AI-Enabled Wearables for Motor Function Assessment and Rehabilitation in Parkinson Disease: Scoping Review

AI-Enabled Wearables for Motor Function Assessment and Rehabilitation in Parkinson Disease: Scoping Review

AI-Enabled Wearables for Motor Function Assessment and Rehabilitation in Parkinson Disease: Scoping Review

Review

1School of Nursing, Zunyi Medical University, Zunyi, China

2Affiliated Hospital of Zunyi Medical College, Zunyi, China

*these authors contributed equally

Corresponding Author:

Mingtao Quan, MSc

School of Nursing

Zunyi Medical University

No. 201 Dalian Road, Huichuan District

Zunyi, 563000

China

Phone: 86 18185066688

Email: 865768818@qq.com


Background: Artificial intelligence (AI)–enabled wearable devices are rapidly emerging in rehabilitation and motor function assessment for patients with Parkinson disease (PD). However, evidence remains fragmented, integration into nursing practice is limited, and comprehensive synthesis is lacking.

Objective: This study aimed to summarize studies on AI-enabled wearable devices for PD rehabilitation and motor function assessment, describing device types, monitored indicators, algorithms, and application characteristics, and identifying research gaps and barriers to clinical translation.

Methods: Guided by the PRISMA-ScR (Preferred Reporting Items for Systematic Reviews and Meta-Analyses extension for Scoping Reviews) framework, 9 databases (China National Knowledge Infrastructure, Wanfang Data, SinoMed, Cochrane Library, PubMed, Web of Science, CINAHL, Scopus, and Embase) were searched from inception to December 2025. Eligible studies were published in English or Chinese from January 1, 2020, onward and enrolled people with PD using noninvasive, body-worn AI-enabled wearable devices for rehabilitation, assessment, or monitoring. Dissertations and full conference papers were included, whereas preprints and conference abstracts were excluded. Methodological quality was appraised using the Mixed Methods Appraisal Tool, 2018 tool. Results were synthesized narratively and mapped to characterize devices, sensing modalities, algorithms, and evaluation methods.

Results: A total of 66 studies involving approximately 3579 participants were included. Wearable devices mainly comprised multisensor modules, smart insoles, and wrist-worn devices, with accelerometers being the most frequently used sensors. Data collection was predominantly passive, and most studies were conducted in laboratory or clinical settings using single- or short-term sessions. Internal validation approaches, particularly leave-one-out and k-fold cross-validation, were common, whereas external validation was rare, and reporting of calibration and clinical decision thresholds was limited. Sensitivity and accuracy were the most frequently reported performance metrics, highlighting substantial heterogeneity in analytical methods and outcome reporting.

Conclusions: This scoping review systematically synthesized evidence on AI-enabled wearable devices for motor function assessment and rehabilitation in PD, complemented by an evidence map and guided by a rehabilitation- and nursing-oriented perspective, and identified key translational gaps between proof-of-concept studies and real-world rehabilitation workflows. Compared with previous reviews that primarily focused on monitoring functions or device performance, this review places greater emphasis on rehabilitation applications and nurse-led translation into practice, and proposes a conceptual “challenges and opportunities” framework to inform the design, evaluation, and reporting of devices and algorithms, while further highlighting key considerations for workflow integration and the implementation of decision-support systems. These findings have practical relevance for advancing continuity of rehabilitation across clinical, home, and community settings, and may help guide nurses in delivering continuous monitoring, personalized follow-up, and timely intervention, thereby improving the efficiency and accessibility of rehabilitation management.

J Med Internet Res 2026;28:e85596

doi:10.2196/85596

Keywords



Background

Parkinson disease (PD) is a common neurodegenerative disorder characterized by tremor, bradykinesia, rigidity, and postural or gait disturbances [1,2]. Its high prevalence and disability rates profoundly impair patients’ quality of life and impose substantial burdens on families and health care systems. Since 1990, the number of patients with PD has more than doubled—from approximately 2.8 million to over 6.2 million—and is projected to exceed 12 million by 2040 [3]. In China, accelerated population aging has led to a rising prevalence, with an estimated 1.37% among adults aged 60 years and older and more than 3.62 million patients nationwide [4,5]. Rehabilitation remains essential for improving motor function and slowing functional decline [6-8]. However, conventional face-to-face rehabilitation is time- and resource-intensive, often inaccessible in remote regions, and frequently associated with poor adherence [9,10].

In recent years, the rapid development and application of wearable devices have provided significant advantages for assessment and intervention in neurological and psychiatric disorders, such as stroke and depression [11,12]. Wearable devices encompass a wide range of intelligent electronic devices, including smartwatches, wristbands, insoles, exoskeletons, multisensor wearable devices, and electronic textiles [13-16]. These devices support continuous, real-time, and multidimensional monitoring during daily activities or rehabilitation training, capturing key motor indicators such as gait, tremor, joint mobility, postural control, motor coordination, and overall physical activity levels [17-19].

Against this backdrop, the rapid emergence of artificial intelligence (AI)–enabled wearable devices has created new opportunities for the rehabilitation of PD. In this review, AI-enabled wearable devices refer to noninvasive wearable technologies that incorporate machine learning (ML) or deep learning (DL) methods for data-driven analysis beyond rule-based approaches. Unlike traditional approaches that rely on manual observation or intermittent scale-based assessments, AI-enabled wearable devices enable the capture of large volumes of fine-grained data. Supported by ML and DL techniques, these data can be efficiently processed and intelligently interpreted within a data-driven rehabilitation framework, highlighting the importance of AI model interpretability for clinical and nursing decision-making [20-22]. Moreover, AI-enabled wearable devices facilitate continuous remote monitoring and long-term follow-up, making it possible to track disease progression and changes in motor function in real time within home-based rehabilitation settings. These data can then be leveraged by clinicians and rehabilitation teams to design more precise training programs [23,24]. Such an approach enhances the continuity and personalized rehabilitation, offers a feasible pathway for remote and home-based rehabilitation, and holds promising potential for optimizing health care resource allocation and improving patient adherence.

Research Questions and Objectives

Although numerous studies and reviews have explored the use of wearable devices in PD rehabilitation, several limitations remain, including (1) most studies involve small sample sizes and short follow-up periods, providing insufficient longitudinal evidence to evaluate long-term rehabilitation outcomes [17,25,26]; (2) current reviews mainly focus on device performance or symptom monitoring, with limited attention to AI data-processing workflows, algorithmic applications, and their potential value for clinical and nursing practice [18,19]; (3) studies differ substantially in device types, monitoring indicators, data-processing strategies, and AI algorithm selection. The absence of a unifying framework to synthesize these variations hinders a comprehensive understanding of the research landscape [27-29]; and (4) evidence in nursing practice remains limited; despite nurses’ central role in rehabilitation, research on AI-enabled tools for nursing decision-making, health education, and adherence support is still scarce [30,31]. Collectively, these issues result in fragmented and unsystematic evidence, limiting its ability to guide clinical rehabilitation nursing practice and theoretical development.

To address these needs, we conducted a scoping review and developed an evidence map to systematically present the current applications of AI-enabled wearable devices in PD rehabilitation and motor function assessment. The scoping review summarized the breadth, depth, and nature of existing evidence [32], while the evidence map visualizes research distribution, patterns, and emerging trends across domains [33]. By integrating these 2 approaches, this study aimed to map and visualize the current evidence on AI-enabled wearable devices for PD rehabilitation and motor function assessment. The objectives were to summarize device types, monitoring indicators, AI algorithms, and application characteristics, and to identify research gaps and future directions to support nursing practice and further digital-health implementation in PD rehabilitation.


Protocol and Registration

The review followed the methodological guidance outlined in the PRISMA-ScR (Preferred Reporting Items for Systematic Reviews and Meta-Analyses extension for Scoping Reviews) [34], and the completed checklist is provided in Multimedia Appendix 1. No protocol was registered for this review.

Eligibility Criteria

The inclusion criteria were defined using the population, concept, and context framework. Eligible studies enrolled participants with a confirmed diagnosis of PD and used noninvasive, body-worn wearable devices integrated with AI-based analytical methods. The focus of included studies was on rehabilitation, motor function assessment, or monitoring. Gray literature such as dissertations and full conference papers was included, whereas preprints and conference abstracts were excluded to ensure methodological consistency and to prioritize evidence that had undergone formal peer review. Only studies published in English or Chinese on or after January 1, 2020, were eligible for inclusion. Detailed inclusion and exclusion criteria are presented in Multimedia Appendix 2.

Information Sources

A comprehensive search was conducted on August 12, 2025, across 9 databases: CNKI (China National Knowledge Infrastructure), Wanfang Data, SinoMed, Cochrane Library, PubMed, Web of Science, CINAHL, Scopus, and Embase. The literature coverage spanned from database inception to the search date. No websites or other nondatabase information sources were searched. Reference lists and forward citations of the included studies were additionally screened to identify further potentially relevant records. To ensure the currency and completeness of the evidence, an updated search using the same strategy was conducted on December 10, 2025.

Search Strategy

The reporting of the search process adhered to the PRISMA-S (Preferred Reporting Items for Systematic Reviews and Meta-Analyses literature search extension) [35]. The search strategy was developed by a researcher trained in systematic literature retrieval and combined MeSH (Medical Subject Headings) terms with free-text keywords across four core concepts: (1) wearable or body-worn mobile devices; (2) rehabilitation, assessment, or monitoring; (3) PD; and (4) AI or ML. No limits or restrictions, including language, publication date, or study design, were applied at the search level. All eligibility criteria were applied manually during title/abstract screening and full-text review. The search strategy did not undergo peer review. Complete search strategies for each database and information source are provided in Multimedia Appendix 3.

Selection of Sources of Evidence

The selection process comprised two stages: (1) title/abstract screening and (2) full-text review. In the first stage, 1 reviewer imported all records into EndNote X9 (version 12062; Clarivate), removed duplicates, and excluded studies outside the eligible date range. Two reviewers then independently screened titles and abstracts to determine provisional inclusion. In the second stage, the same 2 reviewers independently assessed the full texts to establish the final set of included studies. Disagreements were resolved through discussion; unresolved cases were adjudicated by a third reviewer. Cohen κ was calculated to measure the interrater agreement [36]. Agreement was high for title and abstract screening (κ=0.86), whereas it was moderate for full-text review (κ=0.57).

Data Charting Process

Data charting was conducted independently by 1 reviewer using a standardized extraction form developed a priori. The form included key study characteristics, wearable device details, AI methods, outcomes, and findings relevant to the review questions. Following Joanna Briggs Institute guidance, charting was iterative, and the form was refined as needed. A second reviewer verified all data, with disagreements resolved by a third reviewer. The form was pilot-tested on 5 studies before full implementation.

Data Items

The data extraction form was developed by the research team based on the objectives of this review and included three core modules: (1) study details, (2) wearable device details, and (3) AI details. When the relevant information was not reported, it was recorded as not available. Each module contained specific data variables, as detailed in Multimedia Appendix 4.

Critical Appraisal of Individual Sources of Evidence

The methodological quality of all included studies was evaluated using the Mixed Methods Appraisal Tool, 2018 version [37]. Two reviewers conducted the assessments independently and resolved discrepancies through discussion; cases in which consensus could not be reached were adjudicated by a third reviewer. The appraisal was used descriptively to identify methodological strengths and limitations of the included studies and was not used as a criterion for study inclusion. The detailed appraisal results are provided in Multimedia Appendix 5 [38-103].

Synthesis of Results

A narrative synthesis approach was used. Extracted data were summarized thematically and presented in descriptive text, tables, and an evidence map to illustrate research trends, methodological characteristics, and application domains of AI-enabled wearable devices in PD.

Ethical Considerations

This scoping review used only published, peer-reviewed literature and therefore did not require approval from an ethics committee or institutional review board. Findings will be submitted to an open access, peer-reviewed journal and presented at relevant medical and engineering conferences.


Study Selection

A total of 1467 records were initially retrieved. Of these, 475 (32.38%) were removed using EndNote X9 due to duplication, publication date, or language incompatibility. The remaining 992 (67.62%) records underwent title and abstract screening, during which 877 (88.41%) were excluded. Among the 115 records reviewed in full text, 47 (40.87%) were unrelated to AI, 7 (6.09%) were unavailable in full text, and 2 (1.74%) lacked extractable data. An additional 7 relevant studies were identified through reference tracing. In total, 66 studies were included in this review [38-103]. The detailed selection process is illustrated in the PRISMA (Preferred Reporting Items for Systematic Reviews and Meta-Analyses) flow diagram (Figure 1).

Figure 1. Flowchart of the study selection process.

Characteristics of Included Studies

All included studies were published between 2020 and 2025, and the publication volume remained relatively stable over this 6-year period. The highest number of studies was published in 2025 (16/66, 24.24%; Table 1). The included studies originated from multiple countries and regions (Figure 2), with Switzerland contributing the largest proportion (27/66, 40.91%). With respect to publication type, the vast majority were journal articles (61/66, 92.42%).

Table 1. Characteristics of the included studies using artificial intelligence–enabled wearable devices for Parkinson disease (PD; N=66).
FeaturesValuesReferences
Year of publication, n (%)

20208 (12.12)[38,47,50,53,61,63,66,73]

202111 (16.67)[44,49,55,56,59,62,72,82,86,94,96]

20229 (13.64)[39,46,51,58,64,68,80,81,83]

202311 (16.67)[43,57,60,71,74,76,77,84,85,89,91]

202411 (16.67)[48,54,65,67,69,70,78,87,90,93,95]

202516 (24.24)[40-42,45,52,75,79,88,92,97-103]
Type of publication, n (%)

Journal article61 (92.42)[38,40-45,47-69,71-84,86-91,93-103]

Conference paper4 (6.06)[39,46,70,92]

Thesis1 (1.52)[85]
Country/region of publication, n (%)

Switzerland27 (40.91)[41-44,47,49,55,57,58,62,65-68,71,75,79,80,88,90,91,96,97,99,101-103]

United States19 (28.79)[39,46,48,51,54,59,61,64,69,70,73,74,76,78,81,86,89,92,94]

United Kingdom11 (16.67)[50,52,53,56,60,63,72,87,93,95,98]

Netherlands5 (7.58)[38,40,45,82,100]

China3 (4.55)[83-85]

Germany1 (1.52)[77]
Number of participants

Mean (SD; range)55.06 (132.78; 2-1057)[38-99,101-103]

1-30, n (%)41 (62.12)[38-42,44,46-49,52,55-64,67,69-75,82,85,86,90,91,93-97,99,103]

31-60, n (%)8 (12.12)[43,68,76-79,81,92]

61-100, n (%)9 (13.64)[45,50,51,53,65,66,83,89,101]

101-500, n (%)6 (9.09)[54,84,87,88,98,102]

>500, n (%)1 (1.52)[80]

Not reported1 (1.52)[100]
Age distribution of participantsa (years)

Meanb (SD; range)64.59 (10.03; 25.00-76.10)[38-47,49-53,56-61,63-76,80,81,83-85,89,91-93,95,96,98,99,101-103]

<60, n (%)3 (4.55)[39,40,55]

60-69, n (%)39 (59.09)[42-45,47,49,51,53,56,63,65-67,69-77,79-81,83,85,86,88,89,91-93,95,96,98,99,102,103]

70-79, n (%)14 (21.21)[38,41,46,50,52,57-61,64,68,84,101]

≥80, n (%)2 (3.03)[54,97]

Not reported8 (12.12)[48,62,78,82,87,90,94,100]
Sex distributionc

Female (%), mean (SD; range)35.88 (18.08; 0–100)[38-41,45-48,50-68,70-73,75,79-81,83-92,94-96,98,99,101,102]
Inclusion details n (%)

PD only47 (71.21)[38,40,41,43,46-49,51,52,54-72,76,81,82,84-88,90,93-99,102,103]

Mixed18 (27.27)[39,42,44,45,50,53,73-75,77-80,83,89,91,92,101]

Not reported1 (1.52)[100]
Application objectived, n (%)

Motor function assessment54 (81.82)[38,39,41-47,49,50,52-55,57-80,82-89,91,92,95,99-102]

Disease progression/symptom monitoring26 (39.39)[39,46,51-53,55,56,65,68,69,71,74,76,79,87,89,90,93,94,96-100,102,103]

Efficacy evaluation11 (16.67)[46,54,56,61,70,71,77,78,81,90,103]

Rehabilitation training8 (12.12)[40,48,54,55,60,67,77,96]

Others1 (1.52) [88]

aReported as mean (SD), median (IQR), or range, depending on the study.

bAge was summarized from studies reporting mean (SD); studies with only median or without age were excluded. For mixed populations, only PD patient data were used.

cBased only on studies that reported sex data.

dApplication objectives were classified based on the primary functional aim of the wearable system. Motor function assessment refers to quantitative evaluation of motor performance at specific time points; disease progression/symptom monitoring refers to continuous or longitudinal tracking of disease states; efficacy evaluation refers to the assessment of responses to interventions or treatments; and rehabilitation training refers to systems explicitly designed to support training or feedback to improve motor function. The number of studies does not add up, as several studies have used >1 application objective.

Figure 2. Geographical distribution of the included studies using artificial intelligence–enabled wearable devices for Parkinson disease.

Regarding study populations, sample sizes ranged from 2 to 1057 participants, with a mean sample size of 55.06 (SD 132.78), yielding a total of approximately 3579 participants. Overall, most studies involved relatively small samples with substantial heterogeneity, and the majority included only patients with PD (47/66, 71.21%). In terms of the application objectives of wearable devices, most studies focused on motor function assessment (54/66, 81.82%), whereas rehabilitation-oriented applications were relatively scarce, being reported in only 8 (12.12%) studies. Overall, this distribution suggests that intervention-focused research remains limited. Detailed characteristics of the included studies are presented in Table 1 and Multimedia Appendix 6 [38-103]. For categories in which studies could contribute to more than 1 classification, percentages may exceed 100%; detailed coding rules are provided in the table footnotes.

Characteristics of Wearable Devices

A total of 5 major types of wearable devices were identified across the included studies. Sensor module was the predominant device type, accounting for 84.85% (56/66) of studies, followed by smart insoles (7/66, 10.61%) and smartwatches (5/66, 7.58%; Table 2). Regarding device origin, commercially available devices were more common (41/66, 62.12%) than noncommercial devices (21/66, 31.82%); nevertheless, a substantial proportion of studies still relied on customized or research-grade devices. A total of 14 distinct wearing locations were reported (Figure 3). The most common placements were the shank (25/66, 37.88%) and the wrist (23/66, 34.85%), followed by the foot (19/66, 28.79%) and the ankle (16/66, 24.24%). Many studies used multiple wearing locations within a single study, reflecting the diversity of measurement strategies adopted.

Table 2. Features of artificial intelligence–enabled WDsa used in the included Parkinson disease studies (N=66).
FeaturesValues, n (%)References
Type of WDa,b

Sensor module56 (84.85)[38-45,47,48,50-75,78-82,84,86-89,91-98,102,103]

Smart insole7 (10.61)[39,46,55,57,59,100,101]

Smartwatch5 (7.58)[49,63,85,97,99]

Smart wristband3 (4.55)[83,85,90]

Stimulator2 (3.03)[48,103]

Others (adhesive electrodes, ankle band, or smart glove)3 (4.55)[42,76,77]
Status of WDa,b

Commercial41 (62.12)[38,42,44,46,48,49,51-59,63-67,70-72,74,75,80,81,84-89,91-93,97-99,101,102]

Noncommercial21 (31.82)[39-41,43,45,47,48,50,60,67-69,73,76-78,83,84,90,100,103]

Not reported8 (12.12)[61,62,79,82,85,94-96]
Company of WDa,b

APDM Inc6 (9.09)[53,62,64,66,86,87]

Shimmer Sensing5 (7.58)[38,55,59,85,93]

Activinsights Ltd4 (6.06)[43,44,85,97]

MC10 Inc3 (4.55)[53,63,75]

Moticon ReGo AG3 (4.55)[46,57,101]

Noraxon USA Inc2 (3.03)[42,74]

PD Neurotechnology Ltd2 (3.03)[57,71]

Great Lakes NeuroTechnologies Inc2 (3.03)[56,70]

STMicroelectronics2 (3.03)[68,69]

Tekscan Inc2 (3.03)[55,59]

Self-developed13 (19.70)[39,51,60,61,67,73,76,77,83,84,89,100,103]

Others16 (24.24)[43,44,48,50,54,58,63,65,88-92,96,99,102]

Not reported16 (24.24)[40,41,45,47,49,52,72,78-82,94,95,97,98]
Placementb

Shank25 (37.88)[40,42-44,48,52,55,59,60,67,69,71,74,75,80-82,84,86,88,91,93-95,102]

Wrist23 (34.85)[38,44,49,53,56-58,63,69-71,79,83,85,87-92,97,99,102]

Foot19 (28.79)[39,40,46,48,55,57,59,62,72-74,87,88,91,96,100-103]

Ankle16 (24.24)[38,39,41,45,51,52,61,64,66,70,76,83,93,98,99,103]

Waist13 (19.70)[44,47,52,54,57,65,69,71,86,88,89,97,102]

Trunk12 (18.18)[45,66,74,75,78,81,84,87,88,92,98,102]

Arm9 (13.64)[45,74,75,78,81,84,85,97,98]

Thigh7 (10.61)[55,59,68,74,88,95,102]

Hip6 (9.09)[62,74,78,82,93,98]

Hand3 (4.55)[63,77,91]

Others (Neck, knees, back, or head)4 (6.06)[50,51,74,95]
Compatibility with OSc,d

Local logger16 (24.24)[38,40,43,44,46,50,54,57,58,60,61,69,73,87,88,93]

Android6 (9.09)[64,65,76,85,90,97]

iOS5 (7.58)[48,51,63,65,90]

Windows3 (4.55)[68,83,84]

Not applicable7 (10.61)[39,52,67,71,94,95,103]

Not reported31 (46.97)[41,42,45,47,49,53,55,56,59,62,66,70,72,74,75,77-82,86,89,91,92,96,98-102]
Gatewaye

PCf10 (15.15)[52,68,73,74,76,77,80,86,90,102]

Smartphone6 (9.09)[39,48,64,76,85,90]

Tablet3 (4.55)[51,80,92]

IoTg Gateway2 (3.03)[82,103]

Not reported48 (72.73)[38,40-47,49,50,53-63,65-67,69-72,75,78,79,81,83,84,87-89,91,93-101]
Hosth

PCf35 (53.03)[38,41,43,44,52,54-59,61,66,68,69,73-77,80,82-84,86-89,91,93,94,96,98,102,103]

Server6 (9.09)[39,48,58,71,90,97]

Smartphone5 (7.58)[48,64,65,76,85]

On-device4 (6.06)[40,60,65,67]

Tablet3 (4.55)[51,80,92]

Not reported18 (27.27)[42,45-47,49,50,53,62,63,70,72,78,79,81,95,99-101]
Mode of data transferi

Bluetooth15 (22.73)[39,40,48,51,52,64,65,68,76,80,84,86,88,89,102]

Internet8 (12.12)[39,48,58,71,83,85,90,103]

Removable media6 (9.09)[38,40,43,48,61,69]

Wired5 (7.58)[40,44,73,77,103]

Not reported39 (59.09)[41,42,45-47,49,50,53-57,59,60,62,63,66,67,70,72,74,75,78,79,81,82,84,87,91-101]
Sensors in the wearablesj

Accelerometer61 (92.42)[38-54,57-59,61-99,102,103]

Gyroscope50 (75.76)[38-43,45-49,51,52,54-57,59-64,66-74,76,78,80,81,83,84,86-89,92-94,96,98,99,102,103]

Magnetometer11 (16.67)[45,52,53,66,71,76,78,83,86,87,94]

Pressure sensor9 (13.64)[39,40,46,55,57,59,77,100,101]

sEMGk sensor3 (4.55)[40,42,78]

Flex sensor2 (3.03)[40,77]
Measured biosignalsl

Acceleration61 (92.42)[38-54,57-59,61-99,102,103]

Angular velocity50 (75.76)[38-43,45-49,51,52,54-57,59-64,66-74,76,78,80,81,83,84,86-89,92-94,96,98,99,102,103]

Magnetic field signals11 (16.67)[45,52,53,66,71,76,78,83,86,87,94]

Pressure/mechanical signals9 (13.64)[39,40,46,55,57,59,77,100,101]

EMGm signals3 (4.55)[40,42,78]

Bending/Flex sensing2 (3.03)[40,77]
Sensing type

Passive52 (78.79)[38,40,41,44-60,62,65,71-73,75-79,82-103]

Active14 (21.21)[39,42,43,61,63,64,66-70,74,80,81]
Application scenario

Clinical31 (46.97)[38,48,51,54,55,57,59-61,63,66-68,74,76-78,80,81,83,84,86,88,89,92-96,98,99]

Home12 (18.18)[44,47,58,65,71,72,79,82,97,100-102]

Clinical and Home23 (34.85)[39-43,45,46,49,50,52,53,56,62,64,69,70,73,75,85,87,90,91,103]
Duration of monitoring/intervention

Single session32 (48.48)[38,40-44,46,48-50,52,55,64,67-69,73,74,76,77,79,83-85,89-91,96,99-102]

Multiple sessions20 (30.30)[47,51,53,54,56,57,60,61,63,66,70,75,80,92-95,97,98,103]

Long-term monitoring9 (13.64)[39,45,58,62,65,71,72,81,86]

Not reported5 (7.58) [59,78,82,87,88]

aWD: wearable device.

bThe number of studies does not add up, as several studies have used >1 wearable device.

cOS: operating system.

dThe number of studies does not add up, as several studies have used >1 wearable device, and many wearable devices are compatible with >1 operating system.

eThe number of studies does not add up, as several studies used >1 wearable device, and many wearable devices used >1 gateway.

fPC: personal computer.

gIoT: Internet of Things.

hThe number of studies does not add up, as several studies used >1 wearable device, and many wearable devices used >1 host.

iThe number of studies does not add up, as several studies used >1 wearable device, and many wearable devices used >1 mode of data transfer.

jThe number of studies does not add up, as several studies used >1 wearable device, and most wearable devices have >1 sensor.

ksEMG: surface electromyography.

lThe number of studies does not add up, as several studies used >1 wearable device, and most wearable devices assess >1 biosignal.

mEMG: electromyography.

Figure 3. Placement of wearable sensors in studies using artificial intelligence–enabled wearable devices for Parkinson disease.

With regard to sensor configuration, devices integrating multiple sensors predominated, accounting for 75.76% (50/66) of the included studies. The collected biosignals were predominantly inertial signals. Accelerometer data were acquired in the vast majority of studies (61/66, 92.42%), followed by gyroscope data (50/66, 75.76%), whereas magnetic field, pressure, and electromyography signals were used less frequently (Table 2).

With respect to application settings, wearable devices were more frequently deployed in clinical environments (31/66, 46.97%), with an additional 34.85% (23/66) of studies spanning both clinical and home environments, whereas studies conducted exclusively in home settings were relatively uncommon (12/66, 18.18%). Regarding the duration of monitoring or intervention, most studies used either single-session assessments (32/66, 48.48%) or repeated short-term testing designs (20/66, 30.30%), while studies involving long-term continuous monitoring were comparatively limited (9/66, 13.64%). Overall, wearable device research in PD is dominated by sensor module devices, with devices most commonly worn on the lower limbs and wrist. Data collection is primarily passive, and studies largely focus on short-term applications conducted in clinical settings. Additional technical characteristics of the wearable devices are provided in Table 2 and Multimedia Appendix 7 [38-103].

AI Algorithm Characteristics of Wearable Devices

The AI applications in the included studies were categorized into four purposes: (1) monitoring or assessment (35/66, 53.03%), (2) state recognition or functional screening (17/66, 25.76%), (3) prediction (8/66, 12.12%), and (4) rehabilitation and feedback (6/66, 9.09%; Table 3). Among the studies, 62.12% (41/66) used ML algorithms only, 28.79% (19/66) used DL algorithms only, and 9.09% (6/66) combined ML and DL approaches. These studies used algorithms to address classification problems (59/66, 89.39%), regression problems (15/66, 22.73%), and clustering problems (3/66, 4.55%).

Table 3. Features of artificial intelligence (AI) algorithms used in the included Parkinson disease wearable-device studies (N=66).
FeaturesValues, n (%)References
AI category

MLa41 (62.12)[38,39,43,45,48,50,53-55,57-63,65,66,68,71-77,80-84,86-88,92,96,98,99,101-103]

DLb19 (28.79)[41,42,44,46,49,51,52,56,64,67,69,70,78,79,89-91,94,100]

ML and DL6 (9.09)[40,47,85,93,95,97]
Task typec

Classification59 (89.39)[39-55,57-60,62-67,69,71,72,74-79,81-103]

Regression15 (22.73)[38,43-45,53,56,61,68-71,77,80,98,102]

Clustering3 (4.55)[73,77,98]
AI algorithmd

Support vector machine28 (42.42)[38,40,43,45,47,50,53,57,61,65,66,68,74,76,78,81-85,87,88,92,93,96,101-103]

Convolutional neural network23 (34.85)[40-42,44,46,47,49,51,52,54,56,67,69,70,78,85,89-91,93,95,97,100]

Random forest19 (28.79)[39,45,47,53,57,60,63,79,80,82,83,86,87,92,95,97-99,101]

Logistic regression15 (22.73)[39,43,53,61,62,65,71,78-80,87,88,92,98,101]

Decision tree14 (21.21)[45,47-50,55,57,59,61,62,76,82,86,88]

Long short-term memory11 (16.67)[42,47,52,56,58,69,70,85,93,94,100]

Multilayer perceptron8 (12.12)[40,41,44,77,78,88,92,100]

k-nearest neighbors7 (10.61)[50,76,83,88,92,93,101]

Naive Bayes5 (7.58)[43,50,76,88,101]

Extreme gradient boosting4 (6.06)[49,87,93,102]

Gradient boosting3 (4.55)[39,57,86]

Self-attention mechanism3 (4.55)[42,69,85]

Hidden Markov model3 (4.55)[58,72,73]

k-means clustering3 (4.55)[75,77,98]

Elastic net3 (4.55)[80,87,98]

AdaBoost3 (4.55)[76,92,96]

Radial basis function2 (3.03)[78,79]

Others12 (18.18)[43,45,47,64,67,78-80,82,86,97,102]
Aim of AI algorithm

Monitoring or assessment35 (53.03)[38,39,42-47,50,53,56-59,61,63,65,70-73,79,81,83-85,88,90-93,97-99,102]

State recognition or functional screening17 (25.76)[41,49,51,54,62,66,74-78,82,89,94,100,101,103]

Prediction8 (12.12)[52,55,64,68,69,80,87,95]

Rehabilitation and feedback6 (9.09)[40,48,60,67,86,96]
Validation approache

Leave-one-out cross-validation37 (56.06)[39,41,42,45-47,49,50,53-55,57-63,65-68,70,73,74,76,78,79,81,83,86,88,90,91,93,102,103]

k-fold cross-validation27 (40.91)[38-41,47,49-52,54,57,69,72,76,77,84,85,87-92,99,101,103]

Hold-out validation18 (27.27)[40,44,48,49,56,61,64,66,76,78,80,82,88,92,94,100-102]

External validation4 (6.06)[43,60,69,71]

Not reported5 (7.58)[75,95-98]
Performance measuresf

Sensitivity43 (65.15)[39-42,46,47,49-55,57-60,62,64-67,69,72,73,75-79,82-85,87,90,94-97,99,101,103]

Accuracy41 (62.12)[39-43,45,46,48,50-55,57,58,62,65,67,69,74-77,82-87,89-92,95-98,100-102]

Specificity27 (40.91)[46,47,49,50,52-55,59,60,62,64-67,69,76,78,79,82,83,94-96,99,101,103]

F1-score27 (40.91)[39-41,49,51-55,57,67,73,75-78,82,84,85,87,92,93,95,99,101-103]

Area under the curve25 (37.88)[39-41,47,52,55,57,58,62-65,69,74,78,79,82,86-88,92,94,96,98,101]

Precision24 (36.36)[39-42,46,51,53,54,58,59,62,65-67,75-78,84,85,87,92,95,97]

Correlation coefficient14 (21.21)[38,44,53,56,58,61,68,70,71,79-81,88,102]

Receiver operating characteristic7 (10.61)[60,63,66,74,89,94,98]

Mean absolute error7 (10.61)[43,44,56,68,70,98,102]

Root-mean-square error6 (9.09)[38,53,61,68,80,102]

Intraclass correlation coefficient4 (6.06)[38,54,77,93]

Cohen κ3 (3.03)[77,95,102]

Equal error rate2 (3.03)[47,52]

G-mean2 (3.03)[51,52]

Negative predictive value2 (3.03)[52,65]

Coefficient of determination2 (3.03)[70,98]

Others9 (13.64) [38,52,54,59,66,67,82,88,99]

aML: machine learning.

bDL: deep learning.

cThe number of studies does not add up, as many studies have used >1 task type.

dThe number of studies does not add up, as many studies have used >1 AI algorithm.

eThe number of studies does not add up, as many studies have used >1 validation approach.

fThe number of studies does not add up, as most studies used >1 performance measure.

A total of 30 distinct AI algorithms were identified across the included studies. The most frequently used were support vector machines (SVMs; 28/66, 42.42%), followed by convolutional neural networks (CNNs; 23/66, 34.85%), random forests (19/66, 28.79%), logistic regression (15/66, 22.73%), decision trees (14/66, 21.21%), long short-term memory (LSTM) networks (11/66, 16.67%), and multilayer perceptrons (8/66, 12.12%; Table 3).

As shown in Figure 4, SVMs and CNNs were predominantly applied in the “monitoring or assessment” and “state recognition or functional screening” categories. Random forests, decision trees, and logistic regression models were distributed across application purposes but were mainly used for monitoring- or assessment-related tasks. Temporal and probabilistic models, such as LSTM networks and hidden Markov models, were more often observed in state recognition or functional screening tasks, although their overall adoption remained limited. In contrast, algorithmic applications in the “prediction” and “rehabilitation or feedback” categories were comparatively sparse, indicating that prospective and closed-loop applications remain at an early stage of development.

Figure 4. Distribution of artificial intelligence (AI) algorithms and application objectives in the included Parkinson disease wearable-device studies.

A total of 4 different model validation approaches were used across the included studies, with approximately 31.82% (21/66) using more than 1 method (Table 3). Leave-one-out cross-validation was the most frequently used approach (37/66, 56.06%), followed by k-fold cross-validation (27/66, 40.91%), hold-out validation (18/66, 27.27%), and external validation (4/66, 6.06%). A total of 25 performance metrics were used to evaluate model performance. The most commonly reported metrics were sensitivity (43/66, 65.15%), accuracy (41/66, 62.12%), specificity (27/66, 40.91%), F1-score (27/66, 40.91%), area under the curve (25/66, 37.88%), precision (24/66, 36.36%), and correlation coefficient (14/66, 21.21%). Detailed characteristics of the AI algorithms applied in each study are provided in Multimedia Appendix 8 [38-103].

Challenges and Opportunities of Wearable Devices

Based on a synthesis of the included evidence, we developed a conceptually integrated “challenges and opportunities” framework grounded in the evidence base. This framework categorizes the challenges and opportunities of AI-enabled wearable devices in clinical applications into six dimensions: (1) evidence and data quality; (2) technical limitations; (3) usability, adherence, and equity; (4) economic and policy barriers; (5) privacy, security, and data governance; and (6) clinical translation and workflow integration. The framework is intended to inform future research and practice rather than to represent an empirically validated implementation pathway, as illustrated in Figure 5 and detailed in Multimedia Appendix 9.

Figure 5. From monitoring to closed-loop rehabilitation: key opportunity domains for artificial intelligence (AI)–enabled wearable devices in Parkinson disease.

Principal Findings

Overview

We synthesized the evidence on AI-enabled wearable devices for PD rehabilitation and motor function assessment, focusing on device types, monitoring indicators, algorithmic approaches, and application characteristics. Overall, the available evidence remains predominantly monitoring- and assessment-oriented, with relatively limited rehabilitation interventions and workflow-integrated applications. Key translational gaps persist, including limited evidence from real-world home/community settings, scarce external validation, underreporting of model calibration and clinical use, and underrepresentation of diverse populations.

Most prior reviews were published between 2020 and 2023 [104-106], and may not fully capture more recent developments, including the wider adoption of remote monitoring devices [107,108], rapid advances in commercial multisensor systems [109], and the evolution of AI models [110]. In contrast to earlier reviews that primarily focused on diagnosis or device technical performance, the present review adopts a rehabilitation- and nursing-oriented perspective and foregrounds evidence relevant to real-world implementation and nurse-led translation. Building on these findings, we propose a conceptual closed-loop framework linking monitoring, assessment, intervention, feedback, and reevaluation, which can be tested and refined in future research.

Study Characteristics

This review covered the 2020-2025 publication window and found that the volume of studies remained consistently high in recent years, suggesting that the field has entered a relatively stable phase of development rather than a period of rapid expansion, consistent with previous reviews [106]. Nevertheless, notable limitations in methodological rigor and sampling design persist: (1) sample sizes varied substantially, with mean values often influenced by outliers [48,80]; (2) sampling was typically based on single-center convenience cohorts [41,64,96]; and (3) subgroup analyses by sex, age, or special populations (such as early-onset or comorbid PD) were largely absent [41,83,84,96]. Many studies enrolled only people with PD, with relatively few including healthy control groups. Outcome reporting tended to prioritize discrimination metrics (eg, accuracy), while model calibration, uncertainty quantification, and clinically actionable thresholding or decision-curve analysis were rarely reported [57,63,64,72,94-96]. In addition, research has been concentrated in Europe and North America (Table 1), with comparatively limited evidence from low- and middle-income countries and insufficient attention to home/community settings, thereby constraining generalizability and real-world applicability. Overall, persistent methodological limitations—including small, single-center samples, predominantly internal validation, and limited evidence from home/community settings—suggest that growth in study quantity has outpaced advances in translational value.

Application Characteristics and Signal Focus

In terms of application characteristics, AI-enabled wearable devices in PD show a marked imbalance between assessment-oriented and rehabilitation-oriented work, consistent with prior systematic reviews describing the field as “detection-heavy but prediction- and intervention-light” [104-106]. Although several representative studies have demonstrated the feasibility of continuous sensor-based quantification of tremor and bradykinesia [53], objective quantification of medication response [38], and sensor-based early detection of freezing of gait [54,55], most remain at a proof-of-concept stage, with limited evidence for real-world, workflow-integrated, or actionable use in routine care [104,105]. This disparity likely reflects inherent differences in methodological complexity. Assessment-focused studies typically rely on passive data capture and brief, structured tasks that are easier to implement in controlled settings, whereas rehabilitation studies require longer follow-up, personalized feedback, adherence management, and integration with clinical workflows—substantially raising both technical and methodological demands [13,111]. Moreover, most studies adopt motion capture–oriented wearable devices centered on inertial and/or pressure signals, with sensors preferentially placed on the lower limbs or the wrist. This design choice reduces data-collection burden and improves task reproducibility, thereby reinforcing assessment paradigms built around controlled tasks; however, it also structurally steers research toward motor outcomes (eg, gait, freezing of gait, tremor, and bradykinesia), making sleep, mood, and cognition more difficult to capture continuously and comprehensively [38,46,53]. Meanwhile, many applications still rely on short, structured assessments, which further constrain the continuous capture of nonmotor domains in real-world contexts [16,19]. Taken together, the evidence continues to prioritize motor-domain tasks, with insufficient coverage of nonmotor symptoms and their continuous assessment in real-world contexts. As a result, current applications fall short of meeting the comprehensive needs of long-term chronic disease management.

AI and Algorithmic Characteristics

In our review, SVMs and CNNs are the most frequently used approaches, while decision trees and logistic regression are commonly used as interpretable baseline models. Currently, the field reflects a pattern of “traditional ML dominance, parallel advances in convolutional DL, and early-stage development of sequential models and emerging paradigms” [38,40,50,57,68,74], which aligns with the ML/DL distribution observed in our results. This distribution may be attributable to differences in how algorithms adapt to data scale and structure. SVMs tend to provide more stable performance in small-sample, feature-engineered wearable data settings, whereas CNNs are better suited to end-to-end feature learning from high-frequency inertial signals, thereby reducing reliance on manual feature engineering [112,113]. Sequential approaches (eg, LSTM networks and a small number of HMMs) have been explored for gait-phase and state-recognition tasks, but remain relatively uncommon. This suggests that temporal dependencies and state-transition dynamics are still underutilized [36,52,72,73]. Consistent with these constraints, validation remains largely internal (eg, cross-validation), while independent external validation is rare, limiting confidence in generalizability [105]. More complex architectures (eg, self-attention mechanisms and extreme gradient boosting) also appear underused, potentially reflecting feasibility constraints for on-device or real-world deployment [39,75,80,92-96,98].

In summary, AI-enabled wearable devices for PD rehabilitation and assessment remain characterized by a strong emphasis on monitoring, with comparatively weak integration into closed-loop systems. From the perspective of the JMIR mHealth and uHealth taxonomy, most devices included in this review are better characterized as mobile health tools, primarily focused on individual-level monitoring and assessment [114]. Although these functions provide valuable objective data, there is still insufficient evidence that they have been systematically embedded into routine nurse-led rehabilitation workflows or broader health care delivery systems as mature eHealth solutions. Building on these findings, future research should move beyond stand-alone monitoring applications toward workflow-integrated eHealth systems with decision-support functions that can be tested in nurse-led rehabilitation pathways.

Practical Recommendations

Implications for Clinical and Nursing Practice

In both clinical and home-based rehabilitation contexts, nurses increasingly require continuous and objective patient information to enhance the precision of assessment and monitoring, and strengthen patient engagement in the rehabilitation process [115]. AI-enabled wearable devices represent a promising approach to addressing this need. Within clinical rehabilitation settings, continuous monitoring of gait, tremor, sleep, and activity levels can augment routine nursing surveillance, facilitate earlier detection of functional deterioration, and support timely intervention or prioritization—moving beyond the inherently fragmented nature of episodic in-clinic observations [25,115]. In remote or home-based rehabilitation, the continuous and objective data generated by wearables can be used to evaluate therapeutic response and adherence, while personalized feedback, symptom prompts, medication-taking support, and nurse-led follow-up via telerehabilitation platforms promote patient self-management [26,115]. Specifically, when algorithmic detection identifies abnormal changes in patients’ daily activity levels or tremor severity that exceed an individualized threshold, an alert can be triggered on the nursing interface. This may prompt nurses to rapidly assess symptom changes, medication adherence, and potential adverse effects and, as appropriate, arrange earlier review, reinforce medication and safety guidance, or adjust follow-up plans [97,115]. Likewise, when long-term monitoring trend reports indicate sustained deterioration in motor function–related indicators or marked fluctuations in motor symptoms, these data can provide objective evidence for the rehabilitation team to deliver targeted health education, adjust the frequency of follow-up, and, when needed, coordinate involvement of rehabilitation therapists or a multidisciplinary team [56,89,99]. These functions align closely with core nursing responsibilities in patient education, symptom monitoring, and long-term chronic disease management, and may help reinforce the full nursing care cycle of assessment, diagnosis, planning, implementation, and evaluation [116].

Priorities for Future Research and Translation

To advance the field, future research may consider five priority areas. (1) Strengthening the evidence base through multicenter, long-term real-world studies with preregistration, appropriate sample size estimation, and attrition management, alongside reporting of external validation, model calibration, uncertainty, and clinical use [117-120]. (2) Enhancing model feasibility through multimodal signal fusion and data-efficient learning strategies to support practical real-world deployment [104]. (3) Exploring conceptual closed-loop rehabilitation frameworks that integrate assessment and intervention, while examining the feasibility of actionable thresholds and their potential integration into routine clinical workflows [120]. (4) Improving device comfort and usability to balance patient adherence with caregiver and clinician workload, address the digital divide, and enhance accessibility [119,120]. (5) Strengthening privacy and data security frameworks, alongside improving insurance reimbursement mechanisms and regulatory support, to promote responsible cross-institutional data sharing [121,122], including explicitly addressing equity, responsible data use, and related ethical considerations in real-world nursing and rehabilitation contexts.

Limitations

This study has several limitations. First, no prospective protocol was registered; however, this is an acceptable practice for scoping reviews, and we clarify this point to enhance methodological transparency. Second, although a systematic search and screening process was conducted, the scope of databases, search strategies, and inclusion criteria may have led to the omission of eligible studies, posing a risk of incompleteness or potential bias. In addition, most included studies had small sample sizes, single-center designs, short follow-up durations, and limited external validation, alongside substantial heterogeneity across study populations, task paradigms, signal sources, algorithmic approaches, and outcome measures. This heterogeneity may limit cross-study comparability and the generalizability of findings. Given the predominantly exploratory and proof-of-concept nature of the included studies, the closed-loop rehabilitation framework and related recommendations proposed in this review should be interpreted as conceptual and practice-informing rather than empirically validated implementation pathways. Finally, the search was restricted to English and Chinese peer-reviewed publications, with limited inclusion of gray literature or other languages, which may introduce time-lag, linguistic, and publication biases. In particular, studies reporting null or negative findings, early-stage failures, or limited model performance may be less likely to be published and therefore underrepresented, potentially leading to an overestimation of the maturity, effectiveness, and translational readiness of AI-enabled wearable technologies.

Conclusions

This study systematically summarized recent evidence on AI-enabled wearable devices for motor function assessment and rehabilitation in people with PD. By combining a scoping review with an evidence map and adopting a rehabilitation-oriented perspective, we synthesized evidence across multiple dimensions, including application aims, sensing modalities, AI methods, and validation practices, thereby identifying key translational gaps between proof-of-concept studies and real-world rehabilitation workflows. Compared with previous reviews that primarily focused on monitoring functions or device performance, this study places greater emphasis on rehabilitation applications and nurse-led translation into practice, and further clarifies the key actionable gaps that limit real-world implementation. On this basis, a conceptually integrated “challenges and opportunities” framework is proposed to inform the design, evaluation, and reporting of devices and algorithms, and to highlight considerations for developing workflow-integrated, decision-support wearable systems. From a real-world perspective, these findings may support continuity of rehabilitation across clinical, home, and community settings by enabling nurses to deliver continuous monitoring, personalized follow-up, and timely intervention, ultimately improving the efficiency and accessibility of PD rehabilitation management.

Acknowledgments

Generative artificial intelligence tools (ChatGPT [OpenAI] and DeepSeek [Hangzhou DeepSeek Artificial Intelligence Co, Ltd]) were used solely under full human supervision to assist with proofreading/editing and translation, including terminology conversion, textual clarification, and translation of reviewer comments. These tools were not used for study design, data processing, analysis, or the generation of scientific content. All AI-assisted outputs were reviewed and revised by the authors, who take full responsibility for the accuracy and integrity of the manuscript.

Data Availability

This scoping review did not generate any primary data. All data extracted and synthesized in this review were obtained from previously published studies, which are publicly available. No additional datasets were created or analyzed.

Conflicts of Interest

None declared.

Multimedia Appendix 1

PRISMA-ScR checklist.

DOCX File , 71 KB

Multimedia Appendix 2

Eligibility criteria.

DOCX File , 19 KB

Multimedia Appendix 3

Search strategy.

DOCX File , 30 KB

Multimedia Appendix 4

Extracted data form.

DOCX File , 19 KB

Multimedia Appendix 5

Quality assessment of the included studies.

DOCX File , 72 KB

Multimedia Appendix 6

Detailed characteristics of included studies using artificial intelligence–enabled wearable devices in Parkinson disease research.

DOCX File , 50 KB

Multimedia Appendix 7

Detailed characteristics of artificial intelligence–enabled wearable devices used in the included Parkinson disease studies.

DOCX File , 59 KB

Multimedia Appendix 8

Detailed characteristics of artificial intelligence algorithms used in the included wearable device studies for Parkinson disease.

DOCX File , 50 KB

Multimedia Appendix 9

Challenges and opportunities of artificial intelligence–enabled wearable devices in Parkinson disease research.

DOCX File , 20 KB

  1. Kouli A, Torsney KM, Kuan WL. Parkinson’s disease: etiology, neuropathology, and pathogenesis. In: Stoker TB, Greenland JC, editors. Parkinson’s Disease: Pathogenesis and Clinical Aspects Internet. Brisbane (AU). Codon Publications; 2018.
  2. Abusrair AH, Elsekaily W, Bohlega S. Tremor in Parkinson's disease: from pathophysiology to advanced therapies. Tremor Other Hyperkinet Mov (NY). 2022;12:29. [FREE Full text] [CrossRef] [Medline]
  3. Dorsey ER, Sherer T, Okun MS, Bloem BR. The emerging evidence of the Parkinson pandemic. J Parkinsons Dis. 2018;8(s1):S3-S8. [FREE Full text] [CrossRef] [Medline]
  4. Qi S, Yin P, Wang L, Qu M, Kan GL, Zhang H, et al. Prevalence of Parkinson's disease: a community-based study in China. Mov Disord. 2021;36(12):2940-2944. [CrossRef] [Medline]
  5. Zhang ST, Jia XF, Zhang JG. Research progress on the association between dietary nutrition, lifestyle, and Parkinson's disease [in Chinese]. J Environ Occup Med. 2025;40(2):149-155. [FREE Full text]
  6. Garg D, Dhamija RK. Rehabilitation in Parkinson's disease: current status and future directions. Ann Mov Disord. 2020;3(2):79-85. [CrossRef]
  7. Yoon SY. Update on Parkinson's disease rehabilitation. Brain Neurorehabil. 2022;15(2):e15. [FREE Full text] [CrossRef] [Medline]
  8. Zhang L, Liu Y, Han W. Advance of rehabilitation for Parkinson's disease [in Chinese]. Chin J Contemp Neurol Neurosurg. 2017;17(5):328-332. [FREE Full text]
  9. Pastana Ramos LF, Vilacorta-Pereira TDCS, Duarte JDS, Yamada ES, Santos-Lobato BL. Feasibility and effectiveness of a remote individual rehabilitation program for people with Parkinson's disease living in the Brazilian Amazon: a randomized clinical trial. Front Neurol. 2023;14:1244661. [FREE Full text] [CrossRef] [Medline]
  10. Sevcenko K, Lindgren I. The effects of virtual reality training in stroke and Parkinson's disease rehabilitation: a systematic review and a perspective on usability. Eur Rev Aging Phys Act. 2022;19(1):4. [FREE Full text] [CrossRef] [Medline]
  11. Fedor S, Lewis R, Pedrelli P, Mischoulon D, Curtiss J, Picard RW. Wearable technology in clinical practice for depressive disorder. N Engl J Med. 2023;389(26):2457-2466. [CrossRef] [Medline]
  12. Maceira-Elvira P, Popa T, Schmid AC, Hummel FC. Wearable technology in stroke rehabilitation: towards improved diagnosis and treatment of upper-limb motor impairment. J Neuroeng Rehabil. 2019;16(1):142. [FREE Full text] [CrossRef] [Medline]
  13. Porciuncula F, Roto AV, Kumar D, Davis I, Roy S, Walsh CJ, et al. Wearable movement sensors for rehabilitation: a focused review of technological and clinical advances. PM R. 2018;10(9 Suppl 2):S220-S232. [FREE Full text] [CrossRef] [Medline]
  14. Rodríguez-Fernández A, Lobo-Prat J, Font-Llagunes JM. Systematic review on wearable lower-limb exoskeletons for gait training in neuromuscular impairments. J Neuroeng Rehabil. 2021;18(1):22. [FREE Full text] [CrossRef] [Medline]
  15. Meena JS, Choi SB, Jung S, Kim J. Electronic textiles: new age of wearable technology for healthcare and fitness solutions. Mater Today Bio. 2023;19:100565. [FREE Full text] [CrossRef] [Medline]
  16. Huang X, Xue Y, Ren S, Wang F. Sensor-based wearable systems for monitoring human motion and posture: a review. Sensors (Basel). 2023;23(22):9047. [FREE Full text] [CrossRef] [Medline]
  17. Adams JL, Dinesh K, Snyder CW, Xiong M, Tarolli CG, Sharma S, et al. A real-world study of wearable sensors in Parkinson's disease. NPJ Parkinsons Dis. 2021;7(1):106. [FREE Full text] [CrossRef] [Medline]
  18. Lu R, Xu Y, Li X, Fan Y, Zeng W, Tan Y, et al. Evaluation of wearable sensor devices in Parkinson's disease: a review of current status and future prospects. Parkinsons Dis. 2020;2020:4693019. [FREE Full text] [CrossRef] [Medline]
  19. Bougea A. Application of wearable sensors in Parkinson’s disease: state of the art. JSAN. 2025;14(2):23. [CrossRef]
  20. Sotirakis C, Su Z, Brzezicki MA, Conway N, Tarassenko L, FitzGerald JJ, et al. Identification of motor progression in Parkinson's disease using wearable sensors and machine learning. NPJ Parkinsons Dis. 2023;9(1):142. [FREE Full text] [CrossRef] [Medline]
  21. Abujrida H, Agu E, Pahlavan K. Machine learning-based motor assessment of Parkinson's disease using postural sway, gait and lifestyle features on crowdsourced smartphone data. Biomed Phys Eng Express. 2020;6(3):035005. [CrossRef] [Medline]
  22. Benyoucef Y, Melliti I, Harmouch J, Asadi B, Del Mastro A, Lapuente-Hernández D, et al. Wearable sensors and artificial intelligence for the diagnosis of Parkinson's disease. J Clin Med. 2025;14(12):4207. [FREE Full text] [CrossRef] [Medline]
  23. Farabolini G, Baldini N, Pagano A, Andrenelli E, Pepa L, Morone G, et al. Continuous movement monitoring at home through wearable devices: a systematic review. Sensors (Basel). 2025;25(16):4889. [FREE Full text] [CrossRef] [Medline]
  24. Islam MS, Rahman W, Abdelkader A, Lee S, Yang PT, Purks JL, et al. Using AI to measure Parkinson's disease severity at home. NPJ Digit Med. 2023;6(1):156. [FREE Full text] [CrossRef] [Medline]
  25. Moreau C, Rouaud T, Grabli D, Benatru I, Remy P, Marques A, et al. Overview on wearable sensors for the management of Parkinson's disease. NPJ Parkinsons Dis. 2023;9(1):153. [FREE Full text] [CrossRef] [Medline]
  26. Del Din S, Kirk C, Yarnall AJ, Rochester L, Hausdorff JM. Body-worn sensors for remote monitoring of Parkinson's disease motor symptoms: vision, state of the art, and challenges ahead. J Parkinsons Dis. 2021;11(s1):S35-S47. [FREE Full text] [CrossRef] [Medline]
  27. Li H, Zecca M, Huang J. Evaluating the utility of wearable sensors for the early diagnosis of Parkinson disease: systematic review. J Med Internet Res. 2025;27:e69422. [FREE Full text] [CrossRef] [Medline]
  28. Gutowski T, Stodulska O, Ćwiklińska A, Gutowska K, Kopeć K, Betka M, et al. Machine learning-based assessment of Parkinson's disease symptoms using wearable and smartphone sensors. Sensors (Basel). 2025;25(16):4924. [FREE Full text] [CrossRef] [Medline]
  29. Wu X, Ma L, Wei P, Shan Y, Chan P, Wang K, et al. Wearable sensor devices can automatically identify the ON-OFF status of patients with Parkinson's disease through an interpretable machine learning model. Front Neurol. 2024;15:1387477. [FREE Full text] [CrossRef] [Medline]
  30. Tischendorf T, Heitmann-Möller A, Ruppert S, Marchwacka M, Schaffrin S, Schaal T, et al. Sustainable integration of digitalisation in nursing education-an international scoping review. Front Health Serv. 2024;4:1344021. [FREE Full text] [CrossRef] [Medline]
  31. AlDossary MBD, Al-Enezi MAH, Aldosary MM. The new frontier in care: a systematic review of nursing implications of remote patient monitoring (RPM) for managing chronic diseases. Saudi J Med Public Health. 2024;1(2):582-589. [FREE Full text] [CrossRef]
  32. Tricco AC, Antony J, Zarin W, Strifler L, Ghassemi M, Ivory J, et al. A scoping review of rapid review methods. BMC Med. 2015;13:224. [FREE Full text] [CrossRef] [Medline]
  33. Miake-Lye IM, Hempel S, Shanman R, Shekelle PG. What is an evidence map? A systematic review of published evidence maps and their definitions, methods, and products. Syst Rev. 2016;5:28. [FREE Full text] [CrossRef] [Medline]
  34. Tricco AC, Lillie E, Zarin W, O'Brien KK, Colquhoun H, Levac D, et al. PRISMA extension for scoping reviews (PRISMA-ScR): checklist and explanation. Ann Intern Med. 2018;169(7):467-473. [FREE Full text] [CrossRef] [Medline]
  35. Rethlefsen ML, Kirtley S, Waffenschmidt S, Ayala AP, Moher D, Page MJ, et al. PRISMA-S Group. PRISMA-S: an extension to the PRISMA statement for reporting literature searches in systematic reviews. Syst Rev. 2021;10(1):39. [FREE Full text] [CrossRef] [Medline]
  36. Higgins JPT, Green S. Cochrane Handbook for Systematic Reviews of Interventions. Chichester (UK). John Wiley & Sons; 2008.
  37. Hong QN, Gonzalez-Reyes A, Pluye P. Improving the usefulness of a tool for appraising the quality of qualitative, quantitative and mixed methods studies, the Mixed Methods Appraisal Tool (MMAT). J Eval Clin Pract. 2018;24(3):459-467. [CrossRef] [Medline]
  38. Aghanavesi S, Westin J, Bergquist F, Nyholm D, Askmark H, Aquilonius SM, et al. A multiple motion sensors index for motor state quantification in Parkinson's disease. Comput Methods Programs Biomed. 2020;189:105309. [CrossRef] [Medline]
  39. Li Y, Bai Q, Yang X, Zhou X, Sun Y, Yao Z. An abnormal gait monitoring system for patients with Parkinson's disease based on wearable devices. Piscataway, NJ. IEEE; 2022. Presented at: 15th International Congress on Image and Signal Processing, BioMedical Engineering and Informatics (CISP-BMEI); 2022 November 05-07:-15; Beijing, China. [CrossRef]
  40. Ghayvat H, Awais M, Geddam R, Quasim MT, Khowaja SA, Dev K. AiCareGaitRehabilitation: multi-modalities sensor data fusion for AI-IoT enabled realtime electrical stimulation device for pre-fog and post-fog to person with Parkinson’s disease. Information Fusion. 2025;122:103155. [CrossRef]
  41. Delgado-Terán JD, Hilbrants K, Mahmutović D, Silva de Lima AL, Wezel RJAV, Heida T. Ankle sensor-based detection of freezing of gait in Parkinson's disease in semi-free living environments. Sensors (Basel). 2025;25(6):1895. [FREE Full text] [CrossRef] [Medline]
  42. Liu X, Zhang X, Li J, Pan W, Sun Y, Lin S, et al. Automated UPDRS gait scoring using wearable sensor fusion and deep learning. Bioengineering (Basel). 2025;12(7):686. [FREE Full text] [CrossRef] [Medline]
  43. Han Y, Liu X, Zhang N, Zhang X, Zhang B, Wang S, et al. Automatic assessments of Parkinsonian gait with wearable sensors for human assistive systems. Sensors (Basel). 2023;23(4):2104. [FREE Full text] [CrossRef] [Medline]
  44. Peraza LR, Kinnunen KM, McNaney R, Craddock IJ, Whone AL, Morgan C, et al. An automatic gait analysis pipeline for wearable sensors: a pilot study in Parkinson's disease. Sensors (Basel). 2021;21(24):8286. [FREE Full text] [CrossRef] [Medline]
  45. Sánchez-Fernández LP, Sánchez-Pérez LA, Martínez-Hernández JM. Computer model for gait assessments in Parkinson's patients using a fuzzy inference model and inertial sensors. Artif Intell Med. 2025;160:103059. [CrossRef] [Medline]
  46. Channa A, Popescu N, Faisal M. Parkinson's disease gait evaluation leveraging wearable insoles and a deep learning approach. IEEE; 2022. Presented at: 8th International Conference on Control, Decision and Information Technologies (CoDIT); 2022 May 17-20:543-549; Istanbul, Turkey. [CrossRef]
  47. Sigcha L, Costa N, Pavón I, Costa S, Arezes P, López JM, et al. Deep learning approaches for detecting freezing of gait in Parkinson's disease patients through on-body acceleration sensors. Sensors (Basel). 2020;20(7):1895. [FREE Full text] [CrossRef] [Medline]
  48. Dvorani A, Wiesener C, Salchow-Hommen C, Jochner M, Spieker L, Skrobot M, et al. On-demand gait-synchronous electrical cueing in Parkinson's disease using machine learning and edge computing: a pilot study. IEEE Open J Eng Med Biol. 2024;5:306-315. [FREE Full text] [CrossRef] [Medline]
  49. Bikias T, Iakovakis D, Hadjidimitriou S, Charisis V, Hadjileontiadis LJ. DeepFoG: an IMU-based detection of freezing of gait episodes in Parkinson's disease patients via deep learning. Front Robot AI. 2021;8:537384. [FREE Full text] [CrossRef] [Medline]
  50. Aich S, Pradhan PM, Chakraborty S, Kim HC, Kim HT, Lee HG, et al. Design of a machine learning-assisted wearable accelerometer-based automated system for studying the effect of dopaminergic medicine on gait characteristics of Parkinson's patients. J Healthc Eng. 2020;2020:1823268. [FREE Full text] [CrossRef] [Medline]
  51. Shi B, Tay A, Au WL, Tan DML, Chia NSY, Yen S. Detection of freezing of gait using convolutional neural networks and data from lower limb motion sensors. IEEE Trans Biomed Eng. 2022;69(7):2256-2267. [CrossRef] [Medline]
  52. Ezhilarasi J, Senthil Kumar T. Develop a novel, faster mask region-based convolutional neural network model with leave-one-subject-out to predict freezing of gait abnormalities of Parkinson’s disease. Neural Comput Applic. 2025;37(7):5441-5457. [CrossRef]
  53. Mahadevan N, Demanuele C, Zhang H, Volfson D, Ho B, Erb MK, et al. Development of digital biomarkers for resting tremor and bradykinesia using a wrist-worn wearable device. NPJ Digit Med. 2020;3:5. [FREE Full text] [CrossRef] [Medline]
  54. Chan LLY, Yang S, Aswani M, Kark L, Henderson E, Lord SR, et al. Development, validation, and limits of freezing of gait detection using a single waist-worn device. IEEE Trans Biomed Eng. 2024;71(10):3024-3031. [CrossRef] [Medline]
  55. Pardoel S, Shalin G, Nantel J, Lemaire ED, Kofman J. Early detection of freezing of gait during walking using inertial measurement unit and plantar pressure distribution data. Sensors (Basel). 2021;21(6):2246. [FREE Full text] [CrossRef] [Medline]
  56. Hssayeni MD, Jimenez-Shahed J, Burack MA, Ghoraani B. Ensemble deep model for continuous estimation of unified Parkinson's disease rating scale III. Biomed Eng Online. 2021;20(1):32. [FREE Full text] [CrossRef] [Medline]
  57. Tsakanikas V, Ntanis A, Rigas G, Androutsos C, Boucharas D, Tachos N, et al. Evaluating gait impairment in Parkinson's disease from instrumented insole and IMU sensor data. Sensors (Basel). 2023;23(8):3902. [FREE Full text] [CrossRef] [Medline]
  58. Brand YE, Schwartz D, Gazit E, Buchman AS, Gilad-Bachrach R, Hausdorff JM. Gait detection from a wrist-worn sensor using machine learning methods: daily living study in older adults and people with Parkinson's disease. Sensors (Basel). 2022;22(18):7094. [FREE Full text] [CrossRef] [Medline]
  59. Pardoel S, Shalin G, Lemaire ED, Kofman J, Nantel J. Grouping successive freezing of gait episodes has neutral to detrimental effect on freeze detection and prediction in Parkinson's disease. PLoS One. 2021;16(10):e0258544. [FREE Full text] [CrossRef] [Medline]
  60. Li D, Hallack A, Gwilym S, Li D, Hu MT, Cantley J. Investigating gait-responsive somatosensory cueing from a wearable device to improve walking in Parkinson's disease. Biomed Eng Online. 2023;22(1):108. [FREE Full text] [CrossRef] [Medline]
  61. Aghanavesi S, Bergquist F, Nyholm D, Senek M, Memedi M. Motion sensor-based assessment of Parkinson's disease motor symptoms during leg agility tests: results from levodopa challenge. IEEE J Biomed Health Inform. 2020;24(1):111-119. [CrossRef] [Medline]
  62. Atrsaei A, Hansen C, Elshehabi M, Solbrig S, Berg D, Liepelt-Scarfone I, et al. Effect of fear of falling on mobility measured during lab and daily activity assessments in Parkinson's disease. Front Aging Neurosci. 2021;13:722830. [FREE Full text] [CrossRef] [Medline]
  63. Shawen N, O'Brien MK, Venkatesan S, Lonini L, Simuni T, Hamilton JL, et al. Role of data measurement characteristics in the accurate detection of Parkinson's disease symptoms using wearable sensors. J Neuroeng Rehabil. 2020;17(1):52. [FREE Full text] [CrossRef] [Medline]
  64. Naghavi N, Wade E. Towards real-time prediction of freezing of gait in patients with Parkinson's disease: a novel deep one-class classifier. IEEE J Biomed Health Inform. 2022;26(4):1726-1736. [CrossRef] [Medline]
  65. Zampogna A, Borzì L, Rinaldi D, Artusi CA, Imbalzano G, Patera M, et al. Unveiling the unpredictable in Parkinson's disease: sensor-based monitoring of dyskinesias and freezing of gait in daily life. Bioengineering (Basel). 2024;11(5):440. [FREE Full text] [CrossRef] [Medline]
  66. Reches T, Dagan M, Herman T, Gazit E, Gouskova NA, Giladi N, et al. Using wearable sensors and machine learning to automatically detect freezing of gait during a FOG-provoking test. Sensors (Basel). 2020;20(16):4474. [FREE Full text] [CrossRef] [Medline]
  67. Slemenšek J, Geršak J, Bratina B, van Midden VM, Pirtošek Z, Šafarič R. Wearable online freezing of gait detection and cueing system. Bioengineering (Basel). 2024;11(10):1048. [FREE Full text] [CrossRef] [Medline]
  68. Borzì L, Mazzetta I, Zampogna A, Suppa A, Irrera F, Olmo G. Predicting axial impairment in parkinson's disease through a single inertial sensor. Sensors (Basel). 2022;22(2):412. [FREE Full text] [CrossRef] [Medline]
  69. Wang W, Lin J, Le X, Li Y, Liu T, Pan L, et al. Addressing multiple challenges in early gait freezing prediction for Parkinson's disease: a practical deep learning approach. IEEE J Biomed Health Inform. 2025;29(9):6251-6262. [CrossRef] [Medline]
  70. Shuqair M, Jimenez-Shahed J, Ghoraani B. Advancing Parkinson's disease management through multi-shared-task self-supervised signal processing. IEEE; 2024. Presented at: 58th Asilomar Conference on Signals, Systems, and Computers; 2024 October 27-30:957-961; Pacific Grove, CA, USA. [CrossRef]
  71. Kanellos FS, Tsamis KI, Rigas G, Simos YV, Katsenos AP, Kartsakalis G, et al. Clinical evaluation in Parkinson's disease: is the golden standard shiny enough? Sensors (Basel). 2023;23(8):3807. [FREE Full text] [CrossRef] [Medline]
  72. Roth N, Küderle A, Ullrich M, Gladow T, Marxreiter F, Klucken J, et al. Hidden Markov model based stride segmentation on unsupervised free-living gait data in Parkinson's disease patients. J Neuroeng Rehabil. 2021;18(1):93. [FREE Full text] [CrossRef] [Medline]
  73. Perez-Ibarra JC, Siqueira AAG, Krebs HI. Identification of gait events in healthy subjects and with Parkinson's disease using inertial sensors: an adaptive unsupervised learning approach. IEEE Trans Neural Syst Rehabil Eng. 2020;28(12):2933-2943. [CrossRef] [Medline]
  74. Meng L, Pang J, Yang Y, Chen L, Xu R, Ming D. Inertial-based gait metrics during turning improve the detection of early-stage Parkinson's disease patients. IEEE Trans Neural Syst Rehabil Eng. 2023;31:1472-1482. [CrossRef] [Medline]
  75. Dattola S, Ielo A, Quartarone A, De Cola MC. Integrating wearable sensor signal processing with unsupervised learning methods for tremor classification in Parkinson's disease. Bioengineering (Basel). 2025;12(1):37. [FREE Full text] [CrossRef] [Medline]
  76. Zhao Y, Liu Y, Lu W, Li J, Shan P, Lian C, et al. Intelligent IoT anklets for monitoring the assessment of Parkinson’s diseases. IEEE Sensors J. 2023;23(24):31523-31536. [CrossRef]
  77. Li Y, Yin J, Liu S, Xue B, Shokoohi C, Ge G, et al. Learning hand kinematics for Parkinson's disease assessment using a multimodal sensor glove. Adv Sci (Weinh). 2023;10(20):e2206982. [FREE Full text] [CrossRef] [Medline]
  78. Yan F, Gong J, Zhang Q, He H. Learning motion primitives for the quantification and diagnosis of mobility deficits. IEEE Trans Biomed Eng. 2024;71(12):3339-3349. [CrossRef] [Medline]
  79. Evers LJW, Raykov YP, Heskes TM, Krijthe JH, Bloem BR, Little MA. Passive monitoring of Parkinson tremor in daily life: a prototypical network approach. Sensors (Basel). 2025;25(2):366. [FREE Full text] [CrossRef] [Medline]
  80. Greene BR, Premoli I, McManus K, McGrath D, Caulfield B. Predicting fall counts using wearable sensors: a novel digital biomarker for Parkinson's disease. Sensors (Basel). 2021;22(1):54. [FREE Full text] [CrossRef] [Medline]
  81. Ricci M, Lazzaro GD, Errico V, Pisani A, Giannini F, Saggio G. The impact of wearable electronics in assessing the effectiveness of levodopa treatment in Parkinson's disease. IEEE J Biomed Health Inform. 2022;26(7):2920-2928. [CrossRef] [Medline]
  82. Ghosh N, Banerjee I. IoT-based freezing of gait detection using grey relational analysis. Internet Things. 2021;13:100068. [CrossRef]
  83. Lin ZR. Quantitative assessment of Parkinson's motor symptoms based on machine learning. Sci Technol Innov Appl. 2022;12(34):50-54. [CrossRef]
  84. Sun Y, Rong Z, Wang F. Automatic quantitative study of motor function in patients with Parkinson's disease based on wearable inertial sensors [in Chinese]. China Med Equip. 2023;38(10):27-32. [FREE Full text]
  85. Li YM. Quantitative Assessment of Parkinson's Motor Symptoms Based on Deep Learning [dissertation]. Lanzhou, China. Lanzhou Jiaotong University; 2023.
  86. Donisi L, Cesarelli G, Balbi P, Provitera V, Lanzillo B, Coccia A, et al. Positive impact of short-term gait rehabilitation in Parkinson patients: a combined approach based on statistics and machine learning. Math Biosci Eng. 2021;18(5):6995-7009. [FREE Full text] [CrossRef] [Medline]
  87. Sotirakis C, Brzezicki MA, Patel S, Conway N, FitzGerald JJ, Antoniades CA. Predicting future fallers in Parkinson's disease using kinematic data over a period of 5 years. NPJ Digit Med. 2024;7(1):345. [FREE Full text] [CrossRef] [Medline]
  88. Hong G, Mao F, Zhang M, Zhang F, Wang X, Ren K, et al. Modeling and validation of wearable sensor-based gait parameters in Parkinson's disease patients with cognitive impairment. Front Aging Neurosci. 2025;17:1590224. [FREE Full text] [CrossRef] [Medline]
  89. Chen M, Sun Z, Xin T, Chen Y, Su F. An interpretable deep learning optimized wearable daily detection system for Parkinson's disease. IEEE Trans Neural Syst Rehabil Eng. 2023;31:3937-3946. [CrossRef] [Medline]
  90. Channa A, Ruggeri G, Ifrim RC, Mammone N, Iera A, Popescu N. Cloud-connected bracelet for continuous monitoring of Parkinson’s disease patients: integrating advanced wearable technologies and machine learning. Electronics. 2024;13(6):1002. [CrossRef]
  91. Singh M, Prakash P, Kaur R, Sowers R, Brašić JR, Hernandez ME. A deep learning approach for automatic and objective grading of the motor impairment severity in Parkinson's disease for use in tele-assessments. Sensors (Basel). 2023;23(21):9004. [FREE Full text] [CrossRef] [Medline]
  92. Weikert T, Li Y, Paez-Granados D, Easthope CA. Automated prediction of item-level ARAT scores from wearable sensors. IEEE; 2025. Presented at: Proceedings of the 2025 IEEE International Conference on Rehabilitation Robotics (ICORR); 2025 May 12–16:1239-1244; Chicago, IL, USA. [CrossRef]
  93. Yang PK, Filtjens B, Ginis P, Goris M, Nieuwboer A, Gilat M, et al. Freezing of gait assessment with inertial measurement units and deep learning: effect of tasks, medication states, and stops. J Neuroeng Rehabil. 2024;21(1):24. [FREE Full text] [CrossRef] [Medline]
  94. Esfahani AH, Dyka Z, Ortmann S, Langendorfer P. Impact of data preparation in freezing of gait detection using feature-less recurrent neural network. IEEE Access. 2021;9:138120-138131. [CrossRef]
  95. Sun H, Ye Q, Xia Y. Predicting freezing of gait in patients with Parkinson's disease by combination of manually-selected and deep learning features. Biomed Signal Process Control. 2024;88:105639. [CrossRef]
  96. Dvorani A, Waldheim V, Jochner MCE, Salchow-Hömmen C, Meyer-Ohle J, Kühn AA, et al. Real-time detection of freezing motions in Parkinson's patients for adaptive gait phase synchronous cueing. Front Neurol. 2021;12:720516. [FREE Full text] [CrossRef] [Medline]
  97. Khan U, Riaz Q, Hussain M, Zeeshan M, Krüger B. Towards effective Parkinson’s monitoring: movement disorder detection and symptom identification using wearable inertial sensors. Algorithms. 2025;18(4):203. [CrossRef]
  98. Park H, Youm C, Cheon SM, Kim B, Choi H, Hwang J, et al. Using machine learning to identify Parkinson's disease severity subtypes with multimodal data. J Neuroeng Rehabil. 2025;22(1):126. [FREE Full text] [CrossRef] [Medline]
  99. Goubault E, Martin C, Duval C, Daneault JF, Boissy P, Lebel K. Enhanced detection and segmentation of sit phases in patients with Parkinson's disease using a single smartwatch and random forest algorithms. Sensors (Basel). 2025;25(19):6104. [FREE Full text] [CrossRef] [Medline]
  100. Liu T, Li Z, Ji X, Xu M, Wang L, Chi Y. Friction nanogenerators based on asymmetrically adherent conductive hydrogels for multifunctional sensing and Parkinsonian gait diagnostics. Chem Eng J. 2025;524:169026. [CrossRef]
  101. Nanayakkara T, Herath HMKKMB, Malekroodi HS, Madusanka N, Yi M, Lee BI. Multi-domain CoP feature analysis of functional mobility for Parkinson's disease detection using wearable pressure insoles. Sensors (Basel). 2025;25(18):5859. [FREE Full text] [CrossRef] [Medline]
  102. Ma L, Lin S, Jin J, Wang Z, Wang X, Chen Z, et al. Objective assessment of gait and posture symptoms in Parkinson's disease using wearable sensors and machine learning. Front Aging Neurosci. 2025;17:1618764. [FREE Full text] [CrossRef] [Medline]
  103. Lopes T, Reis Carneiro M, Morgadinho A, Reis Carneiro D, Tavakoli M. ParCuR-a novel AI-enabled gait cueing wearable for patients with Parkinson's disease. Sensors (Basel). 2025;25(22):7077. [FREE Full text] [CrossRef] [Medline]
  104. Channa A, Popescu N, Ciobanu V. Wearable solutions for patients with Parkinson's disease and neurocognitive disorder: a systematic review. Sensors (Basel). 2020;20(9):2713. [FREE Full text] [CrossRef] [Medline]
  105. Mughal H, Javed AR, Rizwan M, Almadhor AS, Kryvinska N. Parkinson’s disease management via wearable sensors: a systematic review. IEEE Access. 2022;10:35219-35237. [CrossRef]
  106. Sigcha L, Borzì L, Amato F, Rechichi I, Ramos-Romero C, Cárdenas A, et al. Deep learning and wearable sensors for the diagnosis and monitoring of Parkinson’s disease: a systematic review. Expert Syst Appl. 2023;229:120541. [CrossRef]
  107. Joo JH, Lieu N, Tang Y, Browne DS, Agusala B, Liao JM. Trends in utilization of remote monitoring in the United States. Health Aff Sch. 2025;3(6):qxaf115. [CrossRef] [Medline]
  108. Tan SY, Sumner J, Wang Y, Wenjun Yip A. A systematic review of the impacts of remote patient monitoring (RPM) interventions on safety, adherence, quality-of-life and cost-related outcomes. NPJ Digit Med. 2024;7(1):192. [FREE Full text] [CrossRef] [Medline]
  109. Hsueh WT, Yu CX, Cheng HC, Chen MY, Wang HM, Fu LM. A comprehensive review of wearable devices for non-invasive biosensing. Trends Anal Chem. 2025;193:118425. [CrossRef]
  110. Moor M, Banerjee O, Abad ZSH, Krumholz HM, Leskovec J, Topol EJ, et al. Foundation models for generalist medical artificial intelligence. Nature. 2023;616(7956):259-265. [CrossRef] [Medline]
  111. Smuck M, Odonkor CA, Wilt JK, Schmidt N, Swiernik MA. The emerging clinical role of wearables: factors for successful implementation in healthcare. NPJ Digit Med. 2021;4(1):45. [FREE Full text] [CrossRef] [Medline]
  112. Xu H, Liu J, Hu H, Zhang Y. Wearable sensor-based human activity recognition method with multi-features extracted from Hilbert-Huang transform. Sensors (Basel). 2016;16(12):2048. [FREE Full text] [CrossRef] [Medline]
  113. Ordóñez FJ, Roggen D. Deep convolutional and LSTM recurrent neural networks for multimodal wearable activity recognition. Sensors (Basel). 2016;16(1):115. [FREE Full text] [CrossRef] [Medline]
  114. Fiordelli M, Diviani N, Schulz PJ. Mapping mHealth research: a decade of evolution. J Med Internet Res. 2013;15(5):e95. [FREE Full text] [CrossRef] [Medline]
  115. Evers LJW, Peeters JM, Bloem BR, Meinders MJ. Need for personalized monitoring of Parkinson's disease: the perspectives of patients and specialized healthcare providers. Front Neurol. 2023;14:1150634. [FREE Full text] [CrossRef] [Medline]
  116. Toney-Butler TJ, Thayer JM. Nursing Process Internet. Treasure Island (FL). StatPearls Publishing; 2023.
  117. Kwon SH, Park JK, Koh YH. A systematic review and meta-analysis on the effect of virtual reality-based rehabilitation for people with Parkinson's disease. J Neuroeng Rehabil. 2023;20(1):94. [FREE Full text] [CrossRef] [Medline]
  118. Chuang CS, Chen YW, Zeng BY, Hung CM, Tu YK, Tai YC, et al. Effects of modern technology (exergame and virtual reality) assisted rehabilitation vs conventional rehabilitation in patients with Parkinson's disease: a network meta-analysis of randomised controlled trials. Physiotherapy. 2022;117:35-42. [CrossRef] [Medline]
  119. Yau CE, Ho ECK, Ong NY, Loh CJK, Mai AS, Tan EK. Innovative technology-based interventions in Parkinson's disease: a systematic review and meta-analysis. Ann Clin Transl Neurol. 2024;11(10):2548-2562. [FREE Full text] [CrossRef] [Medline]
  120. Bowman T, Gervasoni E, Arienti C, Lazzarini SG, Negrini S, Crea S, et al. Wearable devices for biofeedback rehabilitation: a systematic review and meta-analysis to design application rules and estimate the effectiveness on balance and gait outcomes in neurological diseases. Sensors (Basel). 2021;21(10):3444. [FREE Full text] [CrossRef] [Medline]
  121. Giannakopoulou KM, Roussaki I, Demestichas K. Internet of things technologies and machine learning methods for Parkinson's disease diagnosis, monitoring and management: A systematic review. Sensors (Basel). 2022;22(5):1799. [FREE Full text] [CrossRef] [Medline]
  122. Mathias R, McCulloch P, Chalkidou A, Gilbert S. Digital health technologies need regulation and reimbursement that enable flexible interactions and groupings. NPJ Digit Med. 2024;7(1):148. [FREE Full text] [CrossRef] [Medline]


AI: artificial intelligence
CNKI: China National Knowledge Infrastructure
CNN: convolutional neural network
DL: deep learning
LSTM: long short-term memory
MeSH: Medical Subject Headings
ML: machine learning
PD: Parkinson disease
PRISMA: Preferred Reporting Items for Systematic Reviews and Meta-Analyses
PRISMA-S: Preferred Reporting Items for Systematic Reviews and Meta-Analyses literature search extension
PRISMA-ScR: Preferred Reporting Items for Systematic Reviews and Meta-Analyses extension for Scoping Reviews
SVM: support vector machine


Edited by S Brini; submitted 10.Oct.2025; peer-reviewed by O Oyetunji, S Jiang, S Mohanadas, S Ajayi; comments to author 01.Dec.2025; accepted 04.Feb.2026; published 26.Feb.2026.

Copyright

©Shengting Li, Siqi Chen, Xiaosong Yu, Huixiang Shang, Tingting Tu, Mingtao Quan. Originally published in the Journal of Medical Internet Research (https://www.jmir.org), 26.Feb.2026.

This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in the Journal of Medical Internet Research (ISSN 1438-8871), is properly cited. The complete bibliographic information, a link to the original publication on https://www.jmir.org/, as well as this copyright and license information must be included.