports used the National Institutes of Health Stroke Scale (NIHSS) to characterize stroke severity [25,79-83], limiting the ability to assess the representativeness of the study populations and the generalizability of findings across severity levels.

Sensor Configuration

Sensor configurations varied considerably across reports in terms of sensor type, number, and placement. In PD, all reports used accelerometers [24,26-28,43,44,48-78], with gyroscopes incorporated in 17 reports [24,26,48,51,52,55,57,58,61-63,69,74-78] and magnetometers in 7 [24,55,61,62,69,74,75]. The number of sensors per report ranged from 1 to 7, with a mean of 1.70 (SD 1.29). In stroke, all reports also used accelerometers, with configurations ranging from 1 to 3 sensors per report [25,79-83]. In PD, the most common configuration was a single sensor positioned on the lower back, used in 20 reports [27,28,43,44,50,52,53,56,60,64-68,70,71,73,74,76,78], secured with tape [28,43,44,53,60,64,67,68,76,78], a belt [27,52,65,70,71,73,74,78], both tape and a belt [78], or by a method that was not reported [50,56,66]. Less frequently, reports paired a lower back sensor with additional sensors on the feet [69] or ankles [55,61,62,75]. Other reports used sensors placed exclusively on one [48] or both feet [26,51,57,58,63,77], while more complex multisensor setups included the inertial module with distance sensors and pressure insoles system, comprising 3 IMUs on the lower back and both feet, distance sensors, and pressure insoles [24]. Additional configurations included sensors placed at the chest [54,72], the left hip [59], and a 5-sensor setup distributed across the thighs, forearms, and trunk [49]. In stroke, 4 reports used a single lower back sensor [25,80-82] secured with a belt [80-82] or tape [25]. One report used a single sensor attached to the right foot [83], and one report used 3 sensors placed on the chest and both thighs [79]. In PD, the most frequently used device was the AX3 triaxial accelerometer (Axivity), used in 12 reports [28,43,44,50,53,56,60,64,66-68,76], followed by the McRoberts DynaPort MM+ in 6 reports [27,52,70,71,73,78], the Opal sensor (APDM) in 5 reports [55,61,62,69,75], and the Mobile-GaitLab system (Portabiles Healthcare Technologies GmbH) in 4 reports [51,57,58,77]. In PD, 20 reports used a sampling frequency of 100 Hz [24,26-28,43,44,50,52-54,56,60,64,66-68,70,71,76,78]. In stroke, the most frequently used device was the McRoberts DynaPort MM+ in 3 reports [80-82]. Notably, all 3 of these reports were based on the same dataset, which substantially limits the independence of the stroke evidence base for device-specific observations. In total, 4 stroke reports used a sampling frequency of 100 Hz [25,80-82], 1 of 104 Hz [83], and 1 of 128 Hz [79].

Technical Validation

Technical validation for real-world gait algorithms was reported in 27 of 37 PD reports [24,26-28,43,44,48,50-53,55-58,61-64,66,67,72,74-78]. For the remaining 10 reports, the technical validation status was unclear, as no comparison against an external reference system was reported or could be clearly traced from the cited methodology [49,54,59,60,65,68-71,73]. Of the 27 validated reports, 3 had been validated exclusively in healthy controls [57,58,72], a limitation when interpreting their applicability to clinical populations. Among the validated reports, 20 validations were conducted indoors [24,28,43,44,48,50,53,55-58,61-64,66,67,72,74,75], 4 outdoors [26,51,52,77], and 3 combined both settings [27,76,78]. Notably, 9 indoor reports [28,43,44,50,53,56,64,66,67] based their algorithms on the work of Del Din et al [85], using GAITRite as the reference system, which was the most commonly used reference overall [28,43,44,50,53,55-58,61,62,64,66,67,75]. Other validation approaches included stereophotogrammetry, the inertial module with distance sensors and pressure insoles system, force-sensing insoles, or video analysis, reflecting considerable variability in reference systems across the literature [24,26,27,48,51,52,63,72,74,76-78]. In total, 4 of the 6 stroke reports described algorithm validation [25,80-82], all conducted indoors using video analysis [80-82] or a combination of GAITRite and Opal sensors [25]. For the remaining 2 reports, the technical validation status was unclear [79,83].

Observation Windows and IMU-Derived Outcomes

A 7-day (range 6‐8 days in 1 report) observation window was the most common monitoring duration for real-world gait assessment across both PD and stroke reports [25,28,43,44,50,53,55,59-62,64-69,75-78,80-82]. In PD, monitoring durations ranged from 2.5 hours to 14 days, though the shortest duration was used exclusively for algorithm validation rather than real-world gait assessment [24,26,27,52]. In stroke, monitoring durations ranged from 1 to 7 days [25,79-83]. Continuous walking sequences, referred to as walking bouts, were extracted from sensor recordings across the included reports. However, the criteria used to define a walking bout varied considerably. Overall, 13 distinct definitions were identified across the included reports (Table 2), with 10 in PD, 2 in stroke, and 1 in both cohorts. Some reports applied minimum step or stride count thresholds, others used minimum duration criteria, and some incorporated rest period rules to determine whether consecutive walking sequences should be merged or treated separately. Notably, 17 reports did not provide a walking bout definition [48,49,54,56,57,59,60,64,65,68,70-74,79,83].

Table 2. Walking bout definitions applied in included reports, presented separately for Parkinson disease and strokea.
References
Parkinson disease reports
Minimum bout length of 3 steps with no threshold set for maximum resting period between consecutive walking bouts.[50,53,66,67]
Minimum bout length of 5 strides with a maximum resting period of 3.5 seconds.[51]
Minimum bout length of 3 steps with a maximum resting period of 2.5 seconds.[28,44]
Minimum bout length of greater than 10 seconds with a maximum resting period of 2.5 seconds.[43]
Minimum bout length of 3 steps and 3 seconds with a maximum resting period of 2.5 seconds.[55,61,62,75]
Minimum bout length of 4 strides with a maximum resting period of 3 seconds.[58]
Minimum of 2 consecutive strides of both feet with a maximum resting period of 3 seconds.[24,26,27,52,78]
Minimum of 2 consecutive strides of both feet with a resting period or any other activity (nonwalking period).[76]
Minimum bout length of 3 consecutive strides and 3 seconds with a maximum resting period of 6 seconds.[69]
Consecutive gait cycles less than 3 seconds apart.[63]
A sequence of consecutive steps.[77]
Stroke reports
Minimum bout length of 3 steps with no threshold set for maximum resting period between consecutive walking bouts.[25]
An episode of 1 or more consecutive gait epochs of 8 seconds.[82]
Gait activity of a minimum of 8 seconds.[80,81]

aThe table summarizes the operational criteria used to define walking bouts, including minimum step or stride count thresholds, minimum duration criteria, and interruption rules. A total of 13 distinct definitions were identified across the included literature. In total, 17 reports did not provide a walking bout definition and are therefore not represented in this table.

The digital outcomes extracted in the included reports were categorized using the framework of macro and micro gait characteristics [42]. Figure 2 summarizes the frequency with which individual parameters were reported across PD and stroke reports. Across both populations, the most commonly reported macro characteristics were the number and length of walking bouts and the number of steps per day. Micro characteristics showed greater variability in reporting across both populations. In PD, the most frequently reported micro characteristics were gait speed, cadence, and stride length, while in stroke, they were gait speed and stride time. Beyond these, a range of additional micro parameters were reported across the literature, though each appeared in too few reports to allow meaningful cross-study comparisons, reflecting the considerable heterogeneity in outcome reporting across the included reports.

‎
Figure 2. Frequency of macro and micro gait characteristics in real-world gait assessment across included reports for Parkinson disease (n=37) and stroke (n=6).

Meta-Analysis

Because of the limited number of compatible stroke reports, the quantitative synthesis was conducted exclusively on the PD cohorts. Four real-world gait parameters met the criteria for meta-analysis: number of steps per day (k=6), gait speed (k=16), cadence (k=10), and step length (k=7). Between-study heterogeneity was extreme for all parameters: in every model, the 95% PI was substantially wider than the CI. The pooled mean for the number of steps per day was 7202 (95% CI 3975-10,429). Notably, the PIs for this macro parameter extended below zero, a mathematical impossibility that quantitatively highlights the profound variance introduced by differing aggregation rules and monitoring durations across reports. For gait speed, the pooled mean was 0.84 m/s (95% CI 0.77-0.91). However, the 95% PI spanned from 0.56 to 1.12 m/s, indicating that a future real-world assessment could reasonably fall anywhere within this wide range. Similarly, cadence yielded a pooled mean of 88.81 steps per minute (95% CI 71.65-105.92; 95% PI 31.94-145.68), and step length was 0.54 m (95% CI 0.41-0.67; 95% PI 0.18-0.91). Forest plots detailing the Hartung-Knapp-Sidik-Jonkman models and PIs for the most frequently reported macro and micro parameters are provided in Figure 3.

‎
Figure 3. Random-effects models for Parkinson disease macro (A) number of steps per day and micro (B) gait speed, (C) cadence, and (D) step length gait parameters measured during real-world gait assessment. The dashed line indicates the pooled mean estimate, while the diamond shape represents the 95% CIs [24,28,44,48,49,51,52,55,56,58-60,62,64-66,69,71,72,74-78]. HKSJ: Hartung-Knapp-Sidik-Jonkman.

While a meta-analysis was not possible for stroke because of the limited number of reports, 4 reports provided the number of bouts, with means of 126.1 (SD 69.8) [80], 104.83 (SD 61.16) [79], 123.3 (SD 61.3) [82], and 535 (SD 213.49) [25] bouts. Bout length was reported in 2 reports, with means of 17.1 (SD 2.6) and 16.46 (SD 11.27) seconds [79,80]. In total, 2 reports provided step counts, with means of 7825 (SD 6428) and 3048 (SD 1983) steps [25,82]. Gait speed was provided in 3 reports (mean 0.45, SD 0.13 m/s [83]; mean 0.71, SD 0.15 m/s [80]; mean 0.69, SD 0.15 m/s [81]), while stride time was reported in 2 (mean 1.34, SD 0.32 seconds and mean 1.37, SD 0.37 seconds) [80,81]. Step length (mean 0.61, SD 0.09 m) [25], step velocity (mean 1.05, SD 0.17 m/s) [25], cadence (mean 90.3, SD 13.8 steps per minute) [82], swing time (mean 0.47, SD 0.03 seconds) [25], and stance time (mean 0.76, SD 0.04 seconds) [25] were each available from a single report.

Certainty of Evidence

Certainty of evidence was very low for all 4 meta-analyzed gait outcomes (Table 3), driven primarily by very serious inconsistency, evidenced by PIs that far exceeded the corresponding CIs. Notably, risk of bias was not serious for any outcome, as no contributing report scored below 75% on the modified quality score. The very low certainty therefore reflects methodological fragmentation across reports rather than deficiencies within them. The 2-level downgrade for inconsistency alone was sufficient to reach the lowest possible certainty rating, making the overall judgment robust regardless of the ratings assigned to the remaining domains.

Table 3. GRADEa certainty of evidence for meta-analyzed real-world gait outcomes in Parkinson disease.
Pooled mean (95% CI)Studies, nRisk of biasInconsistencyIndirectnessImprecisionPublication biasOverall certainty of evidence
Number of steps per day
7202 (3975-10,429)6Not seriousVery seriousbSeriouscSeriousdNot assessede⨁◯◯◯ Very lowb,c,d
Gait speed (m/s)
0.84 (0.77‐0.91)16Not seriousVery seriousbSeriouscNot seriousNot assessede⨁◯◯◯ Very lowb,c
Cadence (steps per minute)
88.81 (71.65‐105.92)10Not seriousVery seriousbSeriouscSeriousdNot assessede⨁◯◯◯ Very lowb,c,d
Step length (m)
0.54 (0.41‐0.67)7Not seriousVery seriousbSeriouscSeriousdNot assessede⨁◯◯◯ Very lowb,c,d

aGRADE: Grading of Recommendations Assessment, Development, and Evaluation.

bDowngraded 2 levels for extreme between-study heterogeneity, indicated by PIs far wider than the CIs.

cDowngraded 1 level because external validity was the weakest quality domain, reflecting limited evidence of participant representativeness, and gait-detection algorithms validated predominantly indoors, in prior studies, or outside the target population.

dDowngraded 1 level because the 95% CI around the pooled mean was wide (relative width ≥30% of the pooled estimate).

eFunnel-plot asymmetry testing was not performed for any outcome. For 2 outcomes, steps per day (k=6) and step length (k=7), fewer than 10 studies were available, which precludes reliable testing. For the remaining 2 outcomes, gait speed (k=16) and cadence (k=10), heterogeneity was too extreme, as reflected in very wide PIs, making any asymmetry test uninterpretable even with an adequate number of studies. No downgrade was applied, although the absence of a gray-literature search and the restriction to English-language publications remain potential unassessed sources of bias.


Principal Findings

This systematic review and meta-analysis provides a comprehensive synthesis of real-world gait assessment using wearable sensors in PD and stroke, examining sensor configurations, validation strategies, outcome reporting practices, and methodological approaches across the available literature. The findings reveal a field that is technically advancing but methodologically fragmented. In PD, a substantial and growing body of research supports the use of IMU-based wearable sensors for real-world gait assessment, with most reports demonstrating high methodological quality and broad coverage of both technical and construct validation. However, considerable heterogeneity was identified across sensor configurations, walking bout definitions, outcome reporting practices, and validation approaches, which the meta-analysis confirmed through consistently wide PIs and high between-study variance. Although overall methodological quality was high, external validity emerged as the weakest domain across both populations, driven by insufficient participant representativeness and limited algorithm validation within the target population, a finding that directly constrains the generalizability of current real-world gait assessment research.

In stroke, the evidence base remains critically underdeveloped, with a small number of reports, several of which drew on overlapping datasets, precluding meaningful quantitative synthesis; therefore, all stroke-related findings in this review must be interpreted as preliminary. This disparity is particularly striking, given the global burden of stroke [3]. Unlike PD research, which has benefited from coordinated multicenter efforts, poststroke monitoring has developed in a fragmented and isolated manner. The substantial methodological groundwork laid by initiatives such as Mobilise-D, including validated sensor protocols, harmonized digital mobility outcomes, and shared algorithmic frameworks, represents a significant opportunity for stroke rehabilitation [93]. Rather than building from scratch, these established frameworks could be critically appraised and adapted for poststroke populations, providing a strong and efficient foundation for developing a robust, independent evidence base for real-world gait monitoring after stroke.

Sensor Configuration and Placement Considerations

Selecting an optimal sensor configuration for real-world gait assessment requires balancing measurement accuracy, clinical usability, patient acceptability, and practical feasibility [94], and the considerable variation in sensor setups observed across the included reports reflects exactly these competing demands. A single sensor positioned at the lower back was the most commonly used configuration in PD and stroke, consistent with broader trends in the field [95]. Placed near the body’s center of mass, the lower back sensor captures whole-body movement patterns, is straightforward to attach, and is generally well tolerated during prolonged monitoring [96,97]. However, trunk-mounted sensors present known challenges in detecting gait events such as initial and final foot contacts, which can affect spatiotemporal accuracy and drift correction [98]. Given that the lower back reflects whole-body movement rather than individual limb kinematics, sensors placed closer to the point of ground contact, such as at the shank or foot, may offer greater accuracy in detecting step-to-step and side-to-side differences that are clinically relevant in this population [99,100]. Whether this translates into meaningfully improved clinical sensitivity remains an open question, as no report in this review directly compared sensor placements in stroke populations. It should also be noted that estimating asymmetry from 2 independently placed sensors introduces the risk of misalignment and calibration errors, a technical challenge that future work in this area will need to carefully account for [101,102]. Beyond placement, the method of sensor attachment influences usability and compliance during prolonged monitoring [103]. Adhesive attachment may reduce patient burden and promote consistent sensor placement during prolonged monitoring, whereas shoe-mounted sensors, although precise, may be constrained by footwear type, sensor fixation, orthopedic considerations, or settings in which people commonly remove shoes indoors [104,105]. Sensor configuration choices should therefore be guided by the target population’s clinical profile, practical context, and the specific gait features of interest [98].

Technical Validation Protocols and Ecological Validity

Despite the well-recognized complexity of real-world walking conditions [24], most validations in the included reports relied on simplified protocols involving uniform walking tasks that do not reflect the dynamic nature of daily life. In response, the most recent studies have introduced multitask and multicontext validation protocols simulating real-world factors such as speed variations, inclines or steps, surface types, path shapes, and cognitive load [106,107]. While these elaborated scenarios provide a more comprehensive assessment of algorithm performance, they are still conducted in controlled settings [106,107]. Ecological validation in outdoor, less standardized environments remains the exception, primarily because of the logistical demands of benchmarking sensor-derived data in natural settings [24]. As the field progresses, advancing and standardizing validation methodologies will be essential to ensure that algorithms are rigorously tested under realistic conditions.

Walking Bout Definitions and Observation Windows

The 13 distinct walking bout definitions identified in this review illustrate a challenge that has been recognized more broadly in the field, as the criteria used to define walking sequences directly determine which data are analyzed and which are excluded, affecting both macro and micro gait parameters [67,108]. Kluge et al [109] used a Delphi consensus process to establish a common terminological framework for real-world walking, defining a walking bout as a sequence of at least 2 consecutive strides of both feet, with start and end determined by a resting period or any other nonwalking activity. The substantial proportion of reports in this review that did not report their walking bout definition suggests that this framework has not yet been widely adopted. Standardization of walking bout criteria is not merely a methodological convenience; it is a prerequisite for meaningful cross-study comparison and the development of reproducible gait benchmarks [110,111].

The observation window over which data are collected represents an equally important consideration. A 7-day monitoring period was the most common duration across both populations in this review, a choice supported by recent evidence from Buekers et al [112], who demonstrated that minimum wear time and number of valid days required for reliable weekly digital mobility outcomes are digital mobility outcome–specific and consistent across health conditions. By expert consensus, more than 12 hours of daily wear time across at least 3 valid days is recommended for studies involving multiple outcomes or conditions, though specific parameters may require more days [112]. Despite this, protocols should still target 7 consecutive days to account for participant adherence and technical issues [112].

Sources of Heterogeneity and Certainty of Evidence

The extreme heterogeneity observed in the meta-analysis, evident in PIs far wider than the CIs, is perhaps the most clinically important finding of this review, formally reflected in very low certainty of evidence ratings across all 4 outcomes, driven by very serious inconsistency and serious indirectness throughout. The wide PIs and high between-study variance confirm that pooled point estimates cannot be interpreted as definitive clinical benchmarks for real-world gait in PD. Rather, they serve as quantitative evidence of the substantial methodological fragmentation in the field. Critically, this uncertainty does not reflect poor quality within individual reports, as reporting and internal validity were consistently strong, but rather the absence of methodological consensus across them. Major procedural inconsistencies concerned hardware placement, walking bout definitions, data aggregation protocols, and outcome nomenclature. These were compounded by limited participant representativeness, and algorithm validation predominantly conducted in simplified laboratory conditions rather than in ecologically valid real-world settings, preventing meaningful comparability. Even when pooling was restricted to reports with comparable bout definitions, considerable residual variance remained. Because inconsistent reporting of clinical covariates (eg, disease severity stages) precluded meta-regression, the exact proportion of variance attributable to clinical versus procedural factors could not be partitioned. A further limitation was the absence of detailed contextual information regarding assistive device use and the severity of gait impairments, both of which substantially influence real-world gait patterns and algorithm performance, further constraining the comparability and generalizability of the findings [52,113]. These findings carry a clear implication, as certainty in real-world gait estimates will not improve through additional studies of the current design alone, and it is therefore imperative that future research, particularly in emerging fields such as poststroke real-world monitoring, adopts harmonized approaches early to avoid compounding the methodological fragmentation currently observed in the PD literature.

The Role of Context in Real-World Gait Assessment

The purpose and context of walking itself remain underexplored in current real-world gait research. In daily life, walking serves diverse functional goals such as navigating the home, community ambulation, exercise, or social participation, each of which may elicit different gait patterns and place different demands on the individual [106,114]. Gait metrics derived from short indoor walking sequences are therefore not directly comparable to those captured during longer community outings, as gait characteristics vary by environment and ambulatory bout length [67]. However, most analyses did not explicitly stratify outcomes by these contextual factors. Kirk et al [76] demonstrated this empirically in PD, showing that integrating GPS-derived indoor and outdoor contextual information with wearable-derived gait outcomes revealed meaningful differences in walking behavior across environments, with between-group differences emerging only during outdoor walking. Without contextual awareness, clinically relevant differences in gait may therefore be obscured or misattributed. Beyond environmental context, internal biological and psychological factors including sleep quality, nutritional state, hormonal fluctuations, and anxiety may introduce within-person variability that is difficult to separate from disease-related fluctuations; yet, these are not accounted for in current real-world monitoring studies [115-118].

Limitations

This review has several limitations that should be considered when interpreting its findings. First, only studies published in English were included, which may have introduced language bias and may have led to the exclusion of relevant evidence published in other languages. Second, multiple included reports were derived from overlapping datasets, which may have introduced bias toward specific methodological approaches and reduced the diversity of the evidence base. This concern is particularly acute in stroke, where the small number of reports and their limited independence substantially constrain the conclusions that can be drawn. Third, the review focused exclusively on macro and micro gait characteristics, and important aspects of real-world mobility such as turning, freezing of gait, and postural transitions were not examined. Finally, despite strict data extraction criteria applied prior to pooling, residual heterogeneity limited the clinical utility of the pooled estimates and precluded meaningful subanalyses by clinical variables such as disease severity, faller status, or motor fluctuation state. Although the small number of reports providing compatible metrics constrained these analyses, this limitation reflects the emerging nature of the field rather than a flaw in the review methodology and should serve as a clear signal of where standardization efforts are most urgently needed.

Conclusions

Unlike previous reviews that have focused narrowly on single diseases, specific hardware, or classification algorithms, this review is novel in contrasting PD and stroke populations to highlight shared assessment challenges. It demonstrates that wearable sensor-based real-world gait assessment in PD is technically advancing but methodologically fragmented, while the stroke evidence base remains critically underdeveloped. By quantifying this heterogeneity, this review brings to the field a critical recognition: the observed variance is substantially compounded by procedural and algorithmic inconsistencies, limiting the ability to isolate genuine clinical differences. Important foundational work has already been undertaken to address this, including a comprehensive roadmap for integrating wearable sensors into clinical practice and a Delphi consensus establishing a common terminological framework for real-world walking [5,109]. In stroke, the sparse evidence base represents an opportunity to critically assess and adopt these harmonized frameworks early. Resolving these methodological inconsistencies carries profound real-world implications. Collaborative standardization of terminology, validation procedures, and core outcomes is essential to transition wearable sensors from exploratory research tools to reliable clinical instruments. Ultimately, establishing reliable digital mobility outcomes will support the International Classification of Functioning, Disability and Health’s underused construct of performance, enabling personalized care, precise remote monitoring, and an improved picture of functional mobility in everyday life.

Acknowledgments

The authors declare the use of generative AI (GAI) in the research and writing process. According to the GAIDeT taxonomy (2025), the following tasks were delegated to GAI tools under full human supervision: proofreading and editing, adapting, and reformatting. The GAI tools used were GPT 5.6 Thinking (OpenAI) and Claude Opus 4.8 (Anthropic). Responsibility for the final manuscript lies entirely with the authors. GAI tools are not listed as authors and do not bear responsibility for the final outcomes.

Funding

This research was funded by the P&K Foundation.

Data Availability

The datasets generated or analyzed during this study are available from the corresponding author on reasonable request.

Authors' Contributions

SN, JA-S, and CEA contributed to the conceptualization and methodology of the study. SN, JA-S, and ACN conducted the investigation. SN and JA-S performed the formal analysis and wrote the original draft. CEA, LC, and ARL contributed to supervision, project administration, review, and editing. All authors reviewed and approved the final manuscript.

Conflicts of Interest

LC is co-founder and owns shares of mHealth Technologies s.r.l. The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Multimedia Appendix 1

Search strategy per database.

DOCX File, 25 KB

Multimedia Appendix 2

Quality checklist, calculation of quality score, and quality score per study.

DOCX File, 75 KB

Checklist 1

PRISMA checklist.

DOCX File, 28 KB

Checklist 2

PRISMA-S checklist.

DOCX File, 23 KB

Checklist 3

PRISMA 2020 Abstract checklist.

DOCX File, 21 KB

  1. Intersectoral global action plan on epilepsy and other neurological disorders 2022-2031. World Health Organization. 2023. URL: https://www.who.int/publications/i/item/9789240076624 [Accessed 2026-08-30]
  2. Celik Y, Stuart S, Woo WL, Godfrey A. Gait analysis in neurological populations: progression in the use of wearables. Med Eng Phys. Jan 2021;87(1):9-29. [CrossRef] [Medline]
  3. Roth GA, Mensah GA, Johnson CO. Global burden of cardiovascular diseases and risk factors, 1990–2019. J Am Coll Cardiol. Dec 2020;76(25):2982-3021. [CrossRef]
  4. Horak FB, Mancini M. Digital gait and balance measures. In: Digital Technologies in Movement Disorders. Elsevier; 2023:115-151. [CrossRef]
  5. Del Din S, Kirk C, Yarnall AJ, Rochester L, Hausdorff JM. Body-worn sensors for remote monitoring of Parkinson’s disease motor symptoms: vision, state of the art, and challenges ahead. J Parkinsons Dis. 2021;11(s1):S35-S47. [CrossRef] [Medline]
  6. Buckley C, Alcock L, McArdle R, et al. The role of movement analysis in diagnosing and monitoring neurodegenerative conditions: insights from gait and postural control. Brain Sci. Feb 6, 2019;9(2):34. [CrossRef] [Medline]
  7. Zanardi APJ, da Silva ES, Costa RR, et al. Gait parameters of Parkinson’s disease compared with healthy controls: a systematic review and meta-analysis. Sci Rep. Jan 12, 2021;11(1):752. [CrossRef] [Medline]
  8. Mak MK, Wong-Yu IS, Shen X, Chung CL. Long-term effects of exercise and physical therapy in people with Parkinson disease. Nat Rev Neurol. Nov 2017;13(11):689-703. [CrossRef] [Medline]
  9. Jørgensen HS, Nakayama H, Raaschou HO, Olsen TS. Recovery of walking function in stroke patients: the Copenhagen Stroke Study. Arch Phys Med Rehabil. Jan 1995;76(1):27-32. [CrossRef] [Medline]
  10. Mazzà C, Alcock L, Aminian K, et al. Technical validation of real-world monitoring of gait: a multicentric observational study. BMJ Open. Dec 2, 2021;11(12):e050785. [CrossRef] [Medline]
  11. Morris ME, Matyas TA, Iansek R, Summers JJ. Temporal stability of gait in Parkinson’s disease. Phys Ther. Jul 1996;76(7):763-777. [CrossRef] [Medline]
  12. de Rooij IJM, van de Port IGL, van der Heijden LLM, Meijer JWG, Visser-Meily JMA. Perceived barriers and facilitators for gait-related participation in people after stroke: From a patients’ perspective. Physiother Theory Pract. Dec 2021;37(12):1337-1345. [CrossRef] [Medline]
  13. Jankovic J. Parkinson’s disease: clinical features and diagnosis. J Neurol Neurosurg Psychiatry. Apr 1, 2008;79(4):368-376. [CrossRef]
  14. Towards a common language for functioning, disability and health. World Health Organization (WHO). 2002. URL: https://www.who.int/docs/default-source/classification/icf/icfbeginnersguide.pdf?sfvrsn=eead63d3_4 [Accessed 2026-08-30]
  15. Kwakkel G, Stinear C, Essers B, et al. Motor rehabilitation after stroke: European Stroke Organisation (ESO) consensus-based definition and guiding framework. Eur Stroke J. Dec 2023;8(4):880-894. [CrossRef] [Medline]
  16. Moore SA, Boyne P, Fulk G, Verheyden G, Fini NA. Walk the Talk: current evidence for walking recovery after stroke, future pathways and a mission for research and clinical practice. Stroke. Nov 2022;53(11):3494-3505. [CrossRef] [Medline]
  17. Warmerdam E, Hausdorff JM, Atrsaei A, et al. Long-term unsupervised mobility assessment in movement disorders. Lancet Neurol. May 2020;19(5):462-470. [CrossRef] [Medline]
  18. Del Din S, Godfrey A, Mazzà C, Lord S, Rochester L. Free-living monitoring of Parkinson’s disease: lessons from the field. Mov Disord. Sep 2016;31(9):1293-1313. [CrossRef] [Medline]
  19. Wendel N, Macpherson CE, Webber K, et al. Accuracy of activity trackers in Parkinson disease: should we prescribe them? Phys Ther. Aug 1, 2018;98(8):705-714. [CrossRef] [Medline]
  20. El-Gohary M, Pearson S, McNames J, et al. Continuous monitoring of turning in patients with movement disability. Sensors (Basel). Dec 27, 2013;14(1):356-369. [CrossRef] [Medline]
  21. Mancini M, El-Gohary M, Pearson S, et al. Continuous monitoring of turning in Parkinson’s disease: rehabilitation potential. NRE. 2015;37(1):3-10. [CrossRef]
  22. Zhang Y, Xu L, Pan X, Chen L, Zhong W, Li J. Contextual factors associated with walking performance after stroke: a systematic review and meta-analysis. Front Neurol. 2025;16:1635024. [CrossRef]
  23. Kirk C, Packer E, Polhemus A, et al. A systematic review of real-world gait-related digital mobility outcomes in Parkinson’s disease. NPJ Digit Med. Sep 30, 2025;8(1):585. [CrossRef] [Medline]
  24. Salis F, Bertuletti S, Bonci T, et al. A multi-sensor wearable system for the assessment of diseased gait in real-world conditions. Front Bioeng Biotechnol. 2023;11:1143248. [CrossRef] [Medline]
  25. Moore SA, Hickey A, Lord S, Del Din S, Godfrey A, Rochester L. Comprehensive measurement of stroke gait characteristics with a single accelerometer in the laboratory and community: a feasibility, validity and reliability study. J NeuroEngineering Rehabil. Dec 2017;14(1):130. [CrossRef]
  26. Romijnders R, Salis F, Hansen C, et al. Ecological validity of a deep learning algorithm to detect gait events from real-life walking bouts in mobility-limiting diseases. Front Neurol. 2023;14:1247532. [CrossRef] [Medline]
  27. Kirk C, Küderle A, Micó-Amigo ME, et al. Mobilise-D insights to estimate real-world walking speed in multiple conditions with a wearable device. Sci Rep. Jan 19, 2024;14(1):1754. [CrossRef] [Medline]
  28. Kirk C, Zia Ur Rehman R, Galna B, et al. Can digital mobility assessment enhance the clinical assessment of disease severity in Parkinson’s disease? J Parkinsons Dis. 2023;13(6):999-1009. [CrossRef] [Medline]
  29. Polvorinos-Fernández C, Sigcha L, Borzì L, et al. Evaluating motor symptoms in Parkinson’s disease through wearable sensors: a systematic review of digital biomarkers. Applied Sciences. 2024;14(22):10189. [CrossRef]
  30. Giannakopoulou KM, Roussaki I, Demestichas K. Internet of Things technologies and machine learning methods for Parkinson’s disease diagnosis, monitoring and management: a systematic review. Sensors (Basel). Feb 24, 2022;22(5):1799. [CrossRef] [Medline]
  31. Sigcha L, Borzì L, Amato F, et al. Deep learning and wearable sensors for the diagnosis and monitoring of Parkinson’s disease: a systematic review. Expert Syst Appl. Nov 2023;229:120541. [CrossRef]
  32. Packer E, Debelle H, Bailey HGB, et al. Systematic review of wearables assessing medication effect on motor function and symptoms in Parkinson’s disease. NPJ Parkinsons Dis. May 22, 2025;11(1):135. [CrossRef] [Medline]
  33. Li H, Zecca M, Huang J. Evaluating the utility of wearable sensors for the early diagnosis of Parkinson disease: systematic review. J Med Internet Res. 2025;27:e69422. [CrossRef]
  34. Boukhennoufa I, Zhai X, Utti V, Jackson J, McDonald-Maier KD. Wearable sensors and machine learning in post-stroke rehabilitation assessment: a systematic review. Biomed Signal Process Control. Jan 2022;71:103197. [CrossRef]
  35. Bernaldo de Quirós M, Douma EH, van den Akker-Scheek I, Lamoth CJC, Maurits NM. Quantification of movement in stroke patients under free living conditions using wearable sensors: a systematic review. Sensors (Basel). Jan 28, 2022;22(3):1050. [CrossRef] [Medline]
  36. Silva RD, Silva SD, Cardoso DCR, et al. Psychometric properties of wearable technologies to assess post-stroke gait parameters: a systematic review. Gait Posture. Sep 2024;113:543-552. [CrossRef]
  37. Jiao Y, Hart R, Reading S, Zhang Y. Systematic review of automatic post-stroke gait classification systems. Gait Posture. Mar 2024;109:259-270. [CrossRef]
  38. Page MJ, McKenzie JE, Bossuyt PM, et al. The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. BMJ. 2021:n71. [CrossRef]
  39. Rethlefsen ML, Kirtley S, Waffenschmidt S, et al. PRISMA-S: an extension to the PRISMA Statement for Reporting Literature Searches in Systematic Reviews. Syst Rev. Jan 26, 2021;10(1):39. [CrossRef] [Medline]
  40. Khan KS, Kunz R, Kleijnen J, Antes G. Five steps to conducting a systematic review. J R Soc Med. Mar 2003;96(3):118-121. [CrossRef] [Medline]
  41. Liberati A, Altman DG, Tetzlaff J, et al. The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate healthcare interventions: explanation and elaboration. BMJ. Jul 21, 2009;339(jul21 1):b2700-b2700. [CrossRef]
  42. Lord S, Galna B, Rochester L. Moving forward on gait measurement: toward a more refined approach. Mov Disord. Sep 15, 2013;28(11):1534-1543. [CrossRef] [Medline]
  43. Del Din S, Galna B, Lord S, et al. Falls risk in relation to activity exposure in high-risk older adults. J Gerontol A Biol Sci Med Sci. May 22, 2020;75(6):1198-1205. [CrossRef] [Medline]
  44. Del Din S, Galna B, Godfrey A, et al. Analysis of free-living gait in older adults with and without Parkinson’s disease and with and without a history of falls: identifying generic and disease-specific characteristics. J Gerontol A Biol Sci Med Sci. Mar 14, 2019;74(4):500-506. [CrossRef] [Medline]
  45. Hubble RP, Naughton GA, Silburn PA, Cole MH. Wearable sensor use for assessing standing balance and walking stability in people with Parkinson’s disease: a systematic review. PLoS One. 2015;10(4):e0123705. [CrossRef] [Medline]
  46. Page MJ, Higgins JP, Sterne JA. Chapter 13: Assessing risk of bias due to missing evidence in a meta-analysis [last updated August 2024]. In: Higgins JP, Thomas J, Chandler J, Cumpston M, Li T, Page MJ, editors. Cochrane Handbook for Systematic Reviews of Interventions Version 6.5. Cochrane; 2024. URL: https://www.cochrane.org/handbook [Accessed 2026-08-18]
  47. Borenstein M. How to understand and report heterogeneity in a meta-analysis: the difference between I-squared and prediction intervals. Integr Med Res. Dec 2023;12(4):101014. [CrossRef] [Medline]
  48. Atrsaei A, Corrà MF, Dadashi F, et al. Gait speed in clinical and daily living assessments in Parkinson’s disease patients: performance versus capacity. NPJ Parkinsons Dis. Mar 5, 2021;7(1):24. [CrossRef] [Medline]
  49. Adams JL, Dinesh K, Snyder CW, et al. A real-world study of wearable sensors in Parkinson’s disease. NPJ Parkinsons Dis. Nov 29, 2021;7(1):106. [CrossRef] [Medline]
  50. Rehman RZU, Guan Y, Shi JQ, et al. Investigating the impact of environment and data aggregation by walking bout duration on Parkinson’s disease classification using machine learning. Front Aging Neurosci. 2022;14:808518. [CrossRef] [Medline]
  51. Roth N, Ullrich M, Kuderle A, et al. Real-world stair ambulation characteristics differ between prospective fallers and non-fallers in Parkinson’s disease. IEEE J Biomed Health Inform. Sep 2022;26(9):4733-4742. [CrossRef] [Medline]
  52. Micó-Amigo ME, Bonci T, Paraschiv-Ionescu A, et al. Assessing real-world gait with digital technology? Validation, insights and recommendations from the Mobilise-D consortium. J NeuroEngineering Rehabil. 2023;20(1):78. [CrossRef]
  53. Coates L, Shi J, Rochester L, Del Din S, Pantall A. Entropy of real-world gait in Parkinson’s disease determined from wearable sensors as a digital marker of altered ambulatory behavior. Sensors (Basel). May 5, 2020;20(9):2631. [CrossRef] [Medline]
  54. Terashi H, Taguchi T, Ueta Y, Okubo Y, Mitoma H, Aizawa H. Analysis of non-invasive gait recording under free-living conditions in patients with Parkinson’s disease: relationship with global cognitive function and motor abnormalities. BMC Neurol. Apr 29, 2020;20(1):161. [CrossRef] [Medline]
  55. Shah VV, McNames J, Mancini M, et al. Laboratory versus daily life gait characteristics in patients with multiple sclerosis, Parkinson’s disease, and matched controls. J NeuroEngineering Rehabil. Dec 2020;17(1):159. [CrossRef]
  56. Bouça-Machado R, Branco D, Fonseca G, et al. Kinematic and clinical outcomes to evaluate the efficacy of a multidisciplinary intervention on functional mobility in Parkinson’s disease. Front Neurol. 2021;12:637620. [CrossRef] [Medline]
  57. Ullrich M, Mucke A, Kuderle A, et al. Detection of unsupervised standardized gait tests from real-world inertial sensor data in Parkinson’s disease. IEEE Trans Neural Syst Rehabil Eng. 2021;29:2103-2111. [CrossRef] [Medline]
  58. Ullrich M, Roth N, Kuderle A, et al. Fall risk prediction in Parkinson’s disease using real-world inertial sensor gait data. IEEE J Biomed Health Inform. Jan 2023;27(1):319-328. [CrossRef] [Medline]
  59. Caballol N, Bayés À, Prats A, Martín-Baranera M, Quispe P. Feasibility of a wearable inertial sensor to assess motor complications and treatment in Parkinson’s disease. PLoS One. 2023;18(2):e0279910. [CrossRef] [Medline]
  60. Cohen M, Herman T, Ganz N, Badichi I, Gurevich T, Hausdorff JM. Multidisciplinary intensive rehabilitation program for people with Parkinson’s disease: gaps between the clinic and real-world mobility. Int J Environ Res Public Health. Feb 21, 2023;20(5):3806. [CrossRef] [Medline]
  61. Shah VV, McNames J, Harker G, et al. Effect of bout length on gait measures in people with and without Parkinson’s disease during daily life. Sensors (Basel). Oct 12, 2020;20(20):5769. [CrossRef] [Medline]
  62. Shah VV, McNames J, Mancini M, et al. Digital biomarkers of mobility in Parkinson’s disease during daily living. J Parkinsons Dis. 2020;10(3):1099-1111. [CrossRef] [Medline]
  63. Corrà MF, Atrsaei A, Sardoreira A, et al. Comparison of laboratory and daily-life gait speed assessment during ON and OFF states in Parkinson’s disease. Sensors (Basel). Jun 9, 2021;21(12):3974. [CrossRef] [Medline]
  64. Mirelman A, Volkov J, Salomon A, et al. Digital mobility measures: a window into real-world severity and progression of Parkinson’s disease. Mov Disord. Feb 2024;39(2):328-338. [CrossRef] [Medline]
  65. Zampogna A, Borzì L, Rinaldi D, et al. Unveiling the unpredictable in Parkinson’s disease: sensor-based monitoring of dyskinesias and freezing of gait in daily life. Bioengineering (Basel). Apr 29, 2024;11(5):440. [CrossRef] [Medline]
  66. Nishi Y, Fujii S, Ikuno K, Terasawa Y, Morioka S. Adjustability of gait speed in clinics and free-living environments for people with Parkinson’s disease. J Mov Disord. Oct 2024;17(4):416-424. [CrossRef] [Medline]
  67. Del Din S, Godfrey A, Galna B, Lord S, Rochester L. Free-living gait characteristics in ageing and Parkinson’s disease: impact of environment and ambulatory bout length. J NeuroEngineering Rehabil. Dec 2016;13(1):46. [CrossRef]
  68. Galperin I, Hillel I, Del Din S, et al. Associations between daily-living physical activity and laboratory-based assessments of motor severity in patients with falls and Parkinson’s disease. Parkinsonism Relat Disord. May 2019;62:85-90. [CrossRef] [Medline]
  69. Mancini M, Curtze C, Stuart S, et al. The impact of freezing of gait on balance perception and mobility in community-living with parkinson’S disease. Presented at: 2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC); Jul 18-21, 2018. [CrossRef]
  70. Weiss A, Herman T, Giladi N, Hausdorff JM. Objective assessment of fall risk in Parkinson’s disease using a body-fixed sensor worn for 3 days. PLoS One. 2014;9(5):e96675. [CrossRef] [Medline]
  71. Weiss A, Herman T, Giladi N, Hausdorff JM. Association between community ambulation walking patterns and cognitive function in patients with Parkinson’s disease: further insights into motor-cognitive links. Parkinsons Dis. 2015;2015:547065. [CrossRef] [Medline]
  72. Toosizadeh N, Mohler J, Lei H, Parvaneh S, Sherman S, Najafi B. Motor performance assessment in Parkinson’s disease: association between objective in-clinic, objective in-home, and subjective/semi-objective measures. PLoS One. 2015;10(4):e0124763. [CrossRef] [Medline]
  73. Weiss A, Herman T, Giladi N, Hausdorff JM. New evidence for gait abnormalities among Parkinson’s disease patients who suffer from freezing of gait: insights using a body-fixed sensor worn for 3 days. J Neural Transm (Vienna). Mar 2015;122(3):403-410. [CrossRef] [Medline]
  74. Atrsaei A, Hansen C, Elshehabi M, et al. Effect of fear of falling on mobility measured during lab and daily activity assessments in Parkinson’s disease. Front Aging Neurosci. 2021;13:722830. [CrossRef] [Medline]
  75. Shah VV, McNames J, Carlson‐Kuhta P, et al. Effect of levodopa and environmental setting on gait and turning digital markers related to falls in people with Parkinson’s disease. Movement Disord Clin Pract. Feb 2023;10(2):223-230. [CrossRef]
  76. Kirk C, Rehman RZU, Galna B, et al. Toward an understanding of real-world mobility in Parkinson’s: insights from enhanced contextualisation using GPS-derived location and data-driven modeling of walking speed. Front Aging Neurosci. 2026;18:1746429. [CrossRef] [Medline]
  77. Sidoroff V, Moradi H, Prigent G, et al. Moving beyond the hospital: in-depth characterization of daily-life mobility in patients with atypical Parkinsonian disorders. NPJ Parkinsons Dis. Jan 12, 2026;12(1):34. [CrossRef] [Medline]
  78. Yarnall AJ, Alcock L, Schlenstedt C, et al. The future of clinical trials? Monitoring mobility as an outcome measure in Parkinson’s disease: the mobilise-D study. J Parkinsons Dis. Feb 2026;16(1):110-124. [CrossRef] [Medline]
  79. Sánchez MC, Bussmann J, Janssen W, et al. Accelerometric assessment of different dimensions of natural walking during the first year after stroke: recovery of amount, distribution, quality and speed of walking. J Rehabil Med. Sep 2015;47(8):714-721. [CrossRef] [Medline]
  80. Punt M, Bruijn SM, van Schooten KS, et al. Characteristics of daily life gait in fall and non fall-prone stroke survivors and controls. J NeuroEngineering Rehabil. Dec 2016;13(1):67. [CrossRef]
  81. Punt M, Bruijn SM, Wittink H, van de Port IG, van Dieën JH. Do clinical assessments, steady-state or daily-life gait characteristics predict falls in ambulatory chronic stroke survivors? J Rehabil Med. May 16, 2017;49(5):402-409. [CrossRef] [Medline]
  82. van de Port I, Punt M, Meijer JW. Walking activity and its determinants in free-living ambulatory people in a chronic phase after stroke: a cross-sectional study. Disabil Rehabil. Mar 2020;42(5):636-641. [CrossRef] [Medline]
  83. Felius RAW, Punt M, Wouda NC, Geerars M, Bruijn SM, van Dieën JH. Predicting community walking after stroke. Front Stroke. 2025;4:1523242. [CrossRef] [Medline]
  84. Aminian K, Najafi B. Capturing human motion using body‐fixed sensors: outdoor measurement and clinical applications. Comput Anim Virtual Worlds. May 2004;15(2):79-94. [CrossRef]
  85. Del Din S, Godfrey A, Rochester L. Validation of an accelerometer to quantify a comprehensive battery of gait characteristics in healthy older adults and Parkinson’s disease: toward clinical and at home use. IEEE J Biomed Health Inform. May 2016;20(3):838-847. [CrossRef] [Medline]
  86. Morris R, Stuart S, McBarron G, Fino PC, Mancini M, Curtze C. Validity of Mobility Lab (version 2) for gait assessment in young adults, older adults and Parkinson’s disease. Physiol Meas. Sep 30, 2019;40(9):095003. [CrossRef] [Medline]
  87. Rampp A, Barth J, Schülein S, Gaßmann KG, Klucken J, Eskofier BM. Inertial sensor-based stride parameter calculation from gait sequences in geriatric patients. IEEE Trans Biomed Eng. Apr 2015;62(4):1089-1097. [CrossRef] [Medline]
  88. Mariani B, Jiménez MC, Vingerhoets FJG, Aminian K. On-shoe wearable sensors for gait and turning assessment of patients with Parkinson’s disease. IEEE Trans Biomed Eng. Jan 2013;60(1):155-158. [CrossRef] [Medline]
  89. Roth N, Küderle A, Prossel D, Gassner H, Eskofier BM, Kluge F. An inertial sensor-based gait analysis pipeline for the assessment of real-world stair ambulation parameters. Sensors (Basel). Sep 30, 2021;21(19):6559. [CrossRef] [Medline]
  90. Prigent G, Aminian K, Cereatti A, et al. A robust walking detection algorithm using a single foot-worn inertial sensor: validation in real-life settings. Med Biol Eng Comput. Sep 2023;61(9):2341-2352. [CrossRef] [Medline]
  91. Punt M, van Alphen B, van de Port IG, et al. Clinimetric properties of a novel feedback device for assessing gait parameters in stroke survivors. J NeuroEngineering Rehabil. 2014;11(1):30. [CrossRef]
  92. Herman T, Rosenberg-Katz K, Jacob Y, et al. White matter hyperintensities in Parkinson’s disease: do they explain the disparity between the postural instability gait difficulty and tremor dominant subtypes? PLoS One. 2013;8(1):e55193. [CrossRef] [Medline]
  93. Mobilise-D project. 2019. URL: https://www.mobilise-d.eu [Accessed 2026-08-18]
  94. Keogh A, Dorn JF, Walsh L, Calvo F, Caulfield B. Comparing the usability and acceptability of wearable sensors among older Irish adults in a real-world context: observational study. JMIR Mhealth Uhealth. Apr 20, 2020;8(4):e15704. [CrossRef] [Medline]
  95. Brognara L, Palumbo P, Grimm B, Palmerini L. Assessing gait in Parkinson’s disease using wearable motion sensors: a systematic review. Diseases. Feb 5, 2019;7(1):18. [CrossRef] [Medline]
  96. Keogh A, Alcock L, Brown P, et al. Acceptability of wearable devices for measuring mobility remotely: observations from the Mobilise-D technical validation study. Digit Health. 2023;9:20552076221150745. [CrossRef] [Medline]
  97. Rispens SM, Pijnappels M, van Schooten KS, Beek PJ, Daffertshofer A, van Dieën JH. Consistency of gait characteristics as determined from acceleration data collected at different trunk locations. Gait Posture. May 2014;40(1):187-192. [CrossRef]
  98. Trojaniello D, Cereatti A, Pelosin E, et al. Estimation of step-by-step spatio-temporal parameters of normal and impaired gait using shank-mounted magneto-inertial sensors: application to elderly, hemiparetic, parkinsonian and choreic gait. J NeuroEngineering Rehabil. 2014;11(1):152. [CrossRef]
  99. Storm FA, Buckley CJ, Mazzà C. Gait event detection in laboratory and real life settings: accuracy of ankle and waist sensor based methods. Gait Posture. Oct 2016;50:42-46. [CrossRef]
  100. Ben Mansour K, Rezzoug N, Gorce P. Analysis of several methods and inertial sensors locations to assess gait parameters in able-bodied subjects. Gait Posture. Oct 2015;42(4):409-414. [CrossRef]
  101. Guimarães V, Sousa I, Correia MV. Orientation-invariant spatio-temporal gait analysis using foot-worn inertial sensors. Sensors (Basel). Jun 7, 2021;21(11):3940. [CrossRef] [Medline]
  102. Küderle A, Roth N, Zlatanovic J, Zrenner M, Eskofier B, Kluge F. The placement of foot-mounted IMU sensors does affect the accuracy of spatial parameters during regular walking. PLoS One. 2022;17(6):e0269567. [CrossRef] [Medline]
  103. Keogh A, Argent R, Anderson A, Caulfield B, Johnston W. Assessing the usability of wearable devices to measure gait and physical activity in chronic conditions: a systematic review. J NeuroEngineering Rehabil. 2021;18(1):138. [CrossRef]
  104. Lewin M, Price C, Nester C. Can a shoe-mounted IMU identify the effects of orthotics in ways comparable to gait laboratory measurements? J Foot Ankle Res. Sep 5, 2023;16(1):54. [CrossRef] [Medline]
  105. Napier C, Willy RW, Hannigan BC, McCann R, Menon C. The effect of footwear, running speed, and location on the validity of two commercially available inertial measurement units during running. Front Sports Act Living. 2021;3:643385. [CrossRef] [Medline]
  106. Scott K, Bonci T, Salis F, et al. Design and validation of a multi-task, multi-context protocol for real-world gait simulation. J NeuroEngineering Rehabil. 2022;19(1):141. [CrossRef]
  107. Warmerdam E, Romijnders R, Geritz J, et al. Proposed mobility assessments with simultaneous full-body inertial measurement units and optical motion capture in healthy adults and neurological patients for future validation studies: study protocol. Sensors (Basel). Aug 30, 2021;21(17):5833. [CrossRef] [Medline]
  108. Storm FA, Nair KPS, Clarke AJ, Van der Meulen JM, Mazzà C. Free-living and laboratory gait characteristics in patients with multiple sclerosis. PLoS One. 2018;13(5):e0196463. [CrossRef] [Medline]
  109. Kluge F, Del Din S, Cereatti A, et al. Consensus based framework for digital mobility monitoring. PLoS One. 2021;16(8):e0256541. [CrossRef] [Medline]
  110. Palmerini L, Reggi L, Bonci T, et al. Mobility recorded by wearable devices and gold standards: the Mobilise-D procedure for data standardization. Sci Data. Jan 19, 2023;10(1):38. [CrossRef] [Medline]
  111. Barry G, Galna B, Lord S, Rochester L, Godfrey A. Defining ambulatory bouts in free-living activity: impact of brief stationary periods on bout metrics. Gait Posture. Oct 2015;42(4):594-597. [CrossRef]
  112. Buekers J, Chernova J, Koch S, et al. Digital assessment of real-world walking in people with impaired mobility: how many hours and days are needed? Int J Behav Nutr Phys Act. Nov 21, 2025;22(1):148. [CrossRef] [Medline]
  113. Kluge F, Brand YE, Micó-Amigo ME, et al. Real-world gait detection using a wrist-worn inertial sensor: validation study. JMIR Form Res. May 1, 2024;8:e50035. [CrossRef] [Medline]
  114. Mc Ardle R, Del Din S, Donaghy P, Galna B, Thomas AJ, Rochester L. The impact of environment on gait assessment: considerations from real-world gait analysis in dementia subtypes. Sensors (Basel). Jan 26, 2021;21(3):813. [CrossRef] [Medline]
  115. Zhang X, Jin Y, Wang M, et al. The impact of anxiety on gait impairments in Parkinson’s disease: insights from sensor-based gait analysis. J NeuroEngineering Rehabil. 2024;21(1):68. [CrossRef]
  116. Dos Anjos L, Rodrigues F, Scataglini S, Baptista RR, Lobo da Costa P, Vieira MF. Trunk variability and local dynamic stability during gait after generalized fatigue induced by incremental exercise test in young women in different phases of the menstrual cycle. PeerJ. 2023;11:e16223. [CrossRef] [Medline]
  117. Mendes J, Borges N, Santos A, et al. Nutritional status and gait speed in a nationwide population-based sample of older adults. Sci Rep. 2018;8(1):4227. [CrossRef]
  118. Agmon M, Shochat T, Kizony R. Sleep quality is associated with walking under dual-task, but not single-task performance. Gait Posture. Sep 2016;49:127-131. [CrossRef]


‎
CI: confidence interval
GRADE: Grading of Recommendations Assessment, Development, and Evaluation
ICICLE-GAIT: Incidence of Cognitive Impairment in Cohorts with Longitudinal Evaluation-GAIT
IMU: inertial measurement unit
PD: Parkinson disease
PI: prediction interval
PRISMA: Preferred Reporting Items for Systematic Reviews and Meta-Analyses
PRISMA-S: Preferred Reporting Items for Systematic Reviews and Meta-Analyses Literature Search Extension
V-TIME: Virtual Reality-Treadmill Combined Intervention for Enhancing Mobility and Reducing Falls in the Elderly


Edited by Stefano Brini; submitted 17.Feb.2026; peer-reviewed by Encarna Mico-Amigo, Jie Fei; final revised version received 04.Aug.2026; accepted 10.Aug.2026; published 25.Sep.2026.

Copyright

© Saskia Neumann, Jose Albites-Sanabria, Aileen C Naef, Andreas R Luft, Lorenzo Chiari, Chris Easthope Awai. Originally published in the Journal of Medical Internet Research (https://www.jmir.org), 25.Sep.2026.

This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in the Journal of Medical Internet Research (ISSN 1438-8871), is properly cited. The complete bibliographic information, a link to the original publication on https://www.jmir.org/, as well as this copyright and license information must be included.